Emerson Micro Motion ELITE CMF200A Installation Manual page 19

Elite sensor installation manual
Table of Contents

Advertisement

Figure 14
Micro Motion cable gland and heat shrink
4. For connection at the core processor housing, prepare shielded cable as follows (for armored cable,
omit steps d, e, f, and g):
a. Strip 4 1/2 inches (114 mm) of cable jacket.
b. Remove the clear wrap that is inside the cable jacket, and remove the filler material between the
wires.
c. Remove the foil shield that is around the insulated wires, leaving 3/4 inch (19 mm) of foil or braid
and drain wires exposed, and separate the wires.
d. Wrap the shield drain wire(s) around the exposed foil twice. Cut off the excess wire.
Figure 15
Wrapping the shield drain wires
e. Place the shielded heat shrink over the exposed shield drain wire(s). The tubing should completely
cover the drain wires.
f. Without burning the cable, apply heat (250 °F or 120 °C) to shrink the tubing.
Installation Manual
3/4 in
(19 mm)
Gland clamping
Gland nut
insert
7/8 in
(22 mm)
Shielded heat shrink
4 1/2 in
(114 mm)
7/8 in
(22 mm)
Gland body
Wiring
17