B) Internal Offset Circuit Test
Press the pH7 TEST key and the display
should show a value within 7±1 pH, to
verify the internal circuit of the meter in
terms of the offset compensation.
pH 7
TEST
C) Amplifier Circuit Test
Press the pH4 TEST key and the display
should show a value within the 3.30 to
4.30 pH range, to verify the amplifier circuit
of the meter.
pH 4
TEST
38
pH
pH
38£

ORP DIAGNOSTIC TESTS

HI 8512 and HI 8720 are ORP controllers
provided with autodiagnostic functions that
allow to check and troubleshoot any
malfunctioning.
The functions are made via front panel keys
to isolate the cause of malfunction.
For HI 8720 only, press
MEASURE
key
proceeding with the following
tests.
A) 0 mV Test
Press the 0 mV TEST
key and the display
should show a value of
0±10 mV, to verify the
"zero" calibration of the
instrument.
B) 500 mV Test
Press the 500 mV TEST
key and the display should
show a value of 500±20
mV, to verify the slope
at 500 mV.
39
MEA
SURE
COARSE
before
SET
FINE
mV
0 mV
500 mV
TEST
TEST
mV
0 mV
500 mV
TEST
TEST