Specifications - Hanna Instruments HI 8510 Instruction Manual

Panel mounted ph and orp indicators and controllers
Hide thumbs Also See for HI 8510:
Table of Contents

Advertisement

SPECIFICATIONS

HI 8510E
Range
Resolution
Accuracy
±0.02 pH
Typical EMC Dev.
Installation Category
Input
10
12
Offset: ±2 pH with Δ0 trimmer
Calibration
Slope: 80 to 110% with slope trimmer
Temperature
Fixed or automatic with Pt100
Compensation
from 0 to 100°C (32 to 212°F)
Recorder Output
0-20 mA or 4-20 mA (isolated)
Power Supply
115 or 230 Vac; 50/60 Hz
Environment
-10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout
141 x 69 mm (5.6 x 2.7'')
Weight
HI 8512E
Range
Resolution
Accuracy
±5 mV
Typical EMC Dev.
Installation Category
Input
10
12
Calibration
Slope: 90 to 110% with slope trimmer
Recorder Output
0-20 mA or 4-20 mA (isolated)
Power Supply
115 or 230 Vac; 50/60 Hz
Environment
-10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout
141 x 69 mm (5.6 x 2.7'')
Weight
24
HI 8510T
0.00 to 14.00 pH
0.01 pH
±0.5%
±0.1 pH / ±0.2 mA
II
Ohm
4 to 20 mA
1 kg (2.2 lb.)
HI 8512T
±1000 mV
1 mV
±0.5%
±6 mV / ±0.2 mA
II
Ohm
4 to 20 mA
1 kg (2.2 lb.)
24£
HI 8710E
Range
Resolution
Accuracy
±0.02 pH
Typical EMC Dev.
Installation Category
Input
10
12
Offset: ±2 pH with Δ0 trimmer
Calibration
Slope: 80 to 110% with slope trimmer
Temperature
Fixed or automatic with Pt100
Compensation
from 0 to 100°C (32 to 212°F)
Relays
1 for setpoint and 1 for alarm,
max 2A, 240 V resistive load (isolated)
Recorder Output
0-20 mA or 4-20 mA (isolated)
Power Supply
115 or 230 Vac; 50/60 Hz
Environment
-10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout
141 x 69 mm (5.6 x 2.7'')
Weight
HI 8711E
Range
Resolution
Accuracy
±0.02 pH
Typical EMC Dev.
Installation Category
Input
10
12
Offset: ±2 pH with Δ0 trimmer
Calibration
Slope: 80 to 110% with slope trimmer
Temperature
Fixed or automatic with Pt100
Compensation
from 0 to 100°C (32 to 212°F)
Relays
2 for setpoint and 1 for alarm,
max 2A, 240 V resistive load (isolated)
Recorder Output
0-20 mA or 4-20 mA (isolated)
Power Supply
115 or 230 Vac; 50/60 Hz
Environment
-10 to 50°C (14 to 122°F);
RH max 95% non condensing
Panel Cutout
141 x 69 mm (5.6 x 2.7'')
Weight
25
HI 8710T
0.00 to 14.00 pH
0.01 pH
±0.5%
±0.1 pH / ±0.2 mA
II
Ohm
4 to 20 mA
1 kg (2.2 lb.)
HI 8711T
0.00 to 14.00 pH
0.01 pH
±0.5%
±0.1 pH / ±0.2 mA
II
Ohm
4 to 20 mA
1 kg (2.2 lb.)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hi 8512Hi 8710Hi 8711Hi 8720

Table of Contents