Set Rain Cal; Set Temp & Hum Cal - DAVIS Weather Envoy 6314 Installation Manual

Table of Contents

Advertisement

S
I
O F T W A R E
N S T A L L A T I O N A N D
Station Setup

Set Rain Cal

Based on the type of rain collector you selected in station configuration (See
"Station Configuration" on page 13), the software automatically changes your
station console's rainfall calibration number to the correct setting.
1. Choose Set Rain Cal from the Setup
menu.
The software prompts you to confirm
that you wish to change the rainfall cal-
ibration number.
2. Choose Yes.
The software sets the rainfall calibration
number on your station.
Set Temp & Hum Cal
If you need to adjust your temperature and humidity readings, you can change
the console the temperature calibration numbers using WeatherLink.
1. Choose Set Temp Hum Cal from the Setup menu.
The Set Temp Cal dialog box
appears. The raw reading shown in
this dialog box indicates the unad-
justed reading coming from the sen-
sor. The adjusted reading indicates
what the software displays,
adjusted for any previously entered
calibration number.
2. Enter the adjusted reading for the
variable you want and choose OK.
The software updates the station's
calibration numbers. The station
and software will accept tempera-
ture adjustments up to ±12.7 ºF (~
7.1 ºC) away from the raw reading.
For example, if Weather Envoy is
recording an inside temperature
reading of 60º F (15.6 ºC), you could set the temperature anywhere between
47.3ºF (8.6 ºC) and 72.7ºF (22.6 ºC). (60-12.7=47.3, 60+12.7=72.7)
Note: Weather Envoy minimum and maximum temperature settings are –90ºF and +164ºF.
The Weather Envoy accepts humidity calibrations between 0 and 100%.
You can use the Clear dialog box to reset all temperature and humidity cali-
bration numbers at once.
18
S
E T U P
Remote Computer Installation
Set Temp & Hum Cal

Advertisement

Table of Contents
loading

This manual is also suitable for:

6314cVantage pro weather envoy6314

Table of Contents