Specifications - Fluke ProcessMeter 787 User Manual

Processmeter
Hide thumbs Also See for ProcessMeter 787:
Table of Contents

Advertisement

787
Users Manual

Specifications

All specifications apply from +18 C to +28 C unless
stated otherwise.
All specifications assume a 5 minute warmup period.
DC Volts Measurement
Range (V dc)
4.000
40.00
400.0
1000
Input impedance: 10 M (nominal), < 100 pF
Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz
Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz
Overvoltage protection: 1000V
36
Resolution
0.001V
0.01V
0.1V
1V
The standard specification interval is 1 year.
"Counts" means number of increments or
decrements of the least significant digit.
Accuracy, (% of Reading + Counts)
0.1% + 1
0.1% + 1
0.1% + 1
0.1% + 1
Note