787
Users Manual
AC Current Measurement
Range
45 Hz to 2 kHz
1.000A (Note)
Note: 440 mA continuous, 1A 30 seconds maximum
Specifications are valid from 5% to 100% of amplitude range.
AC conversion: true rms
Maximum crest factor: 3
For non-sinusoidal waveforms, add (2% reading + 2% f.s.) typical
Overload protection 440 mA, 1000V fast-blow fuse
DC Current Measurement
Range
30.000 mA
1.000A (Note)
Note: 440 mA continuous, 1A 30 seconds maximum
Overload protection: 440 mA, 1000V fast-blow fuse
38
Resolution
Accuracy, (% of Reading + Counts)
0.001A
Resolution
Accuracy, (% of Reading + Counts)
0.001 mA
0.001A
1% + 2
0.05% + 2
0.2% + 2
Typical Burden
Voltage
1.5V/A
Typical Burden
Voltage
14 mV/mA
1.5V/A