Frequency Counter Specifications

Table 7-13 Frequency counter (divide by 1) specifications

Range
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz
Table 7-14 Frequency counter (divide by 100
Range
9.9999 MHz
99.999 MHz
Notes for frequency counter specifications:
1 All frequency counters are susceptible to error when measuring low-voltage, low-frequency signals. Shielding inputs
from picking up external noise is critical for minimizing measurements errors. For non-square wave signals, an additional
5 counts needs to be added.
2 The maximum measurement level is < 30 V pp.
3 The minimum measurement frequency of low frequency is set by power-on option to speed up the measurement rate.
4 Shown on the secondary display.
U1253B User's and Service Guide
Resolution
0.001 Hz
0.02% + 3
0.01 Hz
0.0001 kHz
0.002% + 5
0.001 kHz
0.01 kHz
[4]
) specifications
Resolution
0.0001 MHz
0.002% + 5
0.001 MHz
Accuracy
Sensitivity
[1]
100 mVrms
< 985 kHz
200 mVrsm
Accuracy
Sensitivity
400 mVrms
< 20 MHz
600 mVrms
Specifications
7
Minimum input freq.
0.5 Hz
Minimum input freq.
1 MHz
191