Adjustments; Zero Adjustment - Agilent Technologies 34405A User's And Service Manual

5 1/2 digit multimeter
Hide thumbs Also See for 34405A:
Table of Contents

Advertisement

Adjustments

N O TE

Zero Adjustment

CAU T ION
34405A User's and Service Guide
You will need a test input cable and connectors set, and a
Shorting Plug to adjust the instrument (see
Connections"
on page 72).
After each adjustment finishes successfully, the primary display briefly
shows PASS. If the calibration fails, the multimeter beeps, the primary
display shows FAil and an error number is shown in the secondary
display. Calibration error messages are described on
of a calibration failure, correct the problem and repeat the procedure.
Each time you perform a zero adjustment, the instrument
stores a new set of offset correction constants for
measurement functions and ranges. The instrument will
sequence through all required functions and ranges
automatically and store new zero offset calibration constants.
Never turn off the instrument during Zero Adjustment. This may cause
ALL calibration memory to be lost.
Zero Adjustment Procedure
Be sure to allow the instrument to warm up and stabilize
for 2 hours before performing the adjustments.
1 Follow the steps outlined below. Review
Considerations"
on page 71 before beginning this test.
2 After unsecuring the instrument, the instrument goes into
the Adjustment Mode (as indicated by the CAL
annunciator) with the secondary display showing Short.
Connect the Shorting Plug (see
Performance Tests and Calibration
"Input
page
"Test
page
73) between the HI
4
110. In the event
93