MESSAGE
remove
strip
100 mg/dl
(Example)
contrl
7: 5 2am 7-01-00
error 1
retest
error2
retest
PROBLEM
3. The meter was used in very
bright light.
The test strip was inserted
before the word
i n s e r t
appeared on the display.
If the word
contrl
after a blood test, your
blood sample was too small,
smeared, or another drop was
added after the test began.
The sample was applied
before the words
apply
sample
appeared.
1. The test strip moved
during the test.
WHAT TO DO
3. Move the meter away from
the light source. Repeat the
test with a new test strip.
Repeat the test with a new test
strip. Wait for the word
before inserting a test strip.
appears
Repeat the test with a new test
strip and a larger drop of
blood.
Repeat the test with a new
test strip.
1. Repeat the test with a new
test strip.
77
insert