Co2 incubators/with o2 control/with humidity control, with sterilizable ndir sensor system for co2 and microprocessor display program controller mb2 with touchscreen
Cb (e7), co2 incubators, co2 incubators with o2 control, co2 incubators with active humidification.
with sterilizable ndir sensor system for co2 and microprocessor program controller mb2 (175 pages)
Cb (e6.1)
co2 - incubators
co2 – incubators with o2 control
with sterilizable ndir sensor system for co2
and microprocessor controller t4.12 (145 pages)
Cb (e6.1)
co2 - incubators
co2 – incubators with o2 control
with sterilizable ndir sensor system for co2
and microprocessor controller t4.12 (145 pages)
Cb (e6.1)
co2 - incubators
co2 – incubators with o2 control
with sterilizable ndir sensor system for co2
and microprocessor controller t4.12 (145 pages)
Model version: bd056-230v;bd056ul-120v;bd115-230v;bd115ul-120v;bd260-230v;bd260ul-120v;bf056-230v;bf056ul-120v;bf115-230v;bf115ul-120v;bf260-230v;bf260ul-120v;ed056-230v;ed056ul-120v;ed115-230v;ed115ul-120v;ed260-230v;ed260ul-240v;fd056-230v;fd056ul-120v; (99 pages)
Model version:kbf115-230v;kbf115-240v;kbf240-230v;kbf240-240v;kbf720-230v;kbf720-240v;kbf1020-230v;kbf1020-240v;kmf115-230v;kmf115-240v;kmf240-230v;kmf240-240v;kmf720-230v;kmf720-240v.
constant climate chambers with program control.constant climate chamb (155 pages)
Mkf (e3.2)
alternating climate chambers with program control.
model version:mkf115-400v;mkf115-400v-c;mkf240-400v;mkf240-400v-c;mkf720-400v;mkf720-400v-c (115 pages)
Page 1
Fax: +49 7462 2005 100 Internet: http://www.binder-world.com E-mail: info@binder-world.com Service Hotline: +49 7462 2005 555 Service Fax: +49 7462 2005 93 555 Service E-Mail: customerservice@binder-world.com Service Hotline USA: +1 866 885 9794 or +1 631 224 4340 x3 ...
Page 2
Water supply via freshwater bag (standard) ................43 4.5.2.1 Scope of delivery ......................43 4.5.2.2 Installation and connection of the freshwater bag ............44 4.5.2.3 Refilling the already installed freshwater bag ..............45 CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 2/197...
Page 3
....................69 Activating / deactivating CO and humidity control ..............69 CB / CB-UL: Humidity control of the Permadry™ system ..............70 SET-POINT ENTRY IN “FIXED VALUE” OPERATING MODE ......71 Set-point entry through the “Setpoints” menu ................... 72 Direct setpoint entry via Normal display ....................
Page 4
13.3 Log out ............................115 13.4 User change ............................ 115 13.5 Password assignment and password change................. 116 13.5.1 Password change ......................... 116 13.5.2 Deleting the password for an individual authorization level ..........118 CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 4/197...
Page 5
18.2 Silicone access ports 30 mm / 1.18 in, closable from both sides with silicon plugs (option) ..138 18.3 Quick sample access (option) ......................139 18.4 Interior socket 230V (option, available via BINDER INDIVIDUAL Customized Solutions) ..... 140 18.5 Analog outputs for temperature and CO (option) ................
Page 6
Humidity (CBF / CBF-UL) ..................... 171 22.6.8 Controller ..........................171 22.6.9 Open door ..........................172 22.7 Sending the chamber back to BINDER GmbH ................172 23. DISPOSAL......................173 23.1 Disposal of the transport packing ....................173 23.1.1 Outer chamber packing......................173 23.1.2...
Page 7
24.9.3 Chamber size 260 ......................... 187 25. CERTIFICATES AND DECLARATIONS OF CONFORMITY ......188 25.1 EU Declaration of conformity for CB ....................188 25.2 EU Declaration of conformity for CBF ..................... 190 26. CONTAMINATION CLEARANCE CERTIFICATE ..........192 26.1 For chambers located outside the USA and Canada ..............192 26.2 For chambers in the USA and Canada ...................
Page 8
Operating Manual. This Operating Manual is supplemented and updated as needed. Always use the most recent version of the Operating Manual. When in doubt, call the BINDER Service Hotline for information on the up-to- dateness and validity of this Operating Manual.
Page 9
In no event shall BINDER be held liable for any damages, direct or incidental arising out of or related to the use of this manual. This operating manual cannot cover all conceivable applications. If you would like additional information, or if special problems arise that are not sufficiently addressed in this manual, please ask your dealer or contact us directly, e.g.
Page 10
Wear protective gloves Environment protection Wear safety goggles Prohibition signs Do NOT touch Do NOT spray with water Information to be observed in order to ensure optimum function of the product. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 10/197...
Page 11
Service label Hot surface • on the outer chamber door Risk of injury • on the outer door: CB-UL / CBF-UL only • above the access ports (option) Magnetic field • on the outer door above the door handle incubator CB / CBF...
Page 12
Figure 2: Position of additional labels on the CBF / CBF-UL Keep safety labels complete and legible. Replace safety labels that are no longer legible. Contact BINDER service for these replacements. Type plate Position of type plate: left chamber side (seen from front), at the bottom in the middle Nominal temp.
Page 13
(for Germany: DGUV guidelines 213-850 on safe working in laboratories, issued by the employers’ liability insurance association). BINDER GmbH is only responsible for the safety features of the chamber provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts.
Page 14
Danger of injury and damages by the chamber tipping over. Injuries and damage to the chamber and the charging material ∅ Do NOT load the chamber door with heavy objects while it is open. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 14/197...
Page 15
∅ Do NOT set up chambers in non-ventilated recesses. Ensure technical ventilation measures. Observe the relevant regulations for handling O Close the O supply when decommissioning the chamber. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 15/197...
Page 16
∅ Do NOT set up chambers in non-ventilated recesses. Ensure technical ventilation measures. Observe the relevant regulations for handling N Close the N supply when decommissioning the chamber. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 16/197...
Page 17
Close the gas cylinder valve before connecting or removing the gas hose. After connecting the gas cylinder, check all gas connections for leaks (e.g. with leak spray or diluted soap solution). CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 17/197...
Page 18
Other applications than those described in this chapter are not approved. CB / CB-UL and CBF / CBF-UL incubators are suitable for the cultivation of mammal cells under typical conditions of approx. 37 °C / 98.6°F. The chamber permits setting defined pH conditions by common...
Page 19
The requirements described in the Operating Manual for installation site and ambient conditions (Chap. 3.4) must be met. WARNING: If customer should use a BINDER chamber running in non-supervised continuous operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples to split such specimen or samples and store them in at least two chambers, if this is feasible.
Page 20
• Contact with live parts in normal state Cleaning and Decontamination • Penetration of water into the chamber • Inappropriate cleaning and decontamination agents • Enclosure of persons in the interior CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 20/197...
Page 21
• Absence of a plausibility check to rule out erroneous inscription of electrical components • Performance of repair work by untrained/insufficiently qualified personnel • Inappropriate repairs which do not meet the quality standard specified by BINDER • Use of replacement parts other than BINDER original replacement parts •...
Page 22
Warn wearers of such devices of approaching the chamber door. • Protection against touchable surfaces Tested according to EN ISO 13732-1:2008. • Floors See operating manual chap. 3.4 for correct installation • Cleaning See operating manual chap. 21. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 22/197...
Page 23
• Vapors of oil and fat are dangerous for the sensor because they may condensate at the sensor and thus prevent its function (insulating layer). For similar reasons it is not possible to measure smoke gases. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 23/197...
Page 24
Chamber description The CO incubators CB / CB-UL and CBF / CBF-UL are equipped with a multifunctional microprocessor display controller for temperature, CO , and O (chamber with O control) levels and humidity (CBF / CBF- UL) and a digital display accurate to one-tenth of a degree resp. 0.1 vol.-%. With its comprehensive program...
Page 25
A resistance humidifying system humidifies the air. For this purpose, use deionized (demineralized) water. The option BINDER Pure Aqua Service allows using the chamber with any degree of water hardness. Controller: The efficient program controller is equipped with a multitude of operating functions, in addition to recorder and alarm functions.
Page 26
Pt 100 temperature probe sensor (chamber with O control) Humidity sensor (CBF / CBF-UL) Internal socket 230V (max. 3 A) (option, available via BINDER INDIVIDUAL Customized Solutions, chap. 18.4) Perforated shelves, made of stainless steel Glass door handle Measuring access port Inner glass doors Permadry™...
Page 27
Opening for freshwater bag (CBF / CBF-UL) Control panel on the rear of the chamber (10) (11) (1b) (6b) Figure 6: Rear control panel CB with O control and options (1b) (10) (11) (6a) Figure 7: Rear control panel CB-UL with O...
Page 28
If you need to return the chamber, please use the original packing and observe the guidelines for safe lifting and transportation (chap. 3.2). For disposal of the transport packing, see chap. 23.1. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 28/197...
Page 29
Note on second-hand chambers (Ex-Demo-Units): Second-hand chambers are chambers that were used for a short time for tests or exhibitions. They are thoroughly tested before resale. BINDER ensures that the chamber is technically sound and will work flawlessly. Second-hand chambers are marked with a sticker on the chamber door. Please remove the sticker before commissioning the chamber.
Page 30
The chambers can be stacked on top of each other (two chambers maximum). For safe stacking that is easy to maintain, use the original BINDER stacking stand (chap. 18.9.1) or the base on castors (chap. 18.9.2).
Page 31
Vent out any CO gas that may escape via good room ventilation or a suitable connection to an exhaust system. We recommend installing a CO warning system. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 31/197...
Page 32
Ensure technical ventilation measures. Observe the relevant regulations for handling O Take appropriate measures to prevent oxygen enrichment and fire and explosion hazards in areas where oxygen enrichment is possible. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 32/197...
Page 33
Permitted shelf loads: • Maximum load of one single shelf: 10 kg / 22 lb. • Maximum total load of all shelves: 30 kg / 66 lb. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 33/197...
Page 34
Permadry™ water pan The Permadry™ system developed by BINDER is an effective and easy to handle system that ensures high humidity inside the chamber without any condensation forming on the inner surfaces. The Permadry™ water pan consists of two pans in which the outer one is heated and the inner one cooled. With the slight difference of temperature caused by that cooling, the central pan is the specific point for condensation of the surplus humidity.
Page 35
Risk of injury through sudden release of the stored pressure energy when opening the cylinder valve of a not connected cylinder. Injuries. Close the gas cylinder valve before connecting or removing the gas hose. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 35/197...
Page 36
Connect the pre-assembled hose nozzle of the gas hose to the quick acting closure socket (10) DN 6 on the chamber rear, as described in chap. 4.4.4. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 36/197...
Page 37
The real outlet pressure of gas cylinders, sets of gas cylinders or central gas supplies am on the second manometer must not exceed 2.5 bar / 36 psi. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 37/197...
Page 38
99.5 %. The gas connections must be established by qualified personnel who are trained in handling the respective gases and familiar with the required safety measures. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 38/197...
Page 39
Fire and explosion hazard through contact of fat with O Serious injury or death from burns and / or explosion pressure. ∅ All connection parts for the O connection must be degreased. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 39/197...
Page 40
Only use the supplied hose nozzle to connect to the quick acting closure socket. Otherwise, the quick acting closure socket may leak, and/or it may become impossible to connect the original hose nozzle. In this case, please contact BINDER Service. Remove the rubber cover (b) by pulling it off.
Page 41
Adjust the outlet pressure to 2.0 bar / 29 psi above the ambient pressure. The gas connections must be established by qualified personnel who are trained in handling the respective gases and familiar with the required safety measures. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 41/197...
Page 42
1 µS/cm (ultrapure water), may cause acid corrosion due to its low pH). • Tap water that has been treated by the optional water treatment system BINDER Pure Aqua Service (disposable system). A reusable measuring equipment to assess the water quality is included (chap.
Page 43
4.5.1.4 BINDER Pure Aqua Service (option) The optional BINDER water treatment system (disposable system) is available to treat tap water. The lifetime depends on water quality and the amount of treated water used. The measuring equipment to assess the water quality is reusable.
Page 44
32 cm, the tube end for connecting the freshwater bag has a length of approx. 18 cm • Press the tube retainer (12b) into the pump head cover until it snaps into place CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 44/197...
Page 45
• Fill the freshwater bag. Observe water quality and temperature (chap. 4.5.1.3). • Remove the filling tube from the tube connector on the freshwater bag • Reassemble the two parts of the tube connector ①. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 45/197...
Page 46
NW 4 mm Combination 2: Opening Opening (recommended) NW 4 mm NW 7 mm Connecting the freshwater bag Put the filled freshwater bag on the nozzle of the hose connector. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 46/197...
Page 47
③ ② (12b) (12a) Figure 18: Installation of the water supply via prefilled water bottle (12) Pump head (12c) Rotor (12a) Pump head cover (13) Pump connector (12b) Tube retainer CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 47/197...
Page 48
DN 6 mm (transfer set) bottle through the silicone seal. ⑦ • Then remove the second protective cap of the spike connector and insert the piercing pin into the water CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 48/197...
Page 49
∅ Do NOT place the freshwater can tap on or above the rear panel of the chamber Figure 20: Two CBF / CBF-UL incubators with installed water supply via freshwater can (option) CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 49/197...
Page 50
(12a). ① • Lead the silicone tube along the rotor. Insert the rotor with the tube into the pump head cover. Keep the hose tight and do not pinch it.② CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 50/197...
Page 51
Silicone tube DN 3mm to the pump (or via T- type tube connector (T-piece) to another (19a) Removal button chamber) (20) Angular coupling plug with aqua stop (12) Pump (21) Silicone tube d 10 CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 51/197...
Page 52
Alternatively, you can remove the lid and fill the can on site with a suitable aid (water hose, watering can). When filling on site, the existing hose connection can remain on the freshwater can. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 52/197...
Page 53
• Use only the original connection cables from BINDER according to the above specification. • Prior to connection and start-up, check the power supply voltage. Compare the values to the specified data located on the chamber’s type plate (left chamber side, bottom right-hand, see chap.
Page 54
Measuring access port Figure 24: Divided inner door, gas proof (example: CB 170) Pull the handles to open each of the individual glass doors. Do not press too hard while closing them to •...
Page 55
(O control) Humidity values (CBF / CBF- Actual values Set-point values Functional icons Figure 25: Normal display of the MB2 program controller (sample values, CBF / CBF-UL with O control) CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 55/197...
Page 56
Setpoint entry Event list Display of active alarms Access to main menu Figure 26: Operating functions of the MB2 controller in normal display (sample values, CBF / CBF-UL with control) CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 56/197...
Page 57
Press the Normal display icon to return from program display and chart recorder display back to Normal display. Normal display (actual values / setpoint values) Program display (example: time program) Chart recorder display (Sample values, CBF / CBF-UL with O control) CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 57/197...
Page 58
Reset alarm Acknowledge the alarm and mute the buzzer. Change between uppercase and lower case characters, digits Change keyboard and special characters Edit Edit settings of time and week programs CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 58/197...
Page 59
“Interior socket” Interior socket (option) activated Information icon for data processing Icon Information Waiting icon: Data processing is running. Remaining time to touch the display when calibrating the touchscreen. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 59/197...
Page 60
(CBF / CBF-UL) is running. The controller offers 5 program memory places with 100 switching points each. The switching points can be distributed over all days of the week. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 60/197...
Page 61
/ cancel an already entered, respectively a running 9.2, 10.1 week program Unless noted otherwise, the figures show the functional range, which is available for the user with “Admin” authorization level. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 61/197...
Page 62
• Available only for users with “Service” and “Admin” authorization level. Full functional range only for BINDER Service (users with “Service” authorization level). “Programs” submenu • Access to the controller’s program functions (chap. 8, 9, 10) CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 62/197...
Page 63
5.5.3 “Service” submenu The “Service” submenu is available for users with “Service” or “Admin” authorization level. When logged-in with “Admin” authorization level the user will find information to tell the BINDER Service in service case. Path: Main menu > Service Serial number of the chamber, setup version chap.
Page 64
• Performance after power failure during sterilization operation The sterilization process is cancelled and the controller changes to Fixed value operation mode with the setpoints already entered. All setpoint values remain in memory. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 64/197...
Page 65
To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one day and in a well-ventilated location. WARNING: If customer should use a BINDER chamber running in non-supervised continuous operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples to split such specimen or samples and store them in at least two chambers, if this is feasible.
Page 66
If the password function has been deactivated, after turning on the unit without user log-in there are those controller functions available, which correspond to the highest authorization level without a password protection. There is no lock icon in the header. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 66/197...
Page 67
Humidity (CB / CB-UL) • With humidity control effected via the Permadry system equilibration time is max. 15 hours ® Humidity (CBF / CBF-UL) • Equilibration time is max. 1 hour. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 67/197...
Page 68
Therefore, it is not necessary to connect a N cylinder. You can deactivate the N pressure alarm in the controller (chap. 7.3). Then also with activated O control no N pressure alarm will be emitted. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 68/197...
Page 69
The controller continues to show the actual humidity value but no longer the humidity set-point. When operating the CBF / CBF-UL with deactivated humidity control, humidification can be realized using the Permadry™ water pan (chap. 6.8). CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 69/197...
Page 70
CB / CB-UL: Humidity control of the Permadry™ system With the CBF / CBF-UL, humidity setting via the Permadry™ System is not required. The Permadry™ system with its 2-pan water system guarantees a maximum humidity of up to 95 % r.H. in the inner chamber, which remains condensation-free.
Page 71
When setting a lower CO or humidity setpoint, the gas or humidity must be able to escape first. Open both chamber doors for this purpose. Observe the precautions when working with gases (chap. 1.8). CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 71/197...
Page 72
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 72/197...
Page 73
Confirm icon. When entering a humidity setpoint value > 90 % r.h. a message window appears, indicating that condensation is possible. Press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 73/197...
Page 74
• Operation line “N2 pressure alarm off” activates / deactivates the N pressure alarm (chamber with O control) • Operation line “Interior socket” activates / deactivates interior socket (option, available via BINDER INDIVIDUAL Customized Solutions) The other operation lines are without function.
Page 75
∅ Do NOT set up chambers in non-ventilated recesses. Ensure technical ventilation measures. Observe the relevant regulations for handling N Close the N supply when decommissioning the chamber. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 75/197...
Page 76
When the configured moment for program start is reached, the program delay time ends and the program starts running. The controller equilibrates to the values which had been active during program start CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 76/197...
Page 77
Press the Confirm icon. After confirming the message the controller changes to Fixed value operation mode. Temperature and humidity will then equilibrate to the setpoints of Fixed value operation mode. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 77/197...
Page 78
The grey bar shows how much time of the whole time is elapsed. If program duration has been set to infinite, the grey bar is not displayed. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 78/197...
Page 79
(CBF / CBF-UL) shall turn off, activate operation line “Idle mode” in the last program section. After confirming the message the controller changes to Fixed value operation mode. The setpoints of Fixed value operation mode are then equilibrated. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 79/197...
Page 80
There are the following options: The program view opens. Select a program section to open the section editor (chap. 9.6) Press the Edit icon to open the program editor CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 80/197...
Page 81
In the program view press the Edit icon to open the program editor In the program editor select “Delete program” and press the Confirm icon. The program is deleted. The controller returns to the program view. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 81/197...
Page 82
• Insert section: Adding the copied section. This menu point is visible only after a section has been copied. • Delete section • Add new section CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 82/197...
Page 83
The current section (example: section 1) is which the copied section shall be inserted… copied. The controller returns to the section view. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 83/197...
Page 84
In the section view press the Edit icon to open the section editor In the section editor select “Delete section” and press the Confirm icon. The section is deleted. The controller returns to the section view. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 84/197...
Page 85
Select the field “Duration” indicating the time. “Duration” entry menu. Enter the desired section duration with the arrow keys and press the Confirm icon. Setting range: 0 up to 99 hours 59 min 59 sec. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 85/197...
Page 86
• Sudden changes of temperature and CO concentration and humidity The entered setpoint of the section is reached as fast as possible and then held constant for the remaining section duration. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 86/197...
Page 88
• Operation line “N2 pressure alarm off” activates / deactivates the N pressure alarm (chamber with O control) • Operation line “Interior socket” activates / deactivates interior socket (option, available via BINDER INDIVIDUAL Customized Solutions) The other operation lines are without function.
Page 89
When requesting rapid value transitions, we recommend not programming tolerance values in order to enable the maximum speed for heating-up and CO control, and humidification. Section view, showing the temperature tolerance band CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 89/197...
Page 90
• Select the field “Start section for repetition” and enter the section number, at which the repetition should start. Setting range: 1 up to the section before the currently selected section. Confirm entry with Confirm icon. The controller returns to the section view. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 90/197...
Page 91
The controller changes to the program view. To save the programming it is absolutely required to press the Confirm icon. Otherwise all settings will be lost! There is no confirmation prompt! CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 91/197...
Page 92
A confirmation prompt is displayed. Press the Confirm icon to confirm that the program shall really be cancelled. After confirming the message the controller changes to Fixed value operation mode. The chamber will then equilibrate to the setpoints of Fixed value operation mode. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 92/197...
Page 93
Press the Confirm icon. The program view opens. Program view. For the first section no weekday is specified. Therefore the section is first marked in red and cannot be saved. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 93/197...
Page 94
• Replace program: Replacing an new or an existing program with the copied program. This menu point is visible only after a section has been copied. • Delete program • Add new section CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 94/197...
Page 95
In the program view press the Edit icon to open the program editor In the program editor select “Delete program” and press the Confirm icon. The program is deleted. The controller returns to the program view. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 95/197...
Page 96
• Insert section: Adding the copied section. This menu point is visible only after a section has been copied. • Delete section • Create new section CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 96/197...
Page 97
(example: section 2). The current section (example: section 1) is copied. Press the Confirm icon The controller returns to the program view. The controller returns to the section editor CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 97/197...
Page 98
(CBF / CBF-UL) for the entire week program. Select the desired program and press the Edit icon to open the program editor. In the program editor select the “Change program name” function and press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 98/197...
Page 99
10.6.3 Start time Section view. Entry menu “Moment”. Select the field “Moment”. Select with the arrow keys the desired start moment of the section and press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 99/197...
Page 100
• Operation line “N2 pressure alarm off” activates / deactivates the N pressure alarm (chamber with O control) • Operation line “Interior socket” activates / deactivates interior socket (option, available via BINDER INDIVIDUAL Customized Solutions) The other operation lines are without function.
Page 101
Chamber with O control: “N2 pressure alarm immediately off” pressure alarm deactivated Interior socket (option) activated “Interior socket” immediately Notifications are not shown in the event list. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 101/197...
Page 102
After configurable Chamber door is open for longer than delay time (chap. “Door open” ---- the set delay time. 11.5). Factory setting: 1 minute Power failure ---- immediately CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 102/197...
Page 103
“setpoints” menu (chap. 6.7) in order to avoid tolerance range and pressure alarms. CBF / CBF-UL: When operating the chamber without water supply, turn off humidity control in the “setpoints” menu (chap. 6.7) in order to avoid humidity alarms.. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 103/197...
Page 104
The alarm is then no longer in the list of active alarms. • The zero-voltage relay alarm output resets together with the alarm. For appropriate actions in the event of an alarm, please refer to chap. 22.6 “Troubleshooting”. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 104/197...
Page 105
(chap. 11.4) there is an audible alert. The zero-voltage relay alarm output (chap. 11.6) switches to transmit the alarm. The alarm is shown in the list of active alarms (chap. 11.3). CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 105/197...
Page 106
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 106/197...
Page 107
As soon as the cause of the alarm is identified and resolved, you can reset the alarm transmission via the zero-voltage relay outputs together with the alarm message in the “Active alarms” menu by pressing the Reset alarm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 107/197...
Page 108
Check the setting regularly and adjust it following changes of the set-point or charge. The safety controller only activates after the set-point has been reached once. During a running sterilization (chap. 21.3) the safety controller is non-functional CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 108/197...
Page 109
• In the field “Mode” select the desired setting “Limit” or “Offset”. • Select the corresponding field “Limit” or “Offset” according to the selected mode and enter the desired safety controller setpoint. Confirm entry with Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 109/197...
Page 110
Check the safety controller at appropriate intervals for its functionality. It is recommended that the authorized operating personnel should perform such a check, e.g., before starting a longer work procedure. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 110/197...
Page 111
If for an authorization level no password is assigned, the related controller functions of this level are available for every user without login. If passwords have been assigned for all authorization levels, access to the controller functions is locked without login. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 111/197...
Page 112
No lock icon is shown in the display header. User login is neither required nor possible. To activate the password protection and user login, perform new password assignment (chap. 13.5.3). CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 112/197...
Page 113
Display when at least some of the authorization levels are password protected and a user has logged in. The user’s authorization (by password entry) is shown. Example: user with “Admin” authorization. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 113/197...
Page 114
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. Controller with deactivated password CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 114/197...
Page 115
13.4 User change If the password function has been deactivated (chap.13.5.2) this function is not available. Path: Main menu > User > User change Controller with logged-in user CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 115/197...
Page 116
Example: A user with “Admin” authorization can change the passwords for the “Admin” and “User” authorization levels. Path: Main menu > User > Password User logged-in with “Service” or “Admin” authorization CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 116/197...
Page 117
To confirm the entry, press the Confirm icon. Repeat the password entry for confirmation (sample picture). For each character of the password, the required keyboard appears automatically. Then press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 117/197...
Page 118
Do NOT enter anything in the “Password” password shall be deleted. screen. Press the Confirm icon. The password is deleted. Do NOT enter anything in the “Confirm password” screen. Press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 118/197...
Page 119
To confirm the entry, press the Confirm icon. Repeat the password entry for confirmation. For each character of the password, the required keyboard appears automatically. Then press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 119/197...
Page 120
Example: Extended configurations available. Under “Expiration date” the date of expiry of the code is displayed. “Activation code” menu with entered code (sample view). Press OK to take over the entry. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 120/197...
Page 121
Return to Normal display with the Back icon to take over the entries. 14.2 Setting date and time Following start-up of the chamber after language selection: Select the time zone and configure the daylight saving time switch. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 121/197...
Page 122
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 122/197...
Page 123
14.4 Display configuration 14.4.1 Adapting the display parameters This function serves to configure parameters like display brightness and operating times. Path: Main menu > Settings > Display > Display “Display” submenu. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 123/197...
Page 124
Normal display. After completing the calibration, i.e., touching all four boxes, the display changes to Normal display. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 124/197...
Page 125
This menu allows to configure the communication parameters of the RS485 interface. The device address is required to recognize chambers with this interface type in a network, e.g. when connecting it to the optional BINDER APT-COM™ 4 Multi Management Software (chap. 18.1). In this case do not change the other parameters.
Page 126
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 126/197...
Page 127
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 127/197...
Page 128
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 128/197...
Page 129
Export configuration and programs Export configuration and timer / time / week programs Import programs Import timer / time / week programs Export service data Export service data Software update Controller firmware update CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 129/197...
Page 130
• Select “Setpoints“ to see information on the entered setpoints and operation lines. • Select “Actual values” to see information on the current actual values. • Select “Safety controller” to see information on the safety controller status. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 130/197...
Page 131
Service Information on modbus analog and digital for BINDER inputs Service Information on Ethernet connection, MAC chap. 14.5.2 address display Back to main menu CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 131/197...
Page 132
Legend shown on the right side of the display 16.1.2 Switch between legend pages Switch legend Press the Switch legend icon to switch between the legend pages Switching between the legend pages page page CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 132/197...
Page 133
Move the central red line by tapping and holding to the desired position. The legend at the right side shows the values of the current line position. Then further icons appear: CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 133/197...
Page 134
History display: Search the required instant Search Press the Search icon to access the “Search” submenu. “Search” submenu. Select the required instant by entering its date and time and press the Confirm icon CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 134/197...
Page 135
Scroll buttons are shown on the left and on the right. Use them to move along the timeline. Press the Change view icon to return from the pen recorder display to Normal Display. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 135/197...
Page 136
After completing the settings, press the Confirm icon to take over the entries and exit the menu, or press the Close icon to exit the menu without taking over the entries. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 136/197...
Page 137
• Water intake temperature NOT below +5 °C / 41 °F and not exceeding 40 °C / 104 °F. BINDER GmbH is NOT responsible for the water quality provided by the customer. Any problems and malfunctions that might arise following use of water of deviating quality is excluded from liability by BINDER GmbH.
Page 138
Options 18.1 APT-COM™ 4 Multi Management Software (option) The chamber is regularly equipped with an Ethernet interface (5) that can connect the BINDER APT-COM™ 4 Multi Management Software. The MAC Address is indicated in the “Device info” controller menu (chap.
Page 139
This quick method of accessing samples is available for both the standard door hinge on the right and the door hinge on the left. Figure 31: Glass door (detail, example door hinge on the right) with Quick sample access option CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 139/197...
Page 140
18.4 Interior socket 230V (option, available via BINDER INDIVIDUAL Customized Solutions) The interior socket is activated and deactivated in the controller via the switching state of an operation line (chap. 7.3, 9.7.3, 10.6.5). When the interior socket is activated, the notification “Interior socket“ is shown highlighted in black next to the information icon in the screen header in Normal display (chap.
Page 141
0 vol.-% up to 20 vol.-% Temperature range: 0 °C / 32°F up to +200 °C / 392°F Figure 35: Pin configuration of the DIN socket (4) on the chamber rear CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 141/197...
Page 142
BINDER Data Loggers are equipped with a keyboard and a large LCD display, alarm functions and a real- time function. Measurement data are recorded in the Data Logger and can be read out after the measurement via the RS232 interface of the Data Logger.
Page 143
The external bottle changer permits automatic switching to a second gas cylinder as soon as the first cylinder is empty. It can be used for a maximum of two CB chambers. Instructions 7001-0196 delivered with the external bottle changer (Art. no. 8012-0408) describe its installation and operation.
Page 144
This could happen e.g. while opening or closing the door, cleaning, charging and discharging the chamber. BINDER offers stable, vibration-free stands with castors (2 lockable by brakes) for the safe stacking of two chambers.
Page 145
NaHCO buffered media. These test systems are not suitable for calibrating the BINDER sensor system. Recommended procedure: Incubate an empty sample with medium for 1/2 day under the same conditions as the cells. You can •...
Page 146
All the necessary equipment must be supplied by one manufacturer only and in a defined test system. Note: These test systems are not very accurate. A typical accuracy is around 10 % of the full-scale value. These test systems are not suitable for calibrating the BINDER sensor system. 19.1.3...
Page 147
Reduce hand germ count (wash hands with antimicrobial soap, dry with paper tissues, and rub dry • hands with alcohol-based solution). Wear appropriate clothing (work coat, shoes, face mask) • Keep as few personnel as possible in the cell culture lab. • CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 147/197...
Page 148
Condensation in the inner chamber represents a particular risk of contamination. The Permadry™ • system developed by BINDER is an effective and easy way to ensure high humidity (≥95 % r.h.) inside the chamber without any condensation forming on the inner surfaces.
Page 149
37 °C / 98.6 °F = ambient temperature 30 °C / 86 °F and less. The chamber should be calibrated / precisely adjusted. • Let BINDER Service adjust the setting of the door heating in reaction to critical ambient conditions. • Regular cleaning, decontamination and sterilization Clean the shelves, glass door, gaskets, and inner chamber weekly (for cleaning see chap.
Page 150
Alcohol-based solutions or the neutral cleaning agent Art. No. 1002-0016. Zinc coated hinge parts Standard commercial cleaning detergents free from acid or halides. rear chamber wall Do NOT use a neutral cleaning agent on zinc coated surfaces. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 150/197...
Page 151
Do not use cleaning agents that may cause a hazard due to reaction with components of the device or the charging material. If there is doubt regarding the suitability of cleaning products, please contact BINDER service. We recommend using the neutral cleaning agent Art. No. 1002-0016 for thorough cleaning.
Page 152
In case of eye contact, the disinfectant spray may cause eye damage due to chemical burns. Follow the operating instructions and safety hints labeled on the bottle of the disinfectant spray. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 152/197...
Page 153
∅ NEVER spray the disinfectant directly on the CO sensor. Wipe the sensor with a lint-free cloth soaked with the disinfectant. Use only the detergents or disinfectants recommended by BINDER. Other products are not approved, since they could in particular damage the CO sensor.
Page 154
Note: in the case when two chambers are directly stacked on top of each other without using the original BINDER stacking stand or the BINDER stacking adapter, the exact maintenance of the incubation set parameters in one of the chambers while carrying out sterilization in the other one cannot be assured.
Page 155
• In the field “Program” select the program “Sterilization”. • Select the field “Program start” and enter the desired program start time in the “Program start” entry menu. Press the Confirm icon. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 155/197...
Page 156
∅ Do NOT touch the inner surfaces, inner doors, inner door handles, and door gaskets during a sterilization. Note: Since the stored sterilization program cannot be changed, it is not accessible via the program path Main menu > Programs > Time program. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 156/197...
Page 157
“Sterilization failed” and when the temperature has cooled down to a value below < 80 °C. When prematurely terminating the sterilization cycle, it may be that the cells/pathogens inside the chamber have not all been killed. If necessary, you should repeat the sterilization. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 157/197...
Page 158
If required, open the outer door. • Cooling-down time: • Front door open: at least 1 hour • Front door closed: at least 4 hours • • If required, open the glass door. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 158/197...
Page 159
Confirm the message with the Confirm icon. control is deactivated and will only be activated after confirming the information message “Successful sterilization” and when the temperature has cooled down to a value below < 80 °C. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 159/197...
Page 160
For personnel requirements please refer to the Service Manual. • Repair Repair of the chamber can be performed by BINDER Service or by BINDER qualified service partners or technicians, in accordance with the description in the Service Manual. After maintenance, the chamber must be tested prior to resuming operation.
Page 161
• Humidification water set (Art. No. 8009-1104, option): 2 m (chap. 4.5.3) • Water container set (Art. No. 8009-1128, option): 10 m (chap. 4.5.4) It is also available by the meter from BINDER. To operate the freshwater supply via tubular bags (standard), a total length of 32 cm is required.
Page 162
When using gas with a technical grade of 99.5 %, we recommend changing the fine gas filter once a year. Please consult BINDER Service. When using gases with less pureness, the changing intervals should be shorter.
Page 163
• Then mount the new sensor proceeding in reverse order • Reinstall the cover on the chamber rear. The sensor may only be used with the attached filter. After replacing the CO sensor repeat the CO calibration. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 163/197...
Page 164
Replace the door gasket only when cold. Otherwise, the door gasket may become damaged. We recommend taking out a maintenance agreement. Please consult BINDER Service. BINDER telephone hotline: +49 (0) 7462 2005 555...
Page 165
If there are is a technical fault or shortcoming, take the chamber out of operation and inform BINDER Service. If you are not sure whether there is a technical fault, proceed according to the following list. If you cannot clearly identify an error or there is a technical fault, please contact BINDER Service.
Page 167
Acknowledge the alarm on the Controller malfunction. controller. If necessary, contact BINDER Service. Pressure sensor system Acknowledge the alarm on the defective. controller. Contact BINDER Service. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 167/197...
Page 168
(more than 10 service. min.) Alarm message Acknowledge the alarm on the “CO range” controller. Contact BINDER Service. sensor defective. To replace the sensor proceed as described in chap. 22.4. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 168/197...
Page 169
Door gaskets defective. Replace door gaskets Gas sensor not adjusted. Calibrate the sensor. Unusually high gas Gas sensor defective. consumption. Contact BINDER Service Gas fine filter not connected correctly. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 169/197...
Page 170
The Permadry™ pan empty. pan must have thorough contact to the bottom of the inner chamber, see chap. 4.2. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 170/197...
Page 171
For rapid transition phases, do NOT Program duration longer than Tolerances have been program tolerance limits in order to programmed. programmed. permit maximum speed for heating, gas control or humidification. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 171/197...
Page 172
22.7 Sending the chamber back to BINDER GmbH If you return a BINDER product to us for repair or any other reason, we will only accept the product upon presentation of an authorization number (RMA number) that has previously been issued to you. An authorization number will be issued after receiving your complaint either in writing or by telephone prior to your sending the BINDER product back to us.
Page 173
Insulating air cushion foil PE foil Plastic recycling Bag for operating manuals PE foil Plastic recycling If recycling is not possible, all packing parts can also be disposed of with normal waste. CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 173/197...
Page 174
(Elektro- und Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl. I p. 1739) or contact BINDER service who will organize taking back and disposal of the chamber according to the German national law for electrical and electronic equipment (Elektro- und Elektronikgerätegesetz, ElektroG from 20 October 2015, BGBl.
Page 175
According to Annex I of Directive 2012/19/EU of the European Parliament and of the Council on waste electrical and electronic equipment (WEEE), BINDER devices are classified as “monitoring and control instruments” (category 9) only intended for professional use“. They must not be disposed of at public collecting points.
Page 176
Danger of violation against existing law if not disposed of properly. Failure to comply with applicable law. Alteration of the environment. For final decommissioning and disposal of the chamber, please contact BINDER service. Follow the statutory regulations for appropriate, environmentally friendly disposal.
Page 177
The fuse holder is equipped with a fuse clip 5mm x 20 mm (CB) or 6.3 x32 mm (CB-UL). Replace this fuse only with a substitute of the same ratings.
Page 179
Humidity data CB / CB-UL humidity range % r.h. 90 to 95 90 to 95 90 to 95 CBF / CBF-UL humidity range % r.h. 50 to 95 50 to 95 CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 179/197...
Page 180
Chamber fuse 4x20mm / 250V / time-lag (T), external Additional temperature fuse class 1 (DIN °C 12880), internal Different electrical data for CB-UL / CBF-UL constructed for the USA and Canada at 50 Hz power frequency 100-120 100-120 100-120 Nominal voltage...
Page 181
Stacking adapter for direct, thermally decoupled stacking Base with castors Stacking adapter for combination C / C-UL 150 on top of CB / CB-UL / CBF / CBF-UL 170 Silicone access ports closable with 2 silicone plugs BINDER Gas Supply Service: External CO...
Page 182
CBF / CBF-UL: Spatial temperature and humidity measurement including certificate 24.6 Accessories and spare parts (extract) BINDER GmbH is responsible for the safety features of the chamber only, provided skilled electricians or qualified personnel authorized by BINDER perform all maintenance and repair, and if components relating to chamber safety are replaced in the event of failure with original spare parts.
Page 184
Spatial temperature and humidity measurement including certificate (27 measuring 8012-0924 points for temperature, 1 measuring point for humidity) For information on components not listed here, please contact BINDER Service. 24.7 Important conversion data for non-SI units 1 ft = 0.305 m 1 m = 100 cm = 3.28 ft = 39.37 inch...
Page 192
Details about utilized substances / biological substances / Einzelheiten über die eingesetzten Substanzen/biologische Materialien: Designations / Bezeichnungen: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ Safety measures required for handling these substances / Vorsichtsmaßnahmen beim Umgang mit diesen Stoffen: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 192/197...
Page 193
Transportweg/Spediteur: Transport by (means and name of transport company, etc.) Versendung durch (Name Spediteur o.ä.) ___________________________________________________________________________________ Date of dispatch to BINDER GmbH / Tag der Absendung an BINDER GmbH ___________________________________________________________________________________ CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 193/197...
Page 194
We hereby commit ourselves and guarantee that we will indemnify BINDER GmbH for all damages that are a consequence of incomplete or incorrect information provided by us, and that we will exempt BINDER GmbH from eventual damage claims by third parties./ Wir versichern, dass wir gegenüber BINDER für jeden...
Page 195
Pictures of chamber attached? Yes No Pictures have to be attached! Yes No Pictures of Packaging attached? Customer Contact Information Distributor Contact Information Name Company Address Phone E-mail CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 195/197...
Page 196
____________________________________________________________________________ ____________________________________________________________________________ Measures to be taken in case of skin contact or release into the atmosphere: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ Other important information that must be considered: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ CB / CB-UL, CBF / CBF-UL (E7) 07/2021 Page 196/197...
Page 197
4.5 Shipping laws and regulations have not been violated. I hereby commit and guarantee that we will indemnify BINDER Inc. for all damages that are a consequence of incomplete or incorrect information provided by us, and that we will indemnify and hold harmless BINDER Inc.
Need help?
Do you have a question about the CB and is the answer not in the manual?
Questions and answers