Ultrasonics calibration

A-61500 May 2010
NOTE: If calibration fails, check the Operator Log for details. See
Chapter 6, "Accessing the Operator Log" for more
information. You must wait at least 90 seconds before
calibrating again.
Use only an A4 size / 75-80 g/m
perform an Ultrasonics calibration.
NOTE: The screens shown in this section are for the TWAIN
Datasource. The screens displayed on your system may be
different.
1. Center the side guides in the elevator tray and output tray.
2. Place the paper in the elevator tray in portrait orientation.
3. Follow Steps 5 - 12 above to access the Calibration dialog box.
4. Select Type: UDDS. Calibration begins. A confirmation box is
displayed when calibration is complete
5. Click OK.
NOTE: If calibration fails, check the Operator Log for details. See
Chapter 6, "Accessing the Operator Log" for more
information.
2
or lettersize / 20 lb. bond paper to
3-15