11.1.3 Determining Max Voltage
If maximal voltage to use isn't known, it is recommended to use the manual mode
to perform a quick saturation test.
•
Start with the maximal voltage of 2000V.
•
Perform the saturation test slowly and stop just before full saturation.
•
The instantaneous voltage will be indicated in the top right corner. In the
example below, the maximal voltage is 320.14V.
•
If a complete test sequence was required in automatic mode, a max voltage of
320.14V + 10% = ~350V would be appropriate.