Icon Process Controls ProCon DO3000-C Series Quick Start Manual page 13

Dissolved oxygen controller

Advertisement

ProCon
— DO3000-C Series
®
Dissolved Oxygen Controller
Menu Description
Measurement
Sensor
Pressure Compensation Enter current atmospheric pressure value
Salinity Compensation
Temperature
25-0671 © Icon Process Controls Ltd.
Type
Select "DO"
Unit
Select unit (mg/L | PPm | %)
Enter salinity level of the sample
Sensor communication address. The address is preset in the factory.
Modbus Node
User can change according to the demand.
The user can compare the instrument's measured temperature with
that of other standard instruments automatically. If a discrepancy is
found, the "temperature offset" can be used for correction. For example,
Offset
if the instrument measures 77°F and the standard instrument mea-
sures 76.64°F, enter "-0.36°F" as the offset. Then, press "Menu" to exit
the editing interface and select "Save" to apply the correction.
You can choose from three temperature methods: "Auto," "Manual," and
"None."
Auto: The instrument automatically displays the temperature detected
by the sensor.
Input
Manual: The user can manually input a temperature value. The system
defaults to a compensation setting of 77°F.
None: The instrument will not display a temperature value.
The user can choose between two temperature units in the unit
Unit
settings: Celsius (
C) or Fahrenheit (ºF).
o
13