Nidek Medical ONE-I Manual
Nidek Medical ONE-I Manual

Nidek Medical ONE-I Manual

Consultation unit

Advertisement

Quick Links

If you would like to order parts,
please feel free to contact us at
our online store
manual
Consultation unit
For advice and inquiries
please do not hesitate to contact
our after-sales service on 01 49 80 97 97
ONE-I

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ONE-I and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Nidek Medical ONE-I

  • Page 1 Consultation unit ONE-I If you would like to order parts, For advice and inquiries please feel free to contact us at please do not hesitate to contact our online store our after-sales service on 01 49 80 97 97 manual...
  • Page 2 Before use or maintenance read this user's manual This manual contains all the information need to use the NIDEK ONE-i consultation unit, such as operating procedures, precautions, features and maintenance instructions. This manual is essential to ensure correct operation of the unit.
  • Page 3: Table Of Contents

    ........................13 Maintenance .........................14 troubleshooting guide .....................14 Maintenance and checks ....................15 Cleaning ..........................15 List of spare parts ......................15 Features and information .....................16 Classification CE .......................16 Electromagnetic compatibility ..................16 Features t echniques ......................19 Dimensions ........................20 TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 4: The One-I Range

    The ONE-i range Consultation unit indications The ONE-i is a consultation unit for refractive examinations. This device eliminates the need for the practitioner and patient to move around, by adapting the position of the ophthalmic instruments to them either manually or electrically. It is intended for use by vision care professionals, including ophthalmologists, optometrists, orthoptists, opticians and nurses.
  • Page 5: Electrical Connections

    Please note that during examinations, movements are controlled under the constant supervision of the TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 6: Serious Incident Reporting

    (WEEE Pro). This means you can dispose of your consultation unit free of charge. For specific information on disposal in countries other than France, please contact your distributor or consult local regulations for the disposal of electronic products. Labeling 2.6.1 Label TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 7: 2.6.2 Nameplates

    ANSM - See website www.ansm.fr Ecosystem: eco-organization approved by the French public authorities. To find out about all collection solutions, visit www.ecosystem.eco or contact NIDEK or its authorized distributor. EEE: Electrical and Electronic Equipment TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 8: Unit Description

    User Manual ONE-I _________________________________________________________________________________ Unit description manual or translation arm Control panel Examination chair with thrust) children and adults alike TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 9: Use

    Economy mode energy After two hours of inactivity, the unit goes into standby mode (the blue LED goes out) to ensure component life. To restart the unit, simply press the panel's on/off button once. TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 10: Using The Control Panel (Tray Up/Down Option)

    CHAIR Automatic ascent Not used INITIALIZATION Service button CHAIR CHAIR PLATEAU Up/Down Advance / Reverse Up/Down 4.4.1 Motorized table top translation (Option) As an option, the ONE-i consultation unit is equiped an electrically-driven tabletop. TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 11: 4.4.2 Memorization Of Preferred Height (Optional) And Mood Lighting 'Ambiance

    - When you switch on the unit by pressing the "Start/Stop" button (see red-circled button) or the initialization button, the tray will automatically return to this height Operating the control panel (option without tray raising/lowering) Phoropter ARM CHAIR Lighting Up/Down TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 12: Using The Motorized Near Vision Test (Option)

    The test continues: straighten the device with the button used to lower the test. Note: The device can be used with the refractor extended. Parinaud test: the Parinaud scale reading test is a test with a text whose lines written in characters of decreasing size. TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 13: Safety Device

    RT up/down safety device: In the event of contact with patient's hand, this part stops the seat rising and the tray lowering. The red "ALERT" LED flashes slowly when safety is activated. TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 14: Maintenance

    Check that the refractor arm is fully extended If so, please contact your authorized distributor If the actions listed in the table above have not solved the problem(s), please contact NIDEK or your authorized distributor. TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 15: Maintenance And Checks

    List of spare parts The ONE-i consulting unit is equipped with automatic reset fuses, so theres no need to change any fuses. This unit has no wear parts to be replaced periodically. No spare parts are required.
  • Page 16: Features And Information

    Rules 1 and 13 of the classification described in Annex VIII of Regulation (EU) N°2017/745 on medical devices define the ONE-i as a Class I medical device. Protection against electric shock: As a Class I device, the ONE-i consulting unit provides protection against electric shock. addition to basic insulation, it features additional safety devices for grounding the conductive parts of accessible fixed wiring.
  • Page 17 IEC 61000-3-3 PST<0.65 (flickers) In conformity, no account has been taken of the uncertainty associated with the result. (1) Compliant, with modifications - see chapter 2.1 (2) Compliant, with modifications - see chapter 2.2 TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 18 In declaring conformity, no account has been taken of the uncertainty associated with the result. Criterion A: no untimely movements Criterion B: to be defined by the manufacturer Criterion C - permissible loss of capability: reset accepted (IEC 61000-4-11) TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 19: Features T Echniques

    CAUTION: If the maximum acceptable weight on the tray is exceeded, there is a risk of damage to the device and to the patient. It is important not to exceed this maximum load. hPa: Hectopascal TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...
  • Page 20: Dimensions

    Design and assembly site Ecopark consultation units 9, rue Benjamin Franklin Ecoparc du Lyonnais 94370 Sucy-en-Brie - France 25 route de Lyon Phone: +33 (0)1 49 80 97 97 69800 Saint-Priest - France Mail: site@nidek.fr Web: www.nidek.fr TAB-DOC-31-V01- 03-02-2025 – User’s manuel ONE-i...

Table of Contents