Summary of Contents for Emerson Rosemount 3051S
Page 1
Quick Start Guide MS-00825-0300-6200, Rev AA March 2025 Rosemount 3051S Submersible Level Transmitter ® with WirelessHART protocol...
Page 2
The Power Module may be replaced in a hazardous area. The Power Module has surface resistivity greater than one gigaohm and must be properly installed in the wireless device enclosure. Care must be taken during transportation to and from the point of installation to prevent electrostatic charge build-up. www.Emerson.com...
This unit was shipped to you without the Power Module installed. Remove the Power Module prior to shipping. Notice This guide provides basic guidelines for Rosemount 3051S Submersible Level Transmitter. It does not provide instructions for diagnostics, maintenance, service, or troubleshooting. This document is also available electronically on Emerson.com/Global.
Page 4
March 2025 Quick Start Guide www.Emerson.com...
Power up sequence The Wireless Gateway should be installed and functioning properly before any wireless field devices are powered. 1. Install the Power Module into the Rosemount 3051S Submersible Level Transmitter to power the device. Note Wireless devices should be powered up in order of proximity from the Gateway, beginning with the closest device, then working outward from the Gateway.
1-ft. (0.3 m) increments to reference during installation. 1.5 Field Communicator connections The Power Module must be installed before the Field Communicator can interface with the Rosemount 3051S Submersible Level Transmitter. For HART Wireless Transmitter communication via a Field ®...
Page 7
Quick Start Guide March 2025 required. To obtain the latest DD, visit the Field Communicator System Software and Device Description site at Emerson.com/ FieldCommunicator. This transmitter uses the SmartPower Solutions Power Module. For ™ more information, see Power Module replacement. Figure 1-3: Connection diagram...
If ordered with the available bracket (Option Code B4), the transmitter should be mounted directly to the bracket using the included U-bolt hardware kit. The bracket can then be mounted to a panel surface or pipe as appropriate. www.Emerson.com...
Quick Start Guide March 2025 Figure 2-1: 3051S Submersible Level Transmitter Note Dimensions are in inches [mm]. 2.3 Remote antenna (optional) The high-gain, remote antenna options provide flexibility for mounting the device based on wireless connectivity, lightning protection, and current work practices. www.Emerson.com...
Page 10
When installing remote mount antennas for the transmitter, always use established safety procedures to avoid falling or making contact with high-power electrical lines. Note Install remote antenna components for the transmitter in compliance with local and national electrical codes and use best practices for lightning protection. www.Emerson.com...
Page 11
If the supplied remote mount antenna kit is not installed per these instructions, Emerson is not responsible for wireless performance or non-compliance with spectrum regulations. Procedure 1.
Page 12
Transmitter. Coaxial sealant must be applied to guarantee performance of the wireless field network. Figure 2-3 for details on how to apply coaxial sealant. Emerson recommends installing the remote antenna kit with a drip loop as shown in Figure 2-3 to limit stress on the antenna connection and water accumulation.
Quick Start Guide March 2025 3 Connect the Power Module Procedure 1. Remove the housing cover on the field terminal side. 2. Connect the Power Module. Figure 3-1: Installing electronics housing cover www.Emerson.com...
Zero trim A zero trim is a single-point adjustment used for compensating mounting position effects. When performing a zero trim, ensure the transmitter is mounted in a final configuration and the sensor has been deployed to the desired location. www.Emerson.com...
Quick Start Guide March 2025 5 Close the housing Procedure 1. Close the housing cover and tighten. 2. Always ensure a proper seal by installing the electronics housing covers so that metal contacts metal. Notice Do not overtighten. www.Emerson.com...
The LCD display will show the output values based on the wireless update rate. For error codes and other LCD display messages, refer to the Rosemount 3051S Wireless Series Reference Manual (00809-0200-4802). Procedure Press and hold the Diagnostic button for at least five seconds.
This page will show whether the device has joined the network and if it is communicating properly. 6.4 AMS Wireless Configurator When the device has joined the network, it will appear in AMS Wireless Configurator. See Figure 6-2. Device Explorer screen Figure 6-2: AMS Wireless Configurator, www.Emerson.com...
Fast Key sequence: Function Fast Key Menu items sequence Communications 3, 4 • Join Status • Wireless ModeJoin Mode • Number of Available Neighbors • Number of Advertisements Heard • Number of Join Attempts www.Emerson.com...
The sensor cable is factory installed with connection directly to the terminal block. Sensor cable wires correspond to the terminals indexed with color codes listed on the terminal block. For connecting with a Field Communicator, refer to Figure 1-3. www.Emerson.com...
• SmartPower ™ Solutions Quick Start Guide (00825-0100-4701) ™ • Emerson A0701PBU Emerson 701P SmartPower Module - Blue Quick Start Guide (MHM-97913-CC-PBF-EN) Notice Care should be taken to prevent thermal, electrical, or mechanical damage. Contacts should be protected to prevent premature discharge.
Changes or modification to the equipment not expressly approved by Emerson could void the user's authority to operate the equipment. L’émetteur/récepteur exempt de licence contenu dans le present appareil est conforme aux CNR d’Innovation, Sciences et Développement économique Canada applicables aux appareils radio...
L’appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d’en compromettre le fonctionnement. Les changements ou les modifications apportés à l'équipement qui n'est pas expressément approuvé par Emerson pourraient annuler l'autorité de l'utilisateur à utiliser cet équipement. 9.5 NCC Notice 取得審驗證明之低功率射頻器材,非經核准,公司、商號或使...
Quick Start Guide March 2025 9.8 Declaration of conformity www.Emerson.com...
Page 24
2025 Emerson. All rights reserved. Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.