EPS Bio Technology EASYMAX Tag User Manual page 54

Self-monitoring blood glucose system

Advertisement

♦ A state of hyperglycemic-hyperosmolar state (with or without ketosis)
2. Lipemic samples: Cholesterol level up to 400 mg/dL (10.32 mmol/L) and triglycerides up to
800 mg/dL (9.04 mmol/L) do not affect the results. Grossly lipemic patient samples have not
been tested and are not recommended for testing with EASYMAX® Tag glucose meters.
3. Critically ill patients should not be tested with EASYMAX® Tag glucose meters.
4. DO NOT use during xylose absorption testing. Xylose in the blood will interfere
Self-Monitoring Blood Glucose System.
5. Interfering Substances depend on the concentration. The below substances up to the test
concentration will not affect the test results.
Concentrations
of the interference tested
Acetaminophen
Ascorbic Acid
Bilirubin-
unconjugated
Cholesterol
Creatinine
Dopamine
Galactose
Gentisic Acid
Glutathione
Bias
7 mg/dL
(0.46 mmol/L)
4 mg/dL
(0.26 mmol/L)
(0.056
3.3 mg/dL
mmol/L)
(10.32
400 mg/dL
mmol/L)
30 mg/dL
( 2.7 mmol/L)
2.2 mg/dL
(0.14 mmol/L)
20 mg/dL
(1.11 mmol/L)
7 mg/dL
(0.45 mmol/L)
1 mg/dL
(0.03 mmol/L)
Glucose
50-100 mg/dL
Level
(2.8-5.6 mmol/L)
8.1
6.6
0.2
9.6
1.3
8.0
6.2
9.8
3.7
250-350 mg/dL
(13.9-19.4 mmol/L)
5.3%
5.8%
5.2%
7.2%
1.6%
3.2%
2.5%
3.6%
6.5%
54