Self-monitoring blood glucose system

Advertisement

Using EASYMAX® Blood Glucose Test Strips
 Use only with EASYMAX® Tag Blood Glucose Meter.
 Run a control solution test every time you open a new box of test strips (See Chapter 2
"Control Solution Testing.")
 Keep the test strips in their original bottle.
 After you take a test strip out, close the bottle immediately and make sure it has been
closed tightly. This keeps the test strips dry.
 Use the test strip within three minutes after you taking it out of the bottle.
 The strip is for single use only. Do not reuse it.
 Write down the date on strip bottle when the strip bottle is first time opened. Be sure to
check the expiration date on the test strip bottle. The test strip will be expired either in
twelve months after the date of the bottle is opened or till the expiration date printed on the
bottle.
 Store the test strip bottle and your meter in a shady and arid place.
 Store the test strips between 2°C~30°C (36°F~86°F). Do not freeze.
 Do not apply blood or control solution to the test strip until you insert it into the meter.
 Do not touch the test strip with wet hands. Do not bend, cut, or twist the test strips.
 EASYMAX® Tag Self-Monitoring Blood Glucose Test System is a "no code" system and
does not require any meter calibration.