EPS Bio Technology EASYMAX Tag User Manual page 44

Self-monitoring blood glucose system

Advertisement

Screen Messages And Troubleshooting
Never make treatment decisions based on an error message. If you have any concerns, please
contact your local dealer or healthcare professional.
Message
What it means?
System error
There may be a problem with
the meter.
Volume detector error
The volume of blood or control
solution is NOT enough.
What to do?
Replace the batteries first.
Refer to pages 15 and 28. If this
error message appears again,
please contact your local dealer.
Repeat the test with a new
strip.
If Err appears again, please
contact your local dealer.