EPS Bio Technology EASYMAX Tag User Manual page 19

Self-monitoring blood glucose system

Advertisement

Chapter 2: Control Solution Testing
Why Run A Control Solution Test
We recommend that you run EASYMAX® Level 2 (Normal) Control Solution Tests because it lets
you know that your meter and test strips are working properly to give reliable results. You should
run the control solution tests when:
 You use the EASYMAX® Tag Blood Glucose Meter for the first time.
 You open a new bottle of test strips.
 You think the meter or test strips may be working incorrectly.
 You drop the meter.
 You have repeated a test and the test results are still lower or higher than expected.
 You are practicing the test procedure.
Professional users are instructed to follow federal, state, and local guidelines.