PARAMOUNT BED PA-90000 Series Instruction Manual

Bed
Hide thumbs Also See for PA-90000 Series:

Advertisement

Quick Links

INSTRUCTION MANUAL
7B01005800A4
PT. PARAMOUNT BED INDONESIA
PA-90000 Series Bed
Thank you for purchasing the PA-90000 Series Bed.
This instruction manual describes precautions and how to safely use and operate
the bed.
●Before using this bed, please read this instruction manual
to fully understand the safe and proper operating procedures.
●The correct operating procedures should be explained not
only to the persons using the bed but also to caregivers.
●After reading this instruction manual, store it where it can be
easily accessed for later reference.
●Users and/or carers are requested to report information about any
serious incident related to this bed that occurs within the EU to the
manufacturer and the regulating authority of the member country of
residence. 'Serious incident' means any incident that directly or
indirectly led, might have led or might lead to any the following :
a) The death of a user, carer or other person.
b) The temporary or permanent serious deterioration of a user's,
carer's or other person's state of health.
c) A serious public health threat.
●The actual product purchased may differ from detailed
descriptions given in this instruction manual due to
product improvements.
●Values in the specification such as working range, dimensions,
angles and mass are subject to normal manufacturing tolerances.
Besides some exemptions, display purpose words such as
"approximately", "about" etc. are omitted.
●For any queries, please feel free to directly contact Paramount Bed
or your distributor.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PA-90000 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for PARAMOUNT BED PA-90000 Series

  • Page 1 Besides some exemptions, display purpose words such as “approximately”, “about” etc. are omitted. ●For any queries, please feel free to directly contact Paramount Bed or your distributor. PT. PARAMOUNT BED INDONESIA...
  • Page 2: Table Of Contents

    CONTENTS Page INTENDED USE/PART NAMES SAFETY PRECAUTIONS 4 - 12 INSTALLATION ASSEMBLING THE BED (PA-90000) 1. Attaching/detaching the head board and foot board 2. Attaching the mattress stopper ASSEMBLING THE BED (PA-93485V) 1. Fitting component for assembly 2. Head board, foot board, side panel installation ASSEMBLING WITH BED EXIT ALARM (PA-90000) 1.
  • Page 3 CONTENTS Page BEDSIDE TABLE OVER BED TABLE PRODUCT SYMBOL DEFINITIONS 53 - 54 REGULAR INSPECTION AND DAILY MAINTENANCE 55 - 57 TROUBLESHOOTING 1. Model(PA-93**5/ PA-93**6) 2. Model(PA-93**N7) 59-63 LONG-TERM STORAGE SPECIFICATIONS 65-75 AFTER-SALES SERVICE...
  • Page 4: Intended Use/Part Names

    INTENDED USE AND PART NAMES 1. Intended use ■This bed is designed to be used for patient care in medical institutions and elderly care facilities. ※Please refer to the 「SPECIFICATION」for the patient target groups and the intended users. Note: Application environment 1 and 2 in IEC 60601-2-52:2009 2.
  • Page 5 【PA-93485V】 The page number shown next to this symbol indicates where the related explanation can be found p.23-24 Handheld Controller Back Section Base p.31 Head board Leg Section Base p.17 p.17 Foot board Power plug p.16 Mattress stopper Hip Section Base Caster Switching Pedal Knee Section Base p.22...
  • Page 6 【PA-90000 With Bed Exir Alarm】 The page number shown next to this symbol indicates where the related explanation can be found p.16 p.25 Head board Handheld Controller Foot board p.16 Air Mattress Power Outlet 2-Handheld Controller Power Outlet L&R Caster Switching Pedal p.18 Caster p.32...
  • Page 7: Safety Precautions

    ●Precautions requiring special attention by persons using the bed are affixed to the foot-end frame. Do not remove or deface the label. ●If the safety label is removed or defaced, replace it with a new one obtained from Paramount Bed or your distributor.
  • Page 8 SAFETY PRECAUTIONS ■ Do not put any body part (especially the head or neck) between a fold-down side rail or board and the raised base (↓). ●When the head or neck enters the gap, the patient and or/user might not be able to withdraw, and there is a risk of injury.
  • Page 9 ■ Do not use with infants. ●They may fall down through the space of a fold- down side rail. ■ Do not allow dust to accumulate on the power plug. ●When dust adheres to the surface of the power plug, the moisture it holds will allow a current to flow, there will be a worsening of the insulation condition, and a risk of fire.
  • Page 10 SAFETY PRECAUTIONS ■ Do not allow operation of the bed by a child aged 12 or under, or someone deemed incapable of comprehending the operation. ●When a child aged 12 or under, or someone deemed incapable of comprehending the operation (such as someone having a cognitive illness) operates the handheld controller by themselves, there is a risk of unexpected injury such as the body becoming caught in the bed by mistake.
  • Page 11 ■ Do not insert the power plug with wet hands. ●There is a risk of electric shock due to short-circuiting. Power plug Grounded outlet ■ Do not spill liquids on the bed. ●Do not spill liquids on the bed's electrical components, such as the motors, handheld controller, or nurse control panel, otherwise, there is a risk of malfunction.
  • Page 12 SAFETY PRECAUTIONS ■ When performing the height adjustment or inclination angle adjustment, pay attention not to hit the wall, beam, or power outlet on the wall. ●When using the bed, make sure that the beds does not hit the wall beam, or power plug during the adjustment of the height or inclination angle.
  • Page 13 ■ Keep away from heaters. ●Do not use heaters near the bed. Otherwise, the bed may be damaged or deformed, or even catch fire. ■ Check around the bed before operating the bed. ●A position operation could damage items locate around the bed, or could cause the bed to be damage or deformed.
  • Page 14 ■ Do not combine the bed with product other than the suitable products specified by Paramount Bed. ●For suitable product items directly mounted on the bed such as fold-down side rails and mattress, use suitable products that are specified by Paramount Bed.
  • Page 15 SAFETY PRECAUTIONS ■ Do not spray a spray type of insectiside directly on the bed. ●Otherwise, there is a risk that the solvents contained in the insecticide could damage, discolor, or dissolve the bed casters or other resin parts. Pay due attention to this, otherwise, there is a risk of unexpected injury due to damaged or dissolve parts.
  • Page 16: Installation

    INSTALLATION ■To avoid possible malfunctions, do not install the bed in the following locations. ●Locations expose to direct sunlight ●Locations near heat sources such as a stove or heater ●Locations which is heavily electrically charged such as near an MRI. ●Locations that has a high oxygen level ●Locations exposed to excessive amounts of water vapor or oil vapor ●Locations that tend to be exposed to high temperatures, high humidity, low temperatures or...
  • Page 17: Assembling The Bed (Pa-90000)

    ASSEMBLING THE BED (PA-90000) 1. Attaching /detaching the headboard and footboard The head board and foot board can be attached or detached. When washing patients hair, detach the head boad. When performing lower limb training or giving a foot bath, detach the foot board. 1.
  • Page 18: Assembling The Bed (Pa-93485V)

    ASSEMBLING THE BED (PA-93485V) 1. Fitting component for assembly 2. Head board, foot board, side panel installation 1. Take side panel from carton box, attach it on both sides of main frame using bolt and nut. 2. Take head board and foot board from carton box, set board bracket to be able to fit with bracket receiver.
  • Page 19: Assembling With Bed Exit Alarm (Pa-90000)

    ASSEMBLING WITH THE BED EXIT ALARM (PA-90000) 1. Attaching and detaching the handheld controller (RC-1850E) A handheld controller can be attached to this bed * For detail on the handheld controller, see p.29 to 33 ● Select a suitable combination of the bed and handheld controller. If the combination is not suitable, the bed could move unexpectedly, resulting injury or bed failure.
  • Page 20: Handheld Controller Hook

    ASSEMBLING WITH THE BED EXIT ALARM (PA-90000) 2. Handheld controller hook The handhleld controller of the type with a handheld controller and the optional handheld controller (RC- 1850E) have a hook. when not usinf the handheld controller, hook it on the top of the hip sectio-end side rail.
  • Page 21: Connection The Bed Communication Cable

    ●Be sure to check the operation after connecting teh bed communication cable. ●For detail on the suitable bed communication cable and suitable nurse call system, contact Paramount Bed (see p.76) or your distributor. Connecting to the nurse call system enables a nurse call to be made at the following times.
  • Page 22: Connecting To The Power Supply (Electric Hospital Bed Only)

    CONNECTING TO THE POWER SUPPLY (ELECTRIC HOSPITAL BED ONLY) 1. Power outlet ●Use a properly wired grounded outlet for each bed Power plug that meets the power consumption requirements of the bed. Also, firmly insert the power plug into the outlet ●...
  • Page 23: Battery (Pa-93**7N With Bed Exit Alarm)

    3. Battery (PA-93**7N With Bed Exit Alarm) The battery is located under back section base at the head end of the bed. * The illustration shows the bed with the head section raised and the headboard and open the back back raise. When replacing the battery, please Attention to its orientation Battery...
  • Page 24 (Battery-powered bed opeation) When the battery power in on, the bed can be operated by power from the battery while the power plug is not inserted into the outlet (or in the event of a power failure) When the remaining battery is sufficient (after recharging for 2 hours), three up-and down ●...
  • Page 25 (Checking the battery status) ■ The battery status is displayed as shown in the table below. Power lamp on the Battery status Power status Battery status handheld controller indicator Lights in green Remaining battery power is sufficient Flashes green Remaining battery power is low. Since (on for 2 seconds →...
  • Page 26 (Recharging the battery) Inserting the power plug of the bed into the outlet start recharging the battery. and recharging will complete automatically after 24 hours. the battery can be recharged regardless of whether the power in on or off. When using the bed for the first time, recharge the battery before use.
  • Page 27: Operation Overview

    OPERATION OVERVIEW Adjusting the angle of the back or knee section base and the height of the bed ■ The angle of the back and knee section base and the height of the bed can be adjusted with no increments. (For operation procedures, see p.27-34) ■...
  • Page 28: Model(Pa-93**5/ Pa-93**6)

    USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) 1. Model (PA-93**5/ PA-93**6) ■ The angle of the back and knee section base and the height of the bed can be smoothly adjusted with no increments by using the handheld controller. ●When a child aged 12 or under, or someone deemed incapable of comprehending the operation (such as someone having a cognitive illness) operates the handheld controller by themselves, there is a...
  • Page 29 USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) Operation buttons ■ Pressing the button adjust the bed, and releasing the button stops the adjusment, allowing you to adjust the bed to the desired position. Press the button ↑ to raise. Press the button ↓ to lower. Back and knee UP and DOWN buttons The back and knee section base angle can be adjusted simultaneously.
  • Page 30: Model(Pa-93**N7)

    USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) 2. (PA-93**7N Series) ( Part names and function) You can adjust the angle of the head and knee sections and the bed height with no increments with the handheld controller. The handheld controller is intended to be operated by patients. However, the function selection buttons (see p.27 to 29) are not intended to be used by patients.
  • Page 31 USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) ➀ Angle/height/error code display …… Displays the angle of the head and knee sections and the bed height during operation. If an error code is displayed, see p.62 Head angle indication : 0 to 70 ( displayed in increments of 1̊ ) Knee angle indication : 0 to 25 ( indicated with a numerical value only for the upper/lower limit of the knee angle, and becomes “—”...
  • Page 32 USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) (Setting switching) ● Do not allow it to be operated by a person deemed incapable of understanding the operation (a child aged 12 or under or a person with dementia). If a person deemed incapable of understanding the operation operates the handheld controller by oneself, the body may get caught in the bed by mistake, resulting in accidental injury.
  • Page 33 USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) ■ Setting each operation of the bed You can change the settings for SOUND. ● Setting the operation sound You can unmute or mute the operation sound that is emitted by pressing the operation button of the handheld controller.
  • Page 34 USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) ■ Setting procedure Press the settings button Press the function selection button for the setting you wan to change. The setting selection lamps is switched Press the setting button. This completes the setting. When the setting is completed, press the settings button on the back of the handheld controller again to confirm the setting.
  • Page 35: Operation Procedures

    OPERATION PROCEDURES 1. Handle operation (3-Crank Manual Hospital Bed) ■ Use the handle mounted on the frame to raise or lower the back or knee section base and adjust the bed height. ① Grip part a of the handle to be operated and pull it out to the end.
  • Page 36: Adjusting The Angle Of Leg Section Base

    2. Adjusting the angle of the leg section base ■ The angle of the leg section base can be manually adjusted in two levels. The angle of the leg section base must be adjusted by at least two persons. ●Normally, the leg section base stay is at position A. Adjustment to the B position should be performed when the knee section base is raised.
  • Page 37: Locking The Casters

    3. Locking the casters ●Lock the casters at all times except when moving the bed. The patient may be injured getting on or off the bed should it move. ■ Central locking system ●The casters can be locked or unlocked by operating switching pedal on the foot end. ●Locking the four wheels The casters are locked by depressing the caster switching pedal.
  • Page 38: Emergency Cpr (Back-Lowering) Function (Electric Hospital Bed Only)

    4. Emergency CPR (back-lowering) function ■ In case of an emergency or a power failure, the raised back section base can be quickly returned to flat status. ●During the emergency CPR operation, make sure that the patient's head, arms, feet, or other body parts have not entered the gap between the back section base and the frame.
  • Page 39: Bed Exit Alarm (Pa-93**7N)

    The bed communication cable may not be suitable depending on the type of nurse call system. For details, contact Paramount Bed (see p. 120) or your distributor. To use BED EXIT ALARM, you need to set the notification function settings.
  • Page 40 WARNING Notification is not made during bed operation (e.g., head up, height adjustment) even if the patient gets out of the bed or sits up on the bed. Furthermore, getting into or out of the bed during bed operation may cause the patient to fall off or fall over, resulting in injury. If you consider that the patient may get in or out of the bed during bed operation, set Overall lockout or Patient lock-out.
  • Page 41 WARNING If the patient does not stay in bed without moving for 5 seconds or more after returning to the bed, detection may not work, notification may not be made, and the patient may fall off or fall over, resulting in injury. If the patient does not stay in bed without moving for 5 seconds or more after bed operation, detection may not work, notification may not be made, and the patient may fall off or fall over, resulting in injury.
  • Page 42 EDGE OF BED When the patient sits down on the edge of the bed and body weight is applied to the floor, notification is made through the nurse call system. To avoid repeated notifications, notification is made at the following times (or when the following conditions are met).
  • Page 43: Mattress (Optional)

    ●Be sure to use a mattress of dimensions specified by Paramount Bed, described below. If an unsuitable mattress is used, patients could get caught between the mattress and side rail and get injured.
  • Page 44: Suitable Products Table

    Nurse call relay unit NU-1660E, NU-1860E *1) These products may not be available in some countries or regions. For details, contact Paramount bed (see p.51) or your distributor. *2) In some countries or regions, some of the medical device or its accessories fall under the category of medical devices.
  • Page 45: Pole

    2. IV pole This IV pole is designed to be attached to the medical device when giving an intravenous drip to a patient. ●Do not move the bed with holding the IV pole. The IV pole may break causing injury. ●Do not use the IV pole as a support to stand up.
  • Page 46: Battery (Pab250T/ Pa-93**6N)

    3. Battery (PAB250T/ PA-93**6N) This battery is intended to provide power to the medical device when the external power supply is not available. ●When connecting the battery cable, unplug the power plug and confirm the patient not on the bed. There is a risk of injury due to accidental operation or electric shock due to a short circuit.
  • Page 47 ○ A life of the battery span about 2 or 3 years. The battery indicator flash or turn off the light even the battery charging for more than 12 hours, please change the battery. ○ When you change the battery, please contact Paramount Bed or your distributor.
  • Page 48 ●Recharging starts as soon as the table power plug in plugged into the power outlet. ●The battery should be charged (for more than 24 hours) immediately after purchase. ●The battery is optional. For detail on the model number, contact Paramount Bed or your distributor. [Battery life] ●The battery is not designed to be recharged after being totally discharged.
  • Page 49: Oxygen Cylinder Holder

    2. Loosen the knob bolt on the oxygen cylinder, and then slowly insert the oxygen cylinder into the holder from the top. 3. Firmly tighten the knob bolt to fasten the oxygen cylinder Oxygen cylinder holder Loosen Tighten Head board ●For any quaries, please feel free to contact Paramount Bed or your distributor...
  • Page 50: Drain Bag Holder

    5. Drain bag holder (PC-5123) This drain bag holder is designed for hanging drain bags. The drain bag holder are located one each on the left and right at the bottom of hip section. (2 location in total). Use and hook drain bags here Drain bag holder Hip section 1.
  • Page 51: Collapsible Bedside Rail

    6. Collapsible bedside rail (PS-0232) This collapsible bedside rail is intended to be fit on bed side to use, prevent patients and bed clothing falling off from bed. Collapsible bedside rail (PS-0232) Releasing lever facing foot end Driver (Large) ① Fixing metal Driver (Large) Product ID label...
  • Page 52: Bedside Table

    BED SIDE TABLE Bed side table is used for having meals or doing something on the bed. Bed side table Model no. Bed side table KF-282 PF-3100A, PF-3110A PF-3120A, PF-3130A ●Remove the table from the bed when performing positioning operations of the bed. Otherwise, the patient may get caught between the table and bed and get injured.
  • Page 53: Over Bed Table

    97cm 58cm 61cm * Use the over bed table and side table within this range Use it not possible because comes into contact with the bed ●For any queries, please feel free to contact Paramount Bed or your distributor.
  • Page 54: Product Symbol Definitions

    PRODUCT SYMBOL DEFINITIONS ■The symbols below are used for the product. Symbol Description Symbol Description ●Take care to prevent the patient's ●Indoor use symbol head from getting caught in the gap between side rails and the bed. ●Take care to prevent neck from ●Electrical waste getting caught between side-rails and board.
  • Page 55 Symbol Description other product used with the bed are suitable products designated Indicates product covered by by Paramount Bed. the EU directive on waste electrical and electronic equipment (WEEE)> ● Make sure that a side-rails used with the bed are suitable...
  • Page 56: Regular Inspection And Daily Maintenance

    REGULAR INSPECTION AND DAILY MAINTENANCE ■Lubrication (if there is abnormal noise) ●Lubrication is needed if there is abnormal noise. Please feel free to contact Paramount Bed or your distributor to be lubricated. ●Water adhesion to the bed or mattress, or moisture may cause corrosion, abnormal noise or contamination.
  • Page 57 Inspect for loose parts, the operation of moving parts on the bed, and the presence of damage regularly. When an irregularity has been noticed or when a detailed inspection is desired, contact Paramount Bed or your distributor. Operation details Inspection items Proper operation can be achieved.
  • Page 58 REGULAR INSPECTION AND DAILY MAINTENANCE ●Do not wash the padding. There is a risk of padding degradation. ●Do not use a mattress washer, an ozone sterilizer or autocrave apparatus. There is a risk of damage, deterioration or degradation. ●Do not leave any blood, urine, sudor or sebum on the mattress. Clean dirt frequently with neutral detergent or soap.
  • Page 59: Troubleshooting

    ●Unlock the board stopper off. locked. ●When requesting repairs or contacting Paramount Bed or your distributor, please provide the number on the product ID label. ■Product lifetime The lifetime of this equipment is typically eight (8) years. Lifetime is defined as the period during which the product will maintain the specified performance and safety, provided it has been maintained and operated in conditions of normal use in accordance with the requirements in this instruction manual.
  • Page 60: Model(Pa-93**N7)

    ■ Problem may not always indicate faults. Check the following once again before requesting repair. ■ If the problem persists after taking corrective action, stop using the bed immediately, disconnect the power plug from the outlet, turn the battery off and then contact Paramount Bed (see p.76) or your distributor for repair.
  • Page 61 For details on the suitable nurse call Are the nurse call system and system and suitable bed communication ̶ bed communication cable cable, contact Paramount Bed (seep.76) suitable for the bed? or your distributor. Notification is not Are there any obstacles between Remove any objects.
  • Page 62 TROUBLESHOOTING Reference Symptoms and status Check Corrective action page Are the notification settings Configure the notification settings configured properly? properly. Is an object of 3 kg or more If an object of 3 kg or more is removed removed from the bed? from the bed, notification may be made.
  • Page 63 Wait for 20 to 30 minutes and then check orange. time intermittently. the operation, if the error code is displayed again, contact Paramount Bed or you distributor. Release the operation button you are holding A bed operation button is held down.
  • Page 64 There is a possibillity that the If the error code is displayed also after control box has falled. turning the power back on, contact Paramount Bed. If the problem persists after taking the corrective action above, contact Paramount Bed (see p.76) or your distributor.
  • Page 65: Long-Term Storage

    (see p.21 ) ●Inspect the product regularly at least once a year, even if it is during storage. Bed disposal When the PA-90000 series bed has reached the end of its usable life, discard it according to local standards.
  • Page 66: Specifications

    285VA max. Duty cycle 2min. (stop time: 18min (*3) 8 years (based on Paramount Bed data), excluding consumable items such as Service life caster *1) The mattress base height indicates the height from the floor to the mattress base surface.
  • Page 67 285VA max. Duty cycle 2min. (stop time: 18min (*3) 8 years (based on Paramount Bed data), excluding consumable items such as Service life caster *1) The mattress base height indicates the height from the floor to the mattress base surface.
  • Page 68 Materials controller clip: POM resin molding, Cable: PVC resin molding 8 years (based on Paramount Bed data), excluding consumable items such as Service life caster *1) The mattress base height indicates the height from the floor to the mattress base surface.
  • Page 69 285VA max. Duty cycle 2min. (stop time: 18min (*3) 8 years (based on Paramount Bed data), excluding consumable items such as Service life caster *1) The mattress base height indicates the height from the floor to the mattress base surface.
  • Page 70 5 to 40°C conditions Humidity 20 to 90%RH 8 years (based on Paramount Bed data), excluding consumable items such as Service life caster *1) The mattress base height indicates the height from the floor to the mattress base surface. *2) Safe working load: Maximum load with which the bed can be operated. (The total load of user's weight...
  • Page 71 5 to 40°C conditions Humidity 20 to 90%RH 8 years (based on Paramount Bed data), excluding consumable items such as Service life caster *1) The mattress base height indicates the height from the floor to the mattress base surface. *2) Safe working load: Maximum load with which the bed can be operated. (The total load of user's weight...
  • Page 72 SPECIFICATIONS Model PA-93**5/ PA-93**6 Range Environmental conditions When transported or stored When in use Temperature -10 to 50℃ 5 to 40℃ Humidity 20 to 90%RH 20 to 90%RH Air pressure 700 to 1,060hPa 700 to 1,060hPa Model PA-93**7N Range Environmental conditions When transported or stored When in use 10 to 40°C (when using BED...
  • Page 73 Without shield Guidance and Manufacturer's Declaration - Electromagnetic Emissions The PA-90000 Series is intended for use in the following electromagnetic environment. The customer or user of the PA-90000 Series should make sure it is used in such an environment. Emission test...
  • Page 74: Specifications

    SPECIFICATIONS Guidance and Manufacturer's Declaration - Electromagnetic Immunity The PA-90000 Series is intended for use within healthcare facility environment. Users or customers must use bed in this environment. IEC 60601-1-2/EN 60601-1-2 Immunity test Compliance level Test level Static electricity discharge (ESD) Contact discharge: ±8kV...
  • Page 75 Guidance and Manufacturer's Declaration - Electromagnetic Immunity The PA-90000 Series is intended for use in an electromagnetic environment that used wireless RF transciever. Electromagnetic interference can be prevented by maintaining the minimum distance between RF transciever and PA- 90000 series depend on the maximum output and frequency of the communication device as recommended below.
  • Page 76 SPECIFICATIONS Guidance and Manufacturer's Declaration - Electromagnetic Immunity The PA-90000 Series is intended for use in the following electromagnetic environment. The customer or user of the PA- 90000 Series should make sure it is used in such an environment. IEC 60601-1-2...
  • Page 77: After-Sales Service

    ◼ Consumable parts Casters, handheld controller are consumable parts. 2.Minimum stock availability period Paramount Bed keeps replacement parts (necessary to maintain bed performance) in stock for at least 8 years after the discontinuation of production. 3.Questions regarding after-sales service For any queries, please feel free to contact Paramount Bed or your distributor.
  • Page 78 NOTES...
  • Page 79 NOTES...
  • Page 80 NOTES...
  • Page 81 NOTES...
  • Page 82 NOTES...
  • Page 83 NOTES...
  • Page 84 2024-07...