INSTRUCTION MANUAL
7A10625900A4
PARAMOUNT BED CO, LTD.
PA-90000 Series Bed
Preface
Thank you for purchasing the PA-90000 Series Bed.
This instruction manual describes precautions and how to safely use
and operate the bed.
Before using this bed, please read this instruction manual to fully
understand the safe and proper operating procedures.
The correct operating procedures should be explained not only to
the persons using the bed but also to caregivers.
After reading this instruction manual, store it where it can be easily
accessed for later reference.
Users and/or carers are requested to report information about any serious
incident related to this bed that occurs within the EU to the manufacturer
and the regulating authority of the member country of residence.'Serious
incident' means any incident that directly or indirectly led, might have led
or might lead to any of the following:
a) The death of a user, carer or other person.
b) The temporary or permanent serious deterioration of a user's, carer's or
other person's state of health.
c) A serious public health threat.
The actual product purchased may differ from detailed descriptions
given in this instruction manual due to product improvements.
Value in the specification such as working range, dimensions, angles
and mass are subject to normal manufacturing tolerances. Besides some
exemptions, display purpose words such as "approximately", "about"
etc. are omitted.
For any queries, please feel free to contact your dealer directly.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PA-90000 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for PARAMOUNT BED PA-90000 Series

  • Page 1 INSTRUCTION MANUAL PA-90000 Series Bed 7A10625900A4 Preface Thank you for purchasing the PA-90000 Series Bed. This instruction manual describes precautions and how to safely use and operate the bed. ● Before using this bed, please read this instruction manual to fully understand the safe and proper operating procedures.
  • Page 2: Table Of Contents

    CONTENTS Page INTENDED USE / PART NAMES SAFETY PRECAUTIONS 4-12 SPECIFICATIONS 13-16 INSTALLING THE BED ASSEMBLING THE BED (PA-90000) 1. Attaching/detaching the head board and foot board 2. Attaching the mattress stopper ASSEMBLING THE BED (PA-90000VIP) 1. Fitting component for assembly 2.
  • Page 3: Intended Use / Part Names

    INTENDED USE / PART NAMES 1. INTENDED USE ※  Please refer to the 「SPECIFICATION] for the patient target groups and the intended users.        2. PART NAMES 【PA-90000】 The page number shown next to this Bed with The Central Locking System symbol indicates where the related p.26-27 Handheld Controller...
  • Page 4 【PA-90000VIP】 The page number shown next to this        symbol indicates where the related p.23 Head board p.26-27 Handheld controller p.29 Leg section base Back section base p.23 Foot board p.22 Power plug Matress stopper Hip section base p.23 Knee section base Side panel p.30 Caster...
  • Page 5: Safety Precautions

    SAFETY PRECAUTIONS Be sure to read these safety precautions beforehand to ensure proper usage. The safety precautions described here are intended to ensure safe use of the product and prevent personal injury. Precautions are classified as “WARNING”, “CAUTION” to indicate the degree of hazard or injury that may result from improper use.
  • Page 6 SAFETY PRECAUTIONS WARNING Do not put any body part (especially the head or neck) between a fold-down side rail or board and the raised base ( ). When the head or neck enters the gap, the patient and/or user might not able to withdraw, and there is a risk of injury.
  • Page 7 SAFETY PRECAUTIONS WARNING Do not use with infants. They may fall down through the space of a fold-down side rail. Do not allow dust to accumulate on the power plug. When dust adheres to the surface of the power plug, the moisture it holds will allow a current to flow, there will be a worsening of the insulation condition, and a risk of fire.
  • Page 8 WARNING Do not allow operation of the bed by a child aged 12 or under, or someone deemed incapable of comprehending the operation. When a child aged 12 or under, or somenone deemed incapable of comprehending the operation (such as someone having a cognitive illness) operates the handheld controller by themselves, there is a risk of unexpected injury such as the body becoming caught in the bed by mistake.
  • Page 9 SAFETY PRECAUTIONS WARNING Do not insert the power plug with wet hands. There is a risk of electric shock due to short-circuiting. Electric shock Do not spill liquids on the bed. Do not spill liquids on the bed’s electrical components, such as the motors, handheld controller, or nurse control panel, otherwise, there is a risk of malfunction.
  • Page 10 CAUTION the wall, beam, or power outlet on the wall. When using the bed, make sure that the beds does not hit the wall beam, or power plug during the Make sure that there is at least 20 cm clearances from the wall or beam.
  • Page 11 SAFETY PRECAUTIONS CAUTION Keep away from heaters. Do not use heaters near the bed. Otherwise, the bed may be damaged or deformed, or even catch fire. around the bed, or could cause the bed to be damage or deformed. Be sure to use the stopper for the headboard and footboard. Lock If the board is not mounted properly, the board may suddenly come off...
  • Page 12 SAFETY PRECAUTIONS CAUTION Hold the power plug body when unplugging. When holding onto only the power cable and pulling it to disconnect the plug from outlet, the cable could be damaged and there is a risk of electrical shock. Always unplug the bed before cleaning. electrical shock due to a short circuit.
  • Page 13 SAFETY PRECAUTIONS CAUTION Do not spray a spray tipe of insecticide directly on the bed. Otherwise, there is a risk that the solvents contained in the insecticide could damage, discolor, or dissolve the bed casters or other resin parts. Pay due attention to this; othewise, there is a risk of unexpected injury due to damaged or dissolved parts.
  • Page 14: Specifications

    Height Adjustment 220VA max. Duty cycle 8 years (based on Paramount Bed data), excluding consumable items such Service life as caster *3) Safe working load: Maximum load with which the bed can be operated. (The total load of user’s weight and...
  • Page 15 5 to 40 Humidity 20 to 90%RH 8 years (based on Paramount Bed data), excluding consumable items such Service life as caster *3) Safe working load: Maximum load with which the bed can be operated. (The total load of user’s weight and...
  • Page 16 220VA max. Duty cycle 2min. (stop time: 18min (*3)) 8 years (based on Paramount Bed data), excluding consumable items such Service life as caster *1) The mattress base height indicates the height from the floor to the mattress base surface.
  • Page 17 SPECIFICATIONS SAFETY PRECAUTIONS Range When transported or stored When in use Temperature -10 to 50 5 to 40 Humidity 20 to 90%RH 20 to 90%RH Air pressure 700 to 1,060hPa 700 to 1,060hPa User weight (min-max) 40kg to 138kg Minimum user height 146cm Minimum user body Patient...
  • Page 18 Without shield Guidance and Manufacturer's Declaration - Electromagnetic Emissions The PA-90000 Series is intended for use in the following electromagnetic environment. The customer or user of the PA-90000 Series should make sure it is used in such an environment. Emission test...
  • Page 19 SAFETY PRECAUTIONS Guidance and Manufacturer's Declaration - Electromagnetic Immunity The PA-90000 Series is intended for use within healthcare facility environment. Users or customers must use bed in this environment IEC 60601-1-2/EN 60601-1-2 Compliance level Immunity test Test level Static electricity...
  • Page 20 SAFETY PRECAUTIONS Guidance and Manufacturer's Declaration - Electromagnetic Immunity The PA-90000 Series is intended for use in an electromagnetic environment that used wireless RF transciever. Electrromagnetic interference can be prevented by maintaining the minimum distance between RF transciever and PA-90000 Series depend on the maximum output and frequency of the communication device as recommended below.
  • Page 21 SAFETY PRECAUTIONS Guidance and Manufacturer's Declaration - Electromagnetic Immunity The PA-90000 Series is intended for use in the following electromagnetic environment. The customer or user of the PA-90000 Series should make sure it is used in such an environment. IEC 60601-1-2...
  • Page 22: Installing The Bed

    INSTALLING THE BED Area exposed to direct sunlight Near a heat source (stove, heater, etc.) Area exposed to excessive water vapor content or oil vapor Area which is extremely hot, cold, humid, or dry ※ Temperature 5 to 40 C/ Humidity 20 to 90%RH/ Atmospheric pressure 700hpa to 1060hPa. Area with a large amount of dust, smoke, salt, sulfur, or corrosive substances Area where the floor is not strong enough Area where the floor is not level...
  • Page 23: Assembling The Bed (Pa-90000)

    ASSEMBLING THE BED (PA-90000) a foot bath, detach the footboard. Grip Board and mount it so that the board bracket engage with Board bracket sure that it is locked. Detaching Move the board stopper to unlock. Grasp the grip of the board and detach it. Board stopper Lock Unlock...
  • Page 24: Assembling The Bed (Pa-90000Vip)

    ASSEMBLING THE BED (PA-90000VIP) 1. Fitting component for assembly BOLT RACHET WRENCH SPANNER 6 pcs 6 pcs 13 mm 13 mm 2. Head board, foot board, side panel installation Head board Bolt Board bracket Side panel Bracket receiver Foot board Side panel Main frame 1.
  • Page 25: Connecting To The Power Supply (Electric Hospital Bed Only)

    CONNECTING TO THE POWER SUPPLY (ELECTRIC HOSPITAL BED ONLY) 1. Power outlet Each bed should be plugged into a separate, Power plug appropriately wired, single-phase AC 220 V, 10 A or greater grounded power outlet using the power plug Grounded power outlet Never connect more than one cable to a power outlet CAUTION...
  • Page 26: Operation Overview

    OPERATION OVERVIEW 26-28) Electric Hospital Bed Back raising adjusted from 0 to approximately 70 Manual Hospital Bed adjusted from 0 to approximately 75 Electric Hospital Bed Knee raising adjusted from 0 to approximately 25 Manual Hospital Bed adjusted from 0 to approximately 45 Height adjusment PA-93485, PA-93385...
  • Page 27: Using The Handheld Controller (Electric Hospital Bed Only)

    USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) with no increments by using the handheld controller. When a child aged 12 or under, or someone deemed incapable WARNING caught in the bed by mistake. If there is a risk of such person touching the handheld controller, disconnect the power plug in order to prevent an accident caused allowing you to adjust the bed to the...
  • Page 28 USING THE HANDHELD CONTROLLER (ELECTRIC HOSPITAL BED ONLY) can be adjusted simultaneously. adjusted. adjusted. The bed height can be adjusted. Do not drop the handheld controller or pull the cable. The handheld controller may get damaged or deformed and operate incorrectly as a result. When the handheld controller is damaged, request a repair (or replacement).
  • Page 29: Operation Procedures

    OPERATION PROCEDURES 2 10 the bed height. Grip part a of the handle to be operated and pull it out to the end. Raise the grip to 90 degrees, hold the grip, and turn the handle. Up Down Up Down Up Down Knee Height Back...
  • Page 30: Adjusting The Angle Of Leg Section Base

    base must be adjusted by at least two persons. Main frame and main frame and get injured. CAUTION...
  • Page 31: Locking The Casters

    3. Locking the casters WARNING Central locking system switching pedal on the foot end Locking the four wheels The casters are locked by depressing the caster switching pedal. Unlocking the four wheels The casters are unlocked by raising the caster switching pedal.
  • Page 32: Emergency Cpr (Back-Lowering) Function (Electric Hospital Bed Only)

    status. head, arms, feet, or other body parts have not entered the gap WARNING <Emergency CPR procedure> and pull the emergency CPR lever (orange). Levers base are located on both sides of the bed under the hip returns to flat status. If you release the emergency CPR lever halfway through, the back-lowering Emergency CPR lever (orange) Power on status: Back-raising and back lowering can be...
  • Page 33: Mattress (Optional)

    Model Codes : PE-9001/ PE-9101/ PE-9011/ PE-9111/ PE-9002/ PE-9102/ PE-9012 and PE-9112 RPR : Model Codes : PE-7001/ PE-7002/ PE-7011 and PE-7012 PE-6000 Series : Model Codes : PE-6001/ PE-6101/ PE-6002/ PE-6102/ PE-6011/ PE-6111/ PE-6012/ and PE-6112 Paramount Bed or your distributor.
  • Page 34: Optional Accessories

    KC-56B (with 4 hooks) EC-562 Traction equipment KA-83 Oxygen cylinder holder PC-5038 PAB250T Battery ● Recommend accessories may differ due to specification changes or according to product combinations. For any queries, please feel free to contact Paramount Bed or your distributor.
  • Page 35: Pole

    2. IV Pole This IV pole is designed to be attached to the medical device when giving an intravenous drip to a patient. Do not move the bed with holding the IV pole. The IV pole may break causing injury. Do not use the IV pole as a support to stand up.
  • Page 36 3. Battery (PAB250T/ Electric hospital bed only) This battery is intended to provide power to the medical device when the external power supply is not available. WARNING ※ Bolt Side frame If installing with incorrect bracket is incorrect. ※...
  • Page 37 Slide the control box cover Set minus driver in the hole of side on the control box and slide the control box cover to open it. control box. Hole Control box...
  • Page 38: Oxygen Cylinder Holder

    2. Loosen the knob bolt on the oxygen cylinder holder, and then slowly insert the oxygen cylinder into the holder from the top. Oxygen cylinder holder Tighten Loosen Knob bolt Head board For any quaries, please feel free to contact Paramount Bed or your distributor.
  • Page 39: Bed Side Table

    BED SIDE TABLE Bed side table is used for having meals or doing something on the bed. Bed side table Suitable bed side table Bed side table KF-282 PF-3100A, PF-3110A PF-3120A, PF-3130A WARNING between the table and bed and get injured.
  • Page 40: Over Bed Table

    58 cm 97 cm 61 cm *Use the over bed table and side table within this range Use it not possible because comes into contact with the bed For any queries, please feel free to contact Paramount Bed or your distributor.
  • Page 41: Product Symbol Definitions

    PRODUCT SYMBOL DEFINITIONS The symbols below are used for this product. Symbol Symbol CE mark Manufacture Date of manufacture Indoor use symbol Safe working load Electrical waste Sum of the product weight AC input and safe working load IPX4 Serial number Model number in the european community Country of manufacturing...
  • Page 42: Regular Inspection And Daily Maintenance

    REGULAR INSPECTION AND DAILY MAINTENANCE lubricated. <Bed unit> To clean the bed, use a cloth that has been soaked in diluted detergent and wrung out well, then wipe off any residual detergent with a cloth soaked in water and wrung out well. Finally, use a dry cloth to wipe the bed to ensure it is dry.
  • Page 43 For safety in use, inspect the product regularly, at least once a year. The product exhibits wear and There must not be any strange noises. There must not be any strange noises. (p.25) There must not be any strange noises. Bed unit Miscellaneous There must not be any peeling of labels, and the...
  • Page 44 WARNING Remove any soiling (blood, excreta, etc.) detergent diluted with water, and wrung out well. 0.05% - 0.2% benzalkonium chloride (e.g. Osvan) 0.05% - 0.2% benzethonium chloride (e.g. Hiamine) 0.05% chlorhexidine gluconate (e.g. Hibitane) 0.1% - 1.0% sodium hypochlorite (e.g. Purelox) in the sun.
  • Page 45: Troubleshooting

    Paramount Bed or your distributor for repair. Bed damaged in earthquakes, fires, floods or other incidents must be inspected and repaired by Paramount Bed or your distributor. Otherwise, there is a risk of electric shock or fire due to short circuit Problem...
  • Page 46 Product ID label are affixed to the foot end of the main frame and on the inside of the headboard. PA592BOA _ _ _ _ _ _ 007B 1102598 00001 (Label for Electric Type Bed) (Label for Manual Type Bed) the product will maintain the specified performance and safety, provided it has been maintained and...
  • Page 47: Long-Term Storage

    (See p.24). Inspect the product regularly at least once a year, even if it is during storage. Bed Disposal When the PA-90000 Series bed has reached the end of its usable life, discard it according to local standards.
  • Page 48: After-Sales Service

    2. Minimum stock availability period Paramount Bed keeps replacement parts (necessary to maintain bed performance) in stock for at least 3. Question regarding after-sales service For any queries, please feel free to contact your dealer or to the following addresses.
  • Page 49 NOTES...
  • Page 50 NOTES...
  • Page 51 NOTES...