Advertisement
Read "Safety Information" before you use the Meter.
The Fluke 381 is a handheld, battery-operated Clamp Meter (the Meter) that has a remote-display module and detachable iFlex (Flexible Current Probe). The Remote Display can be removed from the Meter body and read away from the measurement source. This lets the display be easily read in difficult-measurement situations such as a hazardous environments, or very tight spaces. The Flexible Current Probe makes it possible to measure higher current (up to 2500 A ac) and larger cables that traditional jawed meters cannot measure.
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
To avoid possible damage to the Meter or to equipment under test:
Table 1. Symbols
Symbol | Meaning | Symbol | Meaning |
~ | AC (Alternating Current) | ![]() | Earth ground |
![]() | DC (Direct Current) | ![]() | AC and dc current. |
| Hazardous voltage | | Risk of Danger. Important information. See Manual. |
![]() | Battery. Low battery when shown. | ![]() | Double insulated |
CAT III | IEC Measurement Category III CAT III equipment has protection against transients in equipment in fixed-equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings. | CAT IV | IEC Measurement Category IV CAT IV equipment has protection against transients from the primary supply level, such as an electricity Meter or an overhead or underground utility service. |
![]() | Do not apply to or remove from HAZARDOUS LIVE conductors. | ![]() | Application around and removal from HAZARDOUS LIVE conductors is permitted. |
Note
The Measurement Category (CAT) and voltage rating of any combination of test probe, test probe accessory, current clamp accessory, and the Meter is the LOWEST rating of any individual component.
The following sections explain the Meter features in detail. See Figure 2 and Table 2.
See Figure 2 and Table 2
Table 2. Meter Features
Item | Description |
![]() | Current sensing Jaw |
![]() | Tactile Barrier |
![]() | Rotary Function Switch, see Table 3. |
![]() | Hazardous-voltage indicator |
![]() | Display release button |
![]() | Display |
![]() | Backlight button: turns the Backlight on and off. The Backlight stays on for 2 minutes when there is no button or switch interaction and then shuts off. |
![]() | Hold button: freezes the display reading and releases the reading when pushed a second time. |
![]() | Min Max button: when first pushed, the Meter shows maximum input. With subsequent pushes, the minimum and the average inputs are shown. Hold ![]() |
![]() | Zero/Shift button: removes dc offset from dc current measurements. Also used to shift and corresponds to the yellow items on the Rotary Function Switch. |
![]() | Inrush button: push to enter inrush mode. Push a second time to exit inrush mode. Integration time is 100 ms. |
![]() | Jaw release |
![]() | Alignment marks: to meet accuracy specifications, the conductor must be aligned with these marks. |
![]() | Common terminal |
![]() | Volts/Ohm input terminal |
![]() | Flexible Current Probe input terminal |
Table 3. Rotary Function Switch
Switch Position | Function |
OFF | Meter is powered down |
![]() | AC voltage |
![]() | DC voltage |
![]() | Resistance and continuity |
![]() | AC current. Push ![]() |
![]() | DC current |
![]() | AC current and frequency measurement using the Flexible Current Probe. Push ![]() |
The Meter uses low-power 802.15.4 wireless technology to let the display module operate in a different location than the Meter base. Although there is control of some Meter functions (Hold, MIN MAX AVG, and Backlight), complete remote control of the Meter is not available through the display module.
The wireless radio signal does not hinder Meter measurements. Usually, the radio signal is off when the display module is docked to the Meter base. It is possible for the radio signal to be on when the display module is docked and the Rotary Function Switch is set to OFF. To make sure that the radio signal is off, remove the batteries from the Meter base and display module.
The display module is synchronized with a Meter base when it is docked on the Meter base and turned on. Different display modules can be synchronized with a Meter base but, only one display module can be synchronized to a Meter base at the same time.
The Meter base and display can be a maximum of 10 meters from each other before the radio signal connection is broken. This distance can change with the obstacles between the Meter base and display. There is a radio connection when shows in the display.
To detach the display from the Meter base, see Figure 1.
When the Meter senses a voltage ±30 V or a voltage overload (OL), is shown on the display and the red high-voltage LED (
) on the Meter base illuminates to tell you a hazardous voltage is at the Meter input.
To avoid electrical shock, do not apply or remove from live hazardous conductors.
The high-performance AC Flexible Current Probe utilizes the Rogowski principle and is used for accurate, non-intrusive measurement of sinusoidal, pulsed, and other complex waveforms. The flexible and lightweight measuring head allows quick and easy installation in hard-to-reach areas and works well with large conductors.
For more information about the Flexible Current Probe, see "Current Measurement (Flexible Current Probe)".
The Meter powers off if there is no button push or Rotary Function Switch operation for 20 minutes. If the Meter powers off, turn the Rotary Function Switch OFF and then back on again. Auto Power Off is disabled during Min Max Avg function use. To disable the Auto Power Off, hold down while turning on the Meter.
Push to toggle the Backlight on and off. The Backlight automatically goes off after 2 minutes. To disable the Backlight Auto Off feature, hold down
while turning on the Meter.
To capture and hold the present display reading, push while taking a reading. Push
again to return to the live reading.
Min Max Avg mode can capture the minimum, maximum, and average readings of a given output signal over an extended time.
Push to enter Min Max Avg mode, push again to toggle between min and max readings. Push a third time to display the average reading. To exit Min Max Avg mode, push and hold
for 2 seconds. When Min Max Avg mode is active, the Auto Power Off feature is disabled.
Push to remove any dc offset that could affect the accuracy of dc readings.
Inrush Current is surge current that occurs when an electrical device is first powered on. The Meter can capture this surge current reading. Current spikes from motor drives are one example of such an event. The Inrush function takes approximately 400 samples over a 100 ms period and calculates the starting current envelope.
The Meter uses two low battery symbols: meter and remote
. When meter
appears, the batteries in the Meter base should be changed. Low batteries on the Meter base will affect the readings. When remote
is displayed, the batteries for the removable display should be changed. Measurements are not affected by low batteries in the display.
To view all segments on the display at once, push while turning the Meter on. See Figure 3 and Table 4.
Note
Prior to first use, remove the battery isolator (small piece of plastic between the batteries and battery contacts).
To avoid electric shock or personal injury:
Note
When measuring current, center the conductor in the Jaw using the alignment marks on the Jaw.
Before taking dc measurements, push to ensure correct readings. Zeroing the Meter removes dc offset from the reading. The Zero function works only in the dc current measurement Rotary Function Switch position.
Note
Before zeroing the Meter, make sure the Jaws are closed and there is no conductor inside the Jaw.
To measure ac or dc current:
Note
Current flowing in opposite directions cancels each other. If current is moving in opposite directions, place one conductor into the clamp at a time. See Figure 4.
To prevent possible electrical shock or personal injury:
Do not apply the Flexible Current Probe around or remove from HAZARDOUS LIVE conductors. Take special care during fitting and removal of the Flexible Current Probe. De-energize the installation under test or wear suitable protective clothing.
To use the Flexible Current Probe, follow these instructions:
If the Flexible Current Probe does not perform as expected:
To measure ac or dc voltage:
To measure resistance or continuity:
If the resistance is < 30 Ω, continuity is indicated by a beeper continuously sounding. If the display reads OL, the circuit is open.
(Jaw and Flexible Current Probe)
The Meter can capture the initial inrush current when starting a device such as a motor or light ballast. To measure the inrush current:
(Jaw and Flexible Current Probe)
To measure frequency:
To avoid possible electric shock or personal injury, repairs or servicing not covered in this manual should be performed only by qualified personnel.
To avoid electrical shock, remove any input signals before cleaning.
To avoid damaging the Meter, do not use aromatic hydrocarbons or chlorinated solvents for cleaning. These solutions will react with the plastics used in the Meter. Do not immerse the Meter in water.
Clean the instrument case with a damp cloth and mild detergent.
To replace the batteries in the Meter body, see Figure 8:
To replace the batteries in the display module, see Figure 8:
Table 5. User-Replaceable Parts
Description | Qty. | Fluke Part Number |
Battery, AAA 1.5 V | 5 | 2838018 |
Battery Door - Display Module | 1 | 3625529 |
Battery Door - Meter Base | 1 | 3766406 |
Fluke 381 Remote Display | 1 | 3766445 |
Soft Case | 1 | 3752973 |
User Manual | 1 | 3538357 |
Electrical Specifications | |
AC Current Via Jaw | |
Range | 999.9 A |
Resolution | 0.1 A |
Accuracy | 2% ± 5 digits (10-100 Hz) 5% ± 5 digits (100-500 Hz) |
Crest Factor (50/60 Hz) | 3 @ 500 A 2.5 @ 600 A 1.42 @1000 A Add 2% for C.F. > 2 |
AC Current via Flexible Current Probe | |
Range | 999.9 A / 2500 A (45 Hz – 500 Hz) |
Resolution | 0.1 A / 1 A |
Accuracy | 3% ±5 digits |
Crest Factor (50/60Hz) | 3.0 at 1100 A 2.5 at 1400 A 1.42 at 2500 A Add 2% for C.F. > 2 |
Position Sensitivity (See Figure 9)
Distance from Optimum | i2500-10 Flex | i2500-18 Flex | Error |
A | 0.5 in (12.7 mm) | 1.4 in (35.6 mm) | ± 0.5 % |
B | 0.8 in (20.3 mm) | 2.0 in (50.8 mm) | ± 1.0 % |
C | 1.4 in (35.6 mm) | 2.5 in (63.5 mm) | ± 2.0 % |
Measurement uncertainty assumes centralized primary conductor at optimum position, no external electrical or magnetic field, and within operating temperature range. | |||
DC Current | |||
Range | 999.9 A | ||
Resolution | 0.1 A | ||
Accuracy | 2% ± 5 digits | ||
AC Voltage | |||
Range | 600 V /1000 V | ||
Resolution | 0.1 V / 1 V | ||
Accuracy | 1.5% ± 5 digits (20 – 500 Hz) | ||
DC Voltage | |||
Range | 600.0 V /1000 V | ||
Resolution | 0.1 V / 1 V | ||
Accuracy | 1% ± 5 digits | ||
Frequency – Via Jaw | |||
Range | 5.0 – 500.0 Hz | ||
Resolution | 0.1 Hz | ||
Accuracy | 0.5% ± 5 digits | ||
Trigger Level | 5 – 10 Hz, ≥10 A 10 – 100 Hz, ≥5 A 100 – 500 Hz, ≥10 A | ||
Frequency via Flexible Current Probe | |||
Range | 5.0 to 500.0 Hz | ||
Resolution | 0.1 Hz | ||
Accuracy | 0.5% ± 5 digits | ||
Trigger Level | 5 to 20 Hz, ≥ 25 A 20 to 100 Hz, ≥ 20 A 100 to 500 Hz, ≥ 25 A | ||
Resistance | |||
Range | 600 Ω/6 kΩ/60 kΩ | ||
Resolution | 0.1 Ω/1 Ω/10 Ω | ||
Accuracy | 1% ± 5 digits | ||
Mechanical Specifications | |||
Size (L x W x H) | 277 mm *88 mm * 43 mm (55 mm for remote unit) | ||
Weight | 350 g | ||
Jaw Opening | 34 mm | ||
Flexible Current Probe Diameter | 7.5 mm | ||
Flexible Current Probe Cable Length (head to electronics connector) | 1.8 m | ||
Environmental Specifications | |||
Operating Temperature | -10°C to +50°C | ||
Storage Temperature | -40°C to +60°C | ||
Operating Humidity | Non condensing (< 10°C) ≤ 90% RH (at 10°C to 30°C) ≤ 75% RH (at 30°C to 40°C) ≤ 45% RH (at 40°C to 50°C) (Without Condensation) | ||
Operating Altitude | 2000 meters | ||
Storage Altitude | 12,000 meters | ||
Temperature Coefficients | Add 0.1 x specified accuracy for each degree C above 28°C or below 18°C | ||
Wireless Frequency | 2.4 GHz ISM Band 10 meter range | ||
Safety Compliance | ANSI/ISA S82.02.01:2004 CAN/CSA-C22.2 No. 61010-1-04 IEC/EN 61010-1:2001 to 1000V CAT III, 600V CAT IV. | ||
Double Insulation Clearance | Per IEC 61010-2-032 | ||
Double Insulation Creepage | Per IEC 61010-1 |
Shop for Fluke products online at: www.MyFlukeStore.com 1.888.610.7664
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the 381 and is the answer not in the manual?
Questions and answers