Advertisement
Special symbols:
![]() | Fault message | ![]() | 1-button adjustment |
![]() | ![]() | Input gas adjustment | |
![]() | Display peak value | ![]() | Password necessary |
![]() | Display TWA | ![]() | Battery 100% full |
![]() | Show STEL | ![]() | Battery 2/3 full |
![]() | Bump test mode | ![]() | Battery 1/3 full |
![]() | Fresh gas adjustment | ![]() | Battery empty |
Marking offset channels:
Feature
ToxicTwins (X-am 5000 / 5600)
CO H 2 compensation (X-am 5000/5600)
H2 offsetting (X-am 5600)
Reading on the display
HCN+
CO+
ch 4 +
For more information on the features, see the technical manual:
Only trained persons are permitted to carry out modifications to the instrument configuration.
To individually configure an instrument with standard configuration, the instrument must be connected to a PC via the USB infrared cable (order no. 83 17 409) or the E-Cal system. The configuration work is carried out using the
PC software Dräger CC Vision. The PC software Dräger CC Vision can be the following web address free of charge:
www.draeger.com/software.
To simplify matters, this document generally refers to gases (e.g. Ex/CO2 or IR Ex). This refers to all existing sensor types.
Standard instrument configuration:
Dräger X-am® 50001) | |
Bump test mode2) | Extended bump test |
Fresh-gas adjustment2)3) | On |
Hydrogen offset4) | On |
Operating signal2)5) | On |
Capture range10) | On |
Switch off2) | allowed |
LEL factor2) ch4 H2 | 4.4 vol. % (4.4 vol. % corresponds to 100 %LEL) 4.0 vol. % (4.0 vol. % corresponds to 100 %LEL) |
STEL2)6)7) (short-term average) | STEL function – disabled Average value duration = 15 minutes |
TWA2)7)8) (shift average) | TWA function – disabled Average value duration = 8 hours |
Alarm A19) | can be acknowledged, non-latching, pre-alarm, rising flank |
Alarm A1 at O2 sensor9) | cannot be acknowledged, latching, like main alarm, falling flank |
Selecting or disabling the capture ranges (only applies for the measuring mode):
The capture range is selected in the measuring mode (factory setting) and permanently disabled in adjustment mode.
The CC-Vision PC software can be used to select or disable the capture ranges for the measuring mode.
The following parameters can be changed on an instrument:
Name | Range | |
Password | Numerical range (3-digit) | |
Operating signal LED[1]) | Yes/No | |
Horn operating signal1) | Yes/No | |
Switch-off mode | "switch-off allowed" or "switch-off prohibited" or "switch-off prohibited at A2" | |
Shift length (TWA)[2]) | 60 - 14400 (in minutes) (Setting for exposure alarm) | |
Short-term exposure level (STEL)[3])[4][5]) | 0 - 15 (in minutes) (setting for exposure alarm) |
The following parameters can be changed on the sensors:
Name | Range | |
Alarm threshold A1 (in measuring unit) | Capture range – A2 | |
Alarm threshold A2 (in measuring unit) | A1 – measuring range end value | |
Interpretation type1) | Inactive, TWA, STEL, TWA+STEL | |
Alarm threshold STEL (in measuring unit)1) | Capture range – full scale value | |
Alarm threshold TWA (in measuring unit)1) | Capture range – full scale value |
To ensure that the values were correctly transferred to the gas detection instrument:
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking on the rechargeable battery for permitted rechargeable batteries and the corresponding temperature class.
Substitution of components may impair intrinsic safety.
1)Only when activated in the instrument configuration. Delivery condition: not activated.
If the gas detector is used for offshore applications, a distance of 5 m to a compass must be complied with.
Explosion hazard! To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements:
Special states in which there is no measuring operation (quick menu, calibration menu, warm-up of sensors, password input) are indicated by a visual signal (slow flashing of the alarm LED
).
If an IR sensor is used in the Dräger X-am 5600, a adjustment of zero point and sensitivity must be carried out after experiencing an impact load that results in a non-zero display when exposed to fresh air.
Incorrect measured values!
If water closes the gas inlets of the gas detector (e.g. by submerging the gas detector under water or by heavy rain), this may result in incorrect measured values.
Shake the gas detector with the display facing downwards in order to remove the water.
An alarm is displayed visually, audibly and through vibration in a specific pattern.
At low temperatures the legibility of the display can be improved by switching on the backlight.
Concentration pre-alarm A1
Intermittent alarm:
Concentration main alarm A2
Intermittent alarm:
Risk of fatal injury! Leave the area immediately.
A main alarm is self-retaining and cannot be acknowledged or cancelled.
After leaving the area, when the concentration has dropped below the alarm threshold:
The measuring range 0 to 100 vol. % CH4 is not suitable for monitoring explosive mixtures in the measuring range of 0 to 100 %LEL.
STEL/TWA exposure alarm
Intermittent alarm:
Display »A2« and (STEL) or respectively
TWA) and measured value alternating:
Health hazard! Leave the area immediately. After this alarm, the deployment of personnel is subject to the relevant national regulations.
The STEL alarm can be triggered with a maximum delay of one minute.
Battery pre-alarm
Intermittent alarm:
Battery main alarm
Intermittent alarm:
Instrument alarm
Intermittent alarm:
Activating the Info mode
Info Off mode
Calling the Quick Menu
1) The fresh-gas adjustment/zero-point adjustment is not supported by the CO2 channel of the infrared sensor and the XXS O3. A zero-point adjustment of these sensors can be made with the Dräger CC-Vision PC software. A suitable zero gas free from carbon dioxide and ozone (e.g. N2) must be used.
Explosion hazard! To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive areas.
Do not use new batteries with used batteries, and do not mix batteries from different manufacturers or of different types. Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
1) Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100) with the appropriate Dräger charger. Charge NiMH single cells for ABT 0100 battery holder as directed by the manufacturer. Ambient temperature during charging: 0 to +40°C.
Charge instrument with NiMH power pack T4 (Type HBT 0000)/ T4 HC (Type HBT 0100)
Explosion hazard! To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following warning statements:
Do not charge underground or in explosion hazard areas! The chargers are not designed in accordance with the regulations for fire damp and explosion protection. Charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100) with the appropriate Dräger charger. Ambient temperature during charging: 0 to +40°C.
Even if the instrument is not used, we recommend that you store the instrument in the charger!
To protect the battery charge only in the temperature range of 5 to 35°C. Outside this temperature range, the charging process is automatically interrupted and automatically continued after the temperature range has been reached again. The charging time is typically 4 hours. A new NiMH power pack reaches its full capacity after three complete charge/discharge cycles. Never store the instrument for extended periods without being connected to a power source (maximum of 2 months) because the internal buffer battery will drain.
In the case of a manual function test the effect of the H2 offsetting must be taken into account accordingly.
A potentially activated H2 offsetting is automatically temporarily deactivated during a manual calibration, a PC calibration or an automatic Bump Test for the relevant duration.
The automatic functional test with the Bump Test Station is described in the instructions for use for the bump test station and in the Technical Manual.
To check the measured value response times, apply t90 test gas to the X-am via the calibration cradle. Check the results in accordance with the information in the table in the enclosed supplementary documentation (order no. 90 33 890) until 90% of the end display is reached.
After the bump test (menu), the display shows a printer icon even if there is no printer connected to the bump test station.
Adjustment may not be possible due to instrument and channel errors.
A potentially activated H2 offsetting is automatically temporarily deactivated during a manual calibration, a PC calibration or an automatic Bump Test for the relevant duration.
Dräger recommends using the extended bump test for cross adjustments (Dräger X-dock technical manual).
Carrying out the fresh air adjustment
Adjust the instrument to fresh air, free of measured gases or other interfering gases. During the fresh air adjustment, the zero point of all sensors (with the exception of the DrägerSensor XXS O2 and XXS CO2) is set to 0. With the DrägerSensor XXS O2, the display is set to 20.9 vol. % and with DrägerSensor XXS CO2 to 0.03 vol. %.
The fresh air adjustment/zero adjustment is not supported by the DrägerSensor XXS O3. A zero-point calibration/adjustment of this sensor can be carried out using the Dräger CC Vision software. A suitable zero gas free from ozone (e.g. N2) must be used.
If a fault has occurred during the fresh air adjustment:
Adjusting the sensitivity for an individual measuring channel
infrared sensor | 0.05 to 5 vol. %2) |
Ex: | 40 to 100 %LEL |
O2 | 10 to 25 vol. % |
CO: | 20 to 999 ppm |
H2S: | 5 to 99 ppm |
NO2 | 5 to 99 ppm |
Test gas concentration of other gases: see instructions for use of the respective DrägerSensors.
1) Depending on the selected data set.
2) Depending on the measuring range and the measuring accuracy.
If a fault has occurred during the span adjustment:
The instrument does not need any special care.
Abrasive cleaning tools (brushes etc.), cleaning agents and cleaning solvents can destroy the dust and water filters.
For information on suitable cleaning and disinfecting agents and their specification, see document 9100081 at www.draeger.com/IFU.
Carefully dry the instrument with a cloth.
The instrument should be inspected and serviced once a year by suitably qualified persons. Comparisons:
Recommended calibration interval for measuring channels Ex, O2, H2S, SO2, NO2 and CO: 6 months.
Calibration intervals of other gases: see Instructions for Use of the respective DrägerSensors.
See the Technical Manual for details of spare parts.
The manufacturer's calibration certificate can be https://www.draeger.com/productioncertificates.
Test gas properties (such as relative humidity, concentration, etc.) can be found on the corresponding data sheet of the sensor. The relative humidity is not relevant for the O2 sensor. Always use dry test gas.
Risk to health. The sensor contains corrosive liquids. In case of leaks, avoid contact with eyes and skin. In case of contact, rinse with plenty of water.
For further information on the use of the Dräger sensor, please access the following link: www.draeger.com/sensorhandbook
Excerpt See the Technical Handbook for details1)
Ambient conditions: during operation and storage
Temperature class T4 (–20 to +50°C):
NiMH power packs type: HBT 0000, HBT 01002)
Power pack type: ABT 01002) with alkaline single cell type: Duracell Procell MN 15002), Duracell Plus Power MN 15002)
Temperature class T3 (–20 to +40°C): Power pack type: ABT 0100 with NiMH single cell type: GP 180AAHC2)
with alkaline single cell type: Panasonic LR6 Powerline
Temperature class T3 (0 to +40°C): Power pack type: ABT 0100 with alkaline single cell type: Varta 40062), Varta 41062)
Temperature range over a short period2):
–40 to +50°C
Maximum of 15 minutes with NiMH power pack T4 (HBT 0000) or T4 HC (HBT 0100)
Requirement: storage of the instrument at room temperature (+20°C) for at least 60 minutes in advance.
Air pressure | 700 to 1300 hPa |
Humidity | 10 to 90% (up to 95% short-term) rel. hum. |
Position of use | any |
Storage time X-am 5600 | 1 year |
Sensors | 1 year |
electrical classification | IP 67 for instrument with sensors |
Alarm volume | Typically 90 dB (A) at 30 cm distance |
Operating time: Alkaline battery | Typically 9 hours under normal conditions |
NiMH power pack: T4 (HBT 0000) T4 HC (HBT 0100) | Typically 9 hours under normal conditions when using Dual IR Ex/CO2 Typically 12 hours under normal conditions when using Dual IR Ex/CO2 ES Typically 10.5 hours under normal conditions |
Dimensions | approx. 130 x 48 x 44 mm (H x W x D) |
Weight | approx. 220 to 250 g |
Refresh interval for display and signals | 1 s |
The technical manual, the quick reference guide and the sensor documentation can be downloaded in electronic form from the database for technical documentation (www.draeger.com/ifu). To do this, enter the part number or product name in the search dialogue.
Safe coupling with electrical device
Electrical connections to devices which are not listed in these instructions for use should only be made following consultation with the respective manufacturers or an expert.
Use in areas subject to explosion hazards
Devices or components for use in explosion-hazard areas which have been tested and approved according to national, European or international explosion protection regulations may only be used under the conditions specified in the approval and with consideration of the relevant legal regulations. The devices or components may not be modified in any manner. The use of faulty or incomplete parts is forbidden. The appropriate regulations must be observed at all times when carrying out repairs on these devices or components.
Intended use
Definitions of alert icons
The following alert icons are used in this document to provide and highlight areas of the associated text that require a greater awareness by the user. A definition of the meaning of each icon is as follows:
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided, could result in physical injury, or damage to the product or environment. It may also be used to alert against unsafe practices.
Indicates additional information on how to use the product.
Portable gas detection instrument for the continuous monitoring of the concentration of several gases in the ambient air within the working area and in explosion-hazard areas.
Independent measurement of up to 6 gases, in accordance with the installed Dräger sensors.
Marking
Copies of the name plate and the declaration of conformity are provided in the enclosed supplementary documentation (order no. 90 33 890). Do not stick anything on the name plate on the gas detector.
The technical suitability tests are valid for the X-am 5600 gas detection instrument and the calibration cradle. The explosion-protection approvals are only valid for the X-am 5600 gas detection instrument; the calibration cradle must not be used in the Ex zone.
The BVS 10 ATEX E 080 X technical suitability test is based on the adjustment with the target gas.
Areas subject to explosion hazards, classified by zones
The instrument is intended for the use in explosion-hazard areas of Zone 0, Zone 1 or Zone 2 or in mines at risk due to fire damp. It is intended for use within a temperature range of –20°C to +50°C, and for areas in which gases of explosion groups IIA, IIB or IIC and temperature class T3 or T4 (depending on the batteries and rechargeable battery) may be present. If used in mines, the instrument is only to be used in areas known to have a low risk of mechanical impact.
Areas subject to explosion hazards, classified by divisions.
The instrument is intended for use in explosion-hazard areas according to Class I & II, Div. 1 or Div. 2 within a temperature range of –20°C to +50°C, and for areas where gases or dusts of groups A, B, C, D, E, F, G and temperature class T3 or T4 may be present (depending on the rechargeable battery and batteries).
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly adhere to the following caution and warning statements:
Increased hydrogen concentrations within the measuring range of the Dräger Sensor XXS H 2 HC may result in false alarms due to the additive effect on the Dräger sensors XXS H2S, and XXS CO, XXS H 2 S (LC) and XXS CO (LC) as well as due to the negative effect on the Dräger Sensor XXS O 2 .
Only use power packs ABT 0100 (83 22 237), HBT 0000 (83 18 704) or HBT 0100 (83 22 244). See marking on power pack for approved batteries and related temperature class.
Do not change batteries in hazardous locations.
To reduce the danger of explosion, do not mix new batteries with old batteries and do not mix batteries made by different manufacturers.
Always disconnect the instrument from the power pack before carrying out any maintenance operations.
Substitution of components may impair intrinsic safety.
Not tested in an oxygen-enriched atmosphere (>21% O2).
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the X-am 5600 and is the answer not in the manual?
Questions and answers