WARNING!
!
Incorrect installation or removal of meters
can result in serious injury or death. Read
the instructions in this guide on the proper
procedures carefully.
Any person installing, inspecting, or maintaining
a McCrometer flowmeter should have a working
understanding of piping configurations and systems
under pressure.
Before adjusting or removing any meter, be certain
the system has depressurized completely.
Be careful when lifting meters. Meters can cause
serious injury if lifted incorrectly or dropped.
Electromagnetic Flow Meter
Quick Start Installation Guide

About This Quick Start Guide

This Quick Start Guide is a supplement to the Installation,
Operation and Maintenance manual supplied with this
meter. It is intended to be a quick reference for the basic
installation and reading of the Ultra Mag.
It is designed to provide installation instructions when the
location of the sensor installation has been predetermined.
Refer to the meter manual (30126-25 Vera Mag 3000 IOM
manual) for information on these topis:
Preparation and planning for installation
Site location
Software configuration
Detailed information on external connections, external
power, outputs
About This Quick Start Guide. . . . . . . . . . . . . . . . .1
1.
Flow Meter Installation . . . . . . . . . . . . . . . .2
2.
3.
4.
Grounding . . . . . . . . . . . . . . . . . . . . . . . .6
5.
Activating The Display . . . . . . . . . . . . . . . . .7
6.
Converter Boot. . . . . . . . . . . . . . . . . . . . . .7
7.
Converter Configuration . . . . . . . . . . . . . . .8
8.
Page 1
Vera Mag 3000
Battery Powered
30126-58 Rev. 1.0 | 13SEP2024
30126-58
13SEP2024

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Vera Mag 3000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for McCrometer Vera Mag 3000

  • Page 1: Table Of Contents

    Ultra Mag. It is designed to provide installation instructions when the location of the sensor installation has been predetermined. Refer to the meter manual (30126-25 Vera Mag 3000 IOM manual) for information on these topis: •...
  • Page 2: Flow Meter Installation

    ® Flow Meter Installation The Ultra Mag requires grounding to ensure proper functioning and accuracy. See section 4 for grounding instructions and methods. Flanged Meter Installation Install the Ultra Mag flow meter inline between two flanged end pipes. The flow meter may require grounding, depending on the environment they are being installed in.
  • Page 3: Remote Mount Converter Installation

    Connections to the sensor must be made with cable supplied by McCrometer specifically for that purpose. Do not substitute the supplied cable with other types of cable, even for short runs. For repairs or added lengths of cable, the entire cable between the sensor and the converter must be replaced.
  • Page 4: Converter Wiring And Connection

    ® Converter Wiring and Connection External Output Connectors The flow meter is configured at the factory for the optional outputs and requested by the customer at the time of order. The external cables attach through a permanent cable gland (standard) or through a screw locking-type waterproof connector (quick connect option).
  • Page 5: Wiring Diagrams

    Green right. Green • Signals and associated wire colors in the McCrometer SmartOutput™ Black interface cable are identified together in the top row of the table above. Black •...
  • Page 6: Grounding

    ® Grounding 1. Preferred method of grounding Ultra Mag meters come standard with a set of grounding rings for use with the preferred method of grounding Ultra Mag meters. This method can be used for all installations, but it is required for non-conductive or internally coated pipe. When pipes are non-conductive, such as PVC or internally coated pipe, you must substitute direct grounding with grounding rings.
  • Page 7: Activating The Display

    Rate unit setting There is an optical sensor embedded in the display 123456789 GAL located under the McCrometer “swirl” logo at the Total unit Total lower left. If the display does not appear, a flashlight x1000...
  • Page 8: Converter Configuration

    Copyright © 2023 McCrometer, Inc. All printed material should not be changed or altered without permission of McCrometer. Any published pricing, technical data, and instructions are subject to change without notice. Contact your McCrometer representative for current pricing, technical data, and instructions.

Table of Contents