9. Care, Maintenance and
Disposal
The CapsoAccess® Capsule Data Access System
is typically used in an office environment with
clean, disinfected, dry capsules. The system
should be cleaned only as needed. More
extensive cleaning may be required if the
system is accidentally soiled or contaminated.
Follow the recommendations herein.
Cleaning: Disconnect the power and USB
cables prior to cleaning. The system chassis
contains electronics. Keep liquids from entering
the chassis. Avoid submerging the system in
liquid or sterilizing it in an autoclave. All
exposed surfaces may be wiped with a swab or
cloth moistened in isopropyl alcohol or other
disinfectant. Clean the receptacle with a swab
including the underside of the rubber gasket
that comes in contact with the capsule.
Disposal: After its useful life or after sustaining
unrepairable damage, the CapsoAccess®
Capsule Data Access System and its power
supply should be disposed of in an
environmentally sound manner. Please consult
and obey national, state, and local laws and
ordinances governing the safe disposal of
electronic equipment.
General: Contact your local distributor or
CapsoVision representative if capsule docking
becomes unreliable.
8
10. Specifications and Ratings
Class I medical (Grounded) device
The CapsoAccess® Capsule Data Access System
is classified as NO APPLIED PARTS EQUIPMENT.
The CapsoAccess® Capsule Data Access System shall
be classified as ORDINARY EQUIPMENT, not intended
or evaluated for use in the presence of flammable
anesthetic mixture with air, oxygen, or nitrous oxide.
The CapsoAccess® system needs special
precautions regarding EMC and needs to be
installed and put into service according to the
EMC information provided in this manual.
Portable and mobile RF communications
equipment can affect the CapsoAccess® system.
The use of accessories, transducers and cables
other than those specified by CapsoVision, Inc.
may result in increased EMISSIONS or decreased
IMMUNITY of the CapsoAccess® system.
This CapsoAccess® system should not be used
adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary,
the CapsoAccess® system should be observed
to verify normal operation in the configuration
in which it will be used.
Essential Performance: Data integrity during
video download is considered to be the essential
performance of the CapsoAccess® system. If the
essential performance is lost or degraded due to
EM disturbances, the video data may be
IFU-2800 Ref F 03/2020