No Pressure Trigger - Zero Transducer
| CONDITION
| CORRECTIVE ACTION
|
ie
Arrempting го use pressure trigger
| 1. Zero the transducer by opening the
|
without having the transducer
©
transducer to air and pressing the
1
zeroed.
i
ZERO key for two seconds.
2, Resume pumping by pressing the
ASSIST/STANDBY key.
| CONDITION
| CORRECTIVE ACTION
| ©
The patient heart rate is less than
1. Assess the patient.
40 bpm.
|
|
|
:
2. Verify proper timing.
|
|
|
|
| e
There is inadequate signal
|1. Check electrode contact and
|
acquisition from the patient
|
placement. Verify the integrity of
|
electrodes.
|
the lead wires and the patient cable.
2. Adjust ECG GAIN by using the
©
i
ECG GAIN Auxiliary keys.
|
|
|
i
!
ul
|
i
TRIGGER SELECT keys, when
|
demand pacing a patient.
|
_
Press the ASSIST/STANDBY key to
resume pumping.
le
The patients AV paced rate is
| 1. Reduce the pacer rate.
above 125 BPM.
|
¡e
The patients V paced rate is above
|
185
BPM.
| e
The system is unable to identify
| 1. Adjust the A-V time interval to
|
the A-V interval.
|
between 80 to 224 msec.
System 98 Service Manual
1-73
Chapter 1, Operation