4.9.2
Display Internal Statistics
This screen shows the following system information:
e
Pump cycle count (number of times the [AB has been inflated)
e
Accumulated assist time (time in hours:minutes that the system has been in
ASSIST Mode)
e
Battery statistics:
- Last 3 start of discharge times
- Last 3 end of discharge times
- Last 3 full charge or top off times
This screen allows the internal statistics to be cleared by pressing the REF. LINE
key twice. This should only be done if the non-volatile Ram has been corrupted
or the Main board has been replaced.
4.9.3
Display Software Revision
Display Software Revisions of System 98.
4.10 REMOTE DIAGNOSTIC CONTROL
For Future Use.