NOTE: If the AutoFill procedure fails to purge and fill the safety disk properly,
the message "AUTOFILL FAILURE' will be displayed and an audible alarm
activated. Corrective action can be obtained by pressing the HELP SCREEN
key (29). MANUAL
FILL help screens, as well as detailed fill illustrations
(See Section 1.3.2.14) are provided should MANUAL FILL become necessary.
1.3.2.13 System Power-Up Defaults
The power-up settings depend upon the amount of time the System has been
powered down.
The System recognizes when it has been powered down for less than 15 minutes.
When the power is restored within the 15 minute period, the System retains the
user settings (with exceptions as shaded in the table below) and trend. The shaded
items always return to their default setting.
When the System is powered down for longer than 15 minutes the factory
default setting are restored. At this point when the System is powered on, the
user settings and the trend are cleared.
To adopt the factory default settings prior to a full 15 minutes power down, press
the CHANGE/SELECT key (25) while powering on the System.
The following table lists all of the controls and their default settings. Default
settings are programmed at Datascope and cannot be changed.
| CONTROLS
DEFAULTS
|
i Trigger Select
ECG
| IAB Frequency
b
IAB Augmentation
LAB Inflation
Manually Set - No Default
|
{LAB Deflation
Manually Set - No Default
|
ECG Lead
II
Pressure Source
Internal
|
Pressure Zero
Not Zeroed
ECG Gain
Pressure Threshold
Normal
Internal Rare
[Normal (Fixed 80 BPM)
Slow Gas
[ON
Timing
[AUTO
| IAB Fill Mode
AUTO
System 98 Service Manual
1-39
Chapter 1, Operation