Order number: 7131-01; 7131-01/002; 7131-02; Note, e.g. specific instructions regarding 7131-02/002; 7131-02/021 efficient and cost-effective use of the unit. V 1200 S Order number: 7136-02; 7136-02/002; 7136-02/021; 7136-02/003; 7136-03 Refer to Operating Instructions. Warnings and symbols Warnings Wear protective gloves.
Important information Safety CE labelling with the number of the noti- fied body Dürr Dental has designed and constructed this unit so that when used properly and for the Serial number intended purpose it does not pose any danger to people or property.
Important information Specialist personnel The unit must not be used as a vacuum ❯ cleaner. Operation Do not use chemicals containing chlorine or ❯ Unit operating personnel must ensure safe and foaming chemicals. correct handling based on their training and Operation in operating theatres of explosive ❯...
Important information Dürr Dental accepts no liability for dam- ages or injury resulting from the use of non-approved accessories or optional accessories, or from the use of non-origi- nal wear parts or replacement parts. The use of non-approved accessories, optional accessories or non-genuine wear parts / replacement parts (e.g.
Product description Technical data V 600 Electrical data 7127-01 Rated voltage 230, 1~ Mains frequency Nominal current ≤ 5.0 Start-up current ~ 22 Motor protection * Type of protection IP 24 Protection class Motor winding overheat protector Connections Æ 50 Suction connection (outside) Exhaust air connection (external) Æ...
Page 12
Product description Classification Medical devices class 9000-606-89/30 1912V003...
Page 13
Product description V 600 Electrical data 7127-02 Rated voltage 400, 3~ Mains frequency Nominal current ≤ 2.2 ≤ 2.3 Start-up current ~ 16 ~ 14 Motor protection * Type of protection IP 24 Protection class Max. permitted setting Connections Æ 50 Suction connection (outside) Exhaust air connection (external) Æ...
Page 14
Product description V 900 S Electrical data 7131-01 7131-02 Rated voltage 230, 1~ 400, 3~ Mains frequency Nominal current ≤ 7.4 ≤ 3.6 Start-up current ~ 32 ~ 25 Motor protection * 10.0 Type of protection IP 24 Protection class Max.
Product description V 1200 S Electrical data 7136-02 7136-03 Rated voltage 230, 3~ 400, 3~ 400, 3~ Mains frequency Nominal current ≤ 6.5 ≤ 3.8 ≤ 4.0 Start-up current ~ 43 ~ 25 ~ 17 Motor protection * Type of protection...
V 600 The type plate is located on the upper part of the turbine housing. Type plate V 900 S, V 1200 S The type plate is located on the upper part of the turbine housing. Type plate Evaluation of conformity...
Product description Operation Motor Turbine wheel Intake nozzle Exhaust air muffler Exhaust air connection The V-suction unit is suitable for use in dry air suction systems. The advantage of this system is that the suction unit, regardless of the actual connection layout, can be installed in any available and suitable room (including upper floors or basements).
Page 18
Product description On the vacuum side the V-suction unit is equipped with a condensation separator that collects any con- densation arising within the pipe system and transports it away to the outside. An auxiliary air valve in the condensation separator protects the suction unit against overheating and provides uniform suction power.
Assembly Hose materials Assembly For waste connections and suction lines only use the following hose types: – Flexible spiral hoses made of PVC with inte- Requirements grated spiral or equivalent hoses The unit can be installed either on the same level –...
Assembly System components Current consumption of Cross-section unit [A] The system components listed below are required or recommended for various procedures > 25 and < 32 or for installation. > 32 and < 40 Control box > 40 and < 50 The unit is connected via a control box.
Assembly Installation The actual connection can vary depending on the chosen installation option. The connection shown is only an example. Installation and routeing of hoses and pipes Establish connections between the pipe system and the unit using the flexible hoses supplied. This ❯...
Assembly 230 V 1/N/PE Electrical connections DANGER Electric shock due to incorrectly con- nected device Never install a mains plug instead of ❯ the fixed connection. NOTICE Short circuit due to defective lead Do not route wires near hot surfaces. ❯...
Assembly 10 Commissioning In many countries technical medical prod- ucts and electrical devices are subject to regular checks at set intervals. The owner must be instructed accordingly. Turn on the unit power switch or the main ❯ surgery switch. Carry out a function check of the system. ❯...
Usage The following are required for disinfection/clean- Usage ing: ü Non-foaming disinfectant/cleaning agent that is compatible with the materials. 11 Disinfection and cleaning ü Unit care system, e.g. OroCup To pre-clean, suck up 2 litres of water with the ❯ NOTICE care system.
Usage 12 Maintenance All maintenance work must be performed by a qualified expert or by one of our Service Techni- cians. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯ Wear protective equipment when working (e. g. impermeable gloves, protective goggles and ❯...
Troubleshooting Troubleshooting 13 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯...
Page 27
Troubleshooting Error Possible cause Remedy Suction performance too low Coarse filters in system clogged Clean coarse filters. ❯ (e.g. at separator devices) Leak in the suction line Check and if necessary ❯ establish leak-tightness of suction system and connec- tions. * Mechanical sluggishness of tur- Disassemble the unit and ❯...
Disinfect a defective unit using a suitable sur- ❯ face disinfection agent. Seal all connections with sealing caps. ❯ Pack the unit securely in preparation for trans- ❯ port. Fig. 3: V 900 S, V 1200 S Sealing cap 9000-606-89/30 1912V003...
Page 29
Appendix Appendix 15 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
Page 32
Hersteller/Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...