Page 4
Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only. Electrical Ground: This symbol inside the instrument marks the central safety grounding point for the instrument. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 5
Note that the safety of a system that may incorporate this product is the end user’s responsibility. For Technical Assistance regarding the use and maintenance of this instrument or any other Teledyne API product, contact Teledyne API’s Technical Support Department: Telephone: +1 800-324-5190 (toll free) or +1 858-657-9800 Email: api-techsupport@teledyne.com or access any of the service options on our website at http://www.teledyne-api.com/...
Page 6
être effectuées uniquement par du personnel de maintenance qualifié. Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le point central de la mise à la terre sécuritaire de l’instrument. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 7
Risque d’explosion et d’incendie Techniques de manutention, Blessure corporelle soulevage et de transport inappropriées Notez que la sécurité d’un système qui peut incorporer ce produit est la responsabilité de l’utilisateur final. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Page 8
PRODUCT RETURN All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.
Page 9
Provides information about that which could either affect accuracy of instrument readings or cause loss of data. Provides information pertinent to the proper care, NOTE operation or maintenance of the instrument or its parts. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
INTRODUCTION, SPECIFICATIONS, APPROVALS, & COMPLIANCE The Teledyne API Models N360 and N360M differ only in specifications; unless clearly differentiated, both models in this manual are referred to as the N360 for simplification. The N360/N360M measures carbon dioxide CO by comparing infrared energy absorbed by a sample to that absorbed by a reference according to the Beer-Lambert law.
0 - 95% RH, non-condensing Dimensions H x W x D 7" x 17" x 24.3" (178 mm x 432 mm x 617 mm) Weight 40 lbs. (18.1 kg) At constant temperature and voltage. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
COMPLIANCE AND CERTIFICATIONS This product is CE compliant and adheres to the Low Voltage and ElectroMagnetic Compatibility Directives. For any other certifications, please refer to this product’s specifications sheet on our website. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Do not operate this instrument without first removing dust plugs from SAMPLE and EXHAUST ports on the rear panel. Note Teledyne API recommends that you store shipping containers and materials for future use if/when the instrument should be returned to the factory for repair and/or calibration service. See Warranty statement in this manual and Return Merchandise Authorization (RMA) on our Website at http://www.teledyne-api.com.
MINIMUM REQUIRED CLEARANCE Back of the instrument 10 cm / 4 in 2.5 cm / 1 in Sides of the instrument Above and below the instrument 2.5 cm / 1 in Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
The front panel (Figure 2-1) includes two USB ports for peripheral device connections, which can be used with mouse and keyboard as alternatives to the touchscreen interface, or with flash drive for uploads/downloads (devices not included). Figure 2-1. Front Panel Layout 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
For voltage or current loop outputs to a strip chart recorder and/or a data logger. ETHERNET Connector for network or Internet remote communication, using Ethernet cable. Serial communications port for RS-232 or RS-485 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Teledyne API recommends that you store shipping containers and materials for future use if/when the instrument should be returned to the factory for repair and/or calibration service.
Setup>Analog Outputs menu. Ground Ground Isolated Current Outputs I Out + I Out - I Out + User definable through the I Out - Setup>Analog Outputs menu. I Out + I Out - 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Outputs; they will work with relays, open collectors, or 3.3 V – 24 V logic. Pull low to activate. DI1 and DI2 are fixed (not mappable) for remote zero and span calibrations. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
NTERFACES For remote communications, the rear panel provides Ethernet or RS-232 connectors. In addition to using the appropriate cables, the communication method must be configured using the SETUP>COMM menu, Section 2.5.10). 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
232Rx (default) DTE RS232 2-3, 5-6, 9-12 232Rx 232Tx RS-232 Baud rate: 115200 bits per second (baud) • Data Bits: 8 data bits with 1 stop bit • Parity: None • Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
• at least 0.2 m long • no more than 2.0 m long • vented outside the shelter or immediate area surrounding the instrument. Dust Plugs: Remove dust plugs from rear panel exhaust and supply 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
TAPI 700-Series and a source of dried air scrubbed of CO such as a bottle of N In the case of CO measurements made with the Teledyne API’s Model N360/N360M Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
7.5 psig. If the source of the purge air is shared by a TAPI 700-Series model, the minimum gas pressure should be 25 psig and should not exceed 35 psig. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Source of is pressurized SAMPLE GAS Calibrated Removed during calibration at span gas concentration SAMPLE VENT EXHAUST VENT SPAN Chassis PRESSURE SPAN PURGE LINE Figure 2-8. Pneumatic Connections–Basic Configuration–Using Bottled Span Gas Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
PTION The following illustrations show the pneumatic connections for the ambient zero and ambient span valve option. The minimum span gas flow rate required for this option is 800 cm /min. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
TAPI 700-Series gas Dilution at 100x span gas concentration Calibrator SAMPLE EXHAUST VENT SPAN Chassis PRESSURE SPAN PURGE LINE VENT Figure 2-10. Pneumatic Connections with Ambient Zero/Ambient Span Valves (OPT 50A) Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
This valve option is intended for applications where zero air and span gas are supplied by a zero air generator and gas dilution calibrator such as those in the Teledyne API 700- Series. Internal zero/span and sample/cal valves control the flow of gas through the instrument, but because the calibrator limits the flow of span gas no shutoff valve is required.
Figure 2-13. Pneumatic Connections with Ambient Zero/Pressurized Span Valves and External Zero Air Scrubber This option requires that both zero air and span gas be supplied from external sources. It is specifically designed for applications where span gas will be supplied from a Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
0 ppm and the concentration of the calibrated gas. Figure 2-14 depicts the pneumatic set up in this sort of application of a Model N360 CO Analyzer with ambient zero/ambient span valve option installed (a common configuration for this type of application).
7. Make sure the zero air supply to the analyzer’s purge line inlet is supplying gas at a stable pressure above 7.5 psig. 8. Adjust the N360’s pressure regulator until the attached gauge reads 7.5 psig. 9. Turn off the instrument.
INLET EXHAUST GAS OUTLET Flow / Pressure SAMPL E PRESS URE Sensor PCA SENSO R PUMP FLOW SENSO R PURGE IN Purge Gas Pressure Control Figure 2-15. Pneumatic Flow, Basic Configuration 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
(Normal State) Zero/Span Open to IZS inlet Sample/Cal Open to ZERO/SPAN valve ZERO CAL Zero/Span Open to IZS inlet Sample/Cal Open to ZERO/SPAN valve SPAN CAL Zero/Span Open to PRESSURE SPAN inlet Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Open to ZERO/SPAN valve SPAN CAL Zero/Span Open to SHUTOFF valve Shutoff Valve Open to PRESSURE SPAN Inlet The minimum span gas flow rate required for this option is 800 cm /min. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
GAS OUTLET Sensor Flow / Pressure SAMPL E PRESS URE Sensor PCA SENSO R FLOW SENSO R PURGE IN Purge Gas Pressure Control Figure 2-18. Pneumatic Flow with the O Sensor Option Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
(Section 2.3.4.4) and run an initial calibration (Section 2.3.4.5). Section 2.4 introduces the menu system, and Section 2.5 provides setup/customization instructions. 2.3.4.2 TARTUP Upon initial startup, a sequence of status screens (Figure 2-19) appear prior to the Home page (Figure 2-20). 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
The Alerts page (Figure 2-21) shows the status of any active warning conditions or user-configured Events. (Section 2.4.3 provides more detailed information about Alerts, and Section 2.5.2 addresses Events). Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
(If any functional parameters are not displayed, configure the Dashboard through the Setup>Dashboard menu to add them; see Section 2.4.2). These functions are also useful tools for diagnosing problems (information provided in Section 5.7.2). Figure 2-22. Sample Dashboard Page 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
The method for performing a calibration differs slightly depending on whether or not any of the available internal zero air or valve options are installed. Follow the appropriate calibration instructions in Section 4. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Time settings*; and check for firmware updates when Section 2.5.9 connected to a network that is connected to the Internet. Time Zone change requires special procedures (Section 5.5). Comm View and configure network and serial communications. Section 2.5.10 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Figure 2-23 presents an orientation to the main display screen; Figure 2-24 shows that pressing the gas name or its concentration value or a meter below displays a live plot of their respective readings. Section 2.5.6 provides configuration instructions. Figure 2-23. User Interface Orientation Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
The demodulated, peak IR detector output during the measure portion of the GFC IR Measure Wheel cycle. The demodulated, peak IR detector output during the reference portion of the GFC IR Reference Wheel cycle. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Alerts menu on the Home screen or by pressing the Alerts shortcut (Figure 2-26). While Alerts can be cleared from the Active Alerts page, they remain recorded in the Utilities>Alerts Log menu. Figure 2-26. Navigating to the Active Alerts Page 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
The Calibration menu is used for multipoint calibrations as well as calibrations for zero and span with valve options, and for external calibration with valve options installed. Calibration procedures are presented in Section 4. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
(See Section 4.4.3). 2.4.6 SETUP The Setup menu is used to configure the instrument’s various features, functions, and data log. Section 2.5 provides details for the menus under Setup. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Sections 2.5.1.1 and 2.5.1.2 providing additional details. To transfer captured instrument data to a flash drive see Section 2.5.1.3. Figure 2-29. Datalog Configuration, New Log Page Figure 2-30. Datalog Configuration, Existing Log Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Figure 2-31. Creating a New Data Log The parameters available in the list of Log Tags include the names of Events configured in the Events page (Section 2.5.2). 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Some Events are standard and not editable while others are user-configurable, described here. Existing Events are listed in the Events page (Figure 2-35) under the Setup menu. Figure 2-35. Events List 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
(Figure 2-35). Press ADD to create a new Event (refer to Figure 2-36 for details), or select an existing Event to either Edit or Delete it (Figure 2-38). Figure 2-36. Event Configuration Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Select an Event from the list (Figure 2-35) and press the Edit button to view or edit the details (Figure 2-37), or press the Delete button to delete the Event. Figure 2-38. Edit or Delete an Event 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Name field of the Events Configuration page will appear in the list of Log Tags of the Datalog Configuration page. The Data Logger is presented in Section 2.5.1. 2.5.3 SETUP>DASHBOARD Figure 2-39. Dashboard Display and Configuration Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
(a concentration that matches the diluted gas concentration prior to having input the Dilution Factor). [Enable] Software Set instrument to continue sampling, while ignoring calibration, diagnostic, Maintenance Mode and reset instrument commands. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Setup>Homescreen menu or from Home page using the configuration icon (Figure 2-40). Figure 2-40. Homescreen Configuration An orientation to the Homescreen was presented in Section 2.4.1, including Figure 2-23 and Figure 2-24. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Map the Digital Relays in the same manner. For testing only, navigate to the Utilities>Diagnostics>Digital Outputs menu to change the state (ON/OFF) of individual digital outputs. Figure 2-41. Digital Outputs Setup 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Map the four user-configurable Analog Outputs for either voltage output or current output to any of a wide variety of “Signals” present in the instrument and customize their respective configurations. Figure 2-42. Analog Output Configuration for Voltage Output, Example Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 67
(Table 2-12). For Current output, press the +100 button several times to get the setting • close to 4mA. 6. Press the Accept button when adjustment reached. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Each range is usable from -5% to +5% of the rated span. For manual calibration adjustments, see Section 2.5.8.1 for voltage and Section 2.5.8.2 for current. Figure 2-44. Analog Output Calibration, Voltage or Current Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
+100 button to arrive at a realistic starting point for 4mA. CAUTION! Do not exceed 60 V peak voltage between current loop outputs and instrument ground. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
When an instrument is connected to a network that is connected to the Internet, follow the instructions on this Remote Update page to check for and activate software/firmware updates. (Also refer to Section 5.3). Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Stop bits Select either 0 or 1 stop bit (typically set in conjunction with Parity and Data bits). 2.5.10.2 TCP P This menu is configured with the port number for MODBUS. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
A string of numbers very similar to the Instrument IP address (e.g. Default 192.168.76.1.) that is the address of the computer used by your LAN and serves Gateway as a router to access the Internet or another network. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 73
7. When the Status field indicates that the USB drive has been detected, press the “Upload Configuration to Instrument” Start button. 8. When the Status field indicates that the upload is complete, remove the flash drive. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Static IP address as soon as practical. See Section 2.5.10.3 for configuration details. NUMAVIEW™ REMOTE For remote operation and data capture through an Ethernet connection, please refer to the NumaView™ Remote Software User Guide, PN 08492, available on our website. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Optionally, a recording device such as a strip-chart recorder and/or data logger. (For • electronic documentation, the internal data acquisition system DAS can be used by configuring the Data Logger through the Setup>Data Logging menu, Section 2.5.1). Traceability Standards • 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
5 ppm span gas must be used if the span gas is NOT routed through the dilution system. 4.1.6 SECOND GAS SENSOR OPTIONS CALIBRATION For the optional O sensor, calibrate per Section 4.5 during installation of the instrument. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Sample port, and check the reading after the Stability falls below 1.0 PPB (either in the gas graph or in the Dashboard). Otherwise, follow the steps presented in Sections 4.2.1.1 and 4.2.1.2. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
If incorrect, wait to reach stability, then press the Span button. d. In the Cal Result window, press OK. 3. Press the Stop button and return to Home screen. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Figure 4-2. Zero and Span Calibration Screens Follow the instructions in Section 4.2.1, except instead of the M-P menu, go to the Calibration>Zero menu for Zero cal and to the Calibration>Span menu for CO Span cal. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
(Zero, Low or High), or be disabled. Table 4-1 and Table 4-2 show how to set up the operating states of each calibration or check, and Table 4-3 shows how to program the execution of each. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
High Span check when the Enabled box is also the Calibrate box is unchecked ( disables the High Span calibration and check. Table 4-2 shows how the selection boxes would be enabled/disabled for calibration checks and calibrations. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
COM ports or the contact closure inputs will override any sequence that is in progress. It is recommended that calibrations be performed using • external sources of Zero Air and Span Gas whose accuracy is traceable to EPA standards. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
CPU’s calculation routine. The two offset voltages from the last calibration procedure may be viewed in the Dashboard as Meas Detector and Ref Detector. Activate this procedure in the Utilities>Diagnostics>Dark Calibration menu. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
IMPORTANT IMPACT ON READINGS OR DATA With the O sensor option, subtract the O sensor flow from the measured flow before entering the value in the Flow Cal menu. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Manual Control Valve PUMP Figure 4-4. O Sensor Calibration Set Up SENSOR ZERO GAS: Teledyne API recommends using pure N when calibration the zero point of your O sensor option. SENSOR SPAN GAS: Teledyne API recommends using 20.8% O in N...
To perform a CHECK of the instrument’s Zero or Span Calibration, refer to Sections 4.2.1.1 and 4.2.1.2, respectively. Alternatively, Auto feature described Section 4.2.2.2 with the CALIBRATE attribute set to OFF (not enabled). Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Annually or Perform Verify Leak after any Leak Check Tight Maintenance or Repair Yes if Pneumatic Examine and As Needed lines Clean cleaned Only if cover Cleaning Clean As Needed removed Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
DATE RECORDED OPERATING FUNCTION MODE* STABILITY ZERO CAL CO2 MEAS ZERO CAL ZERO CAL MR RATIO SPAN CAL PRES SAMPLE SAMPLE PHT DRIVE AFTER WARM- SLOPE SPAN CAL OFFSET ZERO CAL Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
• See above, MR Ratio - Zero Cal Increasing Decreasing • See above, MR Ratio - Span Cal Decreasing Increasing SLOPE Span Cal • See above, MR Ratio – Span Cal Increasing Decreasing 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Figure 5-1: Report Generation Page The report can also be set to generate periodically and sent to a Web services “cloud” where it is available for viewing by Teledyne API technical support personnel. Set this function with two Vars: Setup>Vars>Upload Report to Cloud: set to True.
Figure 5-3. Manual Update Page (and other utilities) 3. Insert the flash drive into a front panel USB port and wait for the Status field to indicate that the drive has been detected. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Page 92
6. When complete, as indicated in the Status field, press the Done button, which replaces the Cancel button, and remove the flash drive. 7. Power off and restart the instrument to complete the new firmware installation. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
5. After the Time Zone is implemented first (Steps 1 through 4), then other changes to the date and/or time can be made, and recycling the power is not necessary. Figure 5-4. Time Zone Change Requirements 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
For analyzers equipped with the long-life DFU filter, if and when it needs to be replaced, the DFU is easily accessible for removal and replacement. To replace: 1. Power OFF the analyzer to prevent drawing debris into the instrument. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
CAUTION General Safety Hazard Do not use bubble solution with vacuum applied to the analyzer. The solution may contaminate the instrument. Do not exceed 15 PSIG pressure. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
5.6.6 CLEANING THE OPTICAL BENCH The sensor assembly and optical bench are complex and delicate. Disassembly and cleaning is not recommended. Please check with the factory before disassembling the optical bench. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
It should be noted that if more than two or three warning Alerts occur at the same time, it is often an indication that some fundamental analyzer sub-system (power supply, mainboard) has failed, rather than being an indication of the specific failures referenced by the warning Alerts. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Blocked cooling vents below GFC Assembly. Make sure that adequate clear space beneath the analyzer. WHEEL TEMP The filter wheel temperature is Analyzer’s top cover removed WARNING outside specified limits. Wheel heater Wheel temperature sensor Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
If the box temperature is out of range, check rear panel fan. Areas to the side and rear of instrument should BOX TEMP allow adequate ventilation. See Table 5-4 for BOX TEMP WARNING. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
The following illustrations show connectors and LEDs that can indicate where issues may lie. Figure 5-6 shows the layout for the Mainboard, and Figure 5-7 shows the layout for the Optical Bench Module Smartboard. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
IMPACT ON READINGS OR DATA IMPORTANT Never swap Circuit Gain shunts: use only low gain for the N360 and N360M; never attempt to apply high gain in the N360 or N360M. Figure 5-8. DC Pump Control Board LEDs 5.7.5 FLOW PROBLEMS When troubleshooting flow problems, it is a good idea to first confirm that the actual flow and not the analyzer’s flow sensor and software are in error, or the flow meter is in...
1. Check that the communication/activity LEDs on the pump board (Figure 5-8) are illuminated and/or blinking. If not, check power to board via its test points. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
If no leaks or loose connections are found in the fittings or the gas line between the orifice and the pump, replace the critical flow orifice(s) inside the sample flow control assembly. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
5. Disconnect the exhaust line from the optical bench near the rear of the instrument and plug this line into the SAMPLE inlet creating a pneumatic loop. The CO concentration (either zero or span) now must be constant. If readings become quiet, Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Check the Box Temp signal using the signal voltage in the Utilities>Diagnostics>Analog Inputs menu (Section 5.7.3). This parameter will vary with ambient temperature, but at C (6-7° above room temperature) the signal should be ~1450 mV. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Insufficient time for purging lines upstream of the analyzer. Wait for stability is • reached. Insufficient time for calibration gas source to stabilize. Wait until stability is reached. • Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Problems with RS-232 connections usually center around such things as incorrect connector configuration or incorrect software settings, or improper/incomplete seating of the internal connector. Do not do anything inside the instrument without first contacting Technical Support (Section 5.10). 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Never pull out fuse drawer without ensuring that the Hard Power switch is in OFF position and power cord disconnected, to ensure there is no power to the instrument before checking/changing fuse. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Ensure to carry out a leak check (Section 5.6.4) and a Note recalibration after the analyzer has warmed up for about 60 minutes. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
(“hot-swap”) as this will damage the PCA and require replacement. 5.8.2.1 ONNECTORS ON AINBOARD Modules can be connected to any of the Mainboard’s six CANBUS connectors. Figure 5-12. Mainboard Connectors Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Figure 5-13. DC Pump Control Board Connectors 5.8.2.3 GFC S ONNECTORS ON ENSOR Figure 5-14. GFC Sensor Board Connectors COULD DAMAGE INSTRUMENT AND VOID WARRANTY ATTENTION Never install or remove shunts while instrument is powered ON. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Figure 5-15. Opening the GFC Wheel Housing 6. Remove the two (2) screws holding the opto-pickup printed circuit assembly to the GFC Wheel housing. 7. Carefully remove the opto-pickup printed circuit assembly. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
8. Remove the three (3) screws holding the GFC Wheel motor/heat sink assembly to the GFC Wheel housing. 9. Carefully remove the GFC Wheel motor/heat sink assembly from the GFC Wheel housing. GFC WHEEL HOUSING Figure 5-17. Removing the GFC Wheel Housing 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
12. Follow the previous steps in reverse order to put the GFC Wheel/motor assembly back together. FREQUENTLY ASKED QUESTIONS The following is a list from the Teledyne API’s Technical Support Department of the most commonly asked questions relating to the analyzer. QUESTION...
Page 117
CO measurements are expected above 100 ppm. Call Teledyne API’s Technical Support Department for more information on this topic. This most commonly occurs when an independent metering device is What do I do if the concentration...
Toll-free Phone: +1 800-324-5190 Phone: +1 858-657-9800 Fax: +1 858-657-9816 Email: api-techsupport@teledyne.com Website: http://www.teledyne-api.com/ Before contacting Teledyne API Tech Support, please fill out the problem report form, available online for electronic submission at http://www.teledyne-api.com. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
THEORY OF OPERATION The Model N360 Gas Filter Correlation Carbon Dioxide Analyzer is a microprocessor- controlled analyzer that determines the concentration of carbon dioxide (CO ) in a sample gas drawn through the instrument. It requires that the sample and calibration gases be supplied at ambient atmospheric pressure in order to establish a stable gas flow through the sample chamber where the gases ability to absorb infrared radiation is measured.
(in the case of this analyzer, Carbon Dioxide α is the absorption coefficient that tells how well CO absorbs light at the specific wavelength of interest. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Because water vapor also absorbs light at 4.3 µm, a Gas Filter Correlation (GFC) Wheel is added to the IR light path to overcome the interfering effects of water vapor. Measurement Cell (Pure N Reference Cell with CO Figure 6-2. GFC Wheel 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
IR light. This causes a fluctuation in the intensity of the IR light striking the photo-detector which results in the output of the detector resembling a square wave. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
This linearized concentration value is combined with calibration SLOPE and OFFSET values to produce the CO concentration which is then normalized for changes in sample pressure. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
The ambient CO concentration can rise as high as 1 000 ppm during the time of the day when people are present. It can fluctuate ± 300 Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
A sensor measures the amount of current generated by the feedback control loop which is directly proportional to the concentration of oxygen within the sample gas mixture. Figure 6-7. Oxygen Sensor - Principle of Operation 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
A spring: Applies mechanical force needed to form the seal between the o-rings, the • critical flow orifice and the assembly housing. The STD PID controlled pump option (Figure 5-10) runs on a PID loop flow control that requires no orifice. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
The critical flow orifice used in the analyzer is designed to provide a flow rate per specification (Table 1-1). 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
In order to maintain the proper pressure differential between the inside of the GFC wheel housing and ambient air, the N360 design includes a manually adjustable pressure regulator that maintains the pressure of the purge air feed at 7.5 psig.
If there is an unexpected loss of source power while the instrument is running, when power is restored, it will power up in the last state it was in before power loss. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
GFC Wheel and the sample chamber and just before it reaches the photo-detector to eliminate all black body radiation and other extraneous IR emitted by their various components. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
The resulting light signal tells the analyzer whether the IR beam is shining through the measurement or the reference side of the GFC Wheel. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
SOFTWARE OPERATION The N360/N360M Gas Filter Correlation Carbon Dioxide Analyzer uses special software developed by Teledyne API, which interprets user commands via the various interfaces, performs procedures and tasks, stores data in memory and calculates the concentration of the sample gas.
CO concentration. This is the value that is displayed on the instrument front panel display and is stored in memory by the analyzer’s data acquisition system (DAS) system, the Data Logger. 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
As ambient pressure increases, the compensated CO concentration is decreased. Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 135
DIAG Diagnostics, the diagnostic settings of the analyzer. Disk On Module, a 44-pin IDE flash drive with up to 128MB storage capacity for instrument’s firmware, configuration settings and data 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Page 136
TERM DESCRIPTION/DEFINITION DRAM Dynamic Random Access Memory Data Terminal Equipment EEPROM Electrically Erasable Programmable Read-Only Memory also referred to as a FLASH chip or drive Electro-Static Discharge Ethernet a standardized (IEEE 802.3) computer networking technology for local area networks (LANs), facilitating communication and sharing resources Fluorinated Ethylene Propylene polymer, one of the polymers that Du Pont markets as Teflon ®...
Page 137
VARS Variables, the variable settings of the instrument Voltage-to-Frequency Zero / Span 092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Page 138
Internal Pump, 28, 91, 92, 94, 99, 100, 101, Set Up, 72 122, 123 Internal Zero Air (IZS), 21, 100 Gas Flow Problems, 98 Electrical Connections AC Power, 24 Analog Outputs, 25 Measure Reference Ratio, 119 Communications connections, 27 Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 140
ZERO CAL, 84 SAMPLE PRESS WARN, 94 Zero/Span Valves Internal Pneumatics, 38 Zero/Span/Shutoff Valves Zero Air, 21, 72, 85, 95, 96, 100, 102, 103, 113 Internal Pneumatics, 39 Zero Ar Inlet, 21 Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 141
Bench temperature control duty cycle AI_WHEEL_TEMP Wheel temperature in degree C AI_WHEEL_DUTY_CYCLE Wheel temperature control duty cycle AI_SAMPLE_TEMP Sample temperature in degree C AI_SAMPLE_PRESSURE Sample pressure in PSIA AI_BOX_TEMP Box temperature in degree C 0092910330B DCN8674 Teledyne API – Models N360/N360M CO Analyzers...
Page 142
Detector temp drive in mV AI_PUMP_FLOW Sample flow in CCM AI_ATMOSPHERIC_PRESSURE Atmospheric pressure in PA Holding Registers CO_TARGET_SPAN_CONC_1 CO target span conc for range 1 CO_TARGET_SPAN_CONC_2 CO target span conc for range 2 Teledyne API – Models N360/N360M CO Analyzers 092910330B DCN8674...
Page 143
APPENDIX B - Interconnects, N360 REVISION HISTORY CHASSIS FRONT PANEL GROUND ZONE DESCRIPTION DATE APPROVED INITIAL RELEASE, PROTOTYPE 8/9/2019 019320100, CBL, GRND STRAP 078640000, ASSY, 084260000 THERMISTOR, 12" PCA, MAINBOARD, 089090000 FOR SOM304 ASSY, FAN, 24VDC 078640000, ASSY, 087960100, CBL, 24" CANBUS THERMISTOR, 12"...
Page 144
APPENDIX B - Interconnects, N360M REVISION HISTORY CHASSIS GROUND FRONT PANEL ZONE DESCRIPTION DATE APPROVED INITIAL RELEASE, PROTOTYPE 8/9/2019 019320100, CBL, GRND STRAP 078640000, ASSY, 084260000 THERMISTOR, 12" PCA, MAINBOARD, 089090000 FOR SOM304 ASSY, FAN, 24VDC 078640000, ASSY, 087960100, CBL, 24" CANBUS THERMISTOR, 12"...
Need help?
Do you have a question about the N360 and is the answer not in the manual?
Questions and answers