Technician Symbol: All operations marked with this symbol are to be performed by qualified maintenance personnel only. Electrical Ground: This symbol inside the instrument marks the central safety grounding point for the instrument. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 5
Note that the safety of a system that may incorporate this product is the end user’s responsibility. For Technical Assistance regarding the use and maintenance of this instrument or any other Teledyne API product, contact Teledyne API’s Technical Support Department: Telephone: +1 800-324-5190 Email: api-techsupport@teledyne.com...
être effectuées uniquement par du personnel de maintenance qualifié. Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le point central de la mise à la terre sécuritaire de l’instrument. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 7
Échantillonnage de gaz combustibles Risque d’explosion et d’incendie Techniques de manutention, Blessure corporelle soulevage et de transport inappropriées Notez que la sécurité d’un système qui peut incorporer ce produit est la responsabilité de l’utilisateur final. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
PRODUCT RETURN All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.
Provides information about that which could either affect accuracy of instrument readings or cause loss of data. Provides information pertinent to the proper care, Note operation or maintenance of the instrument or its parts. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Using Events as Triggers for Data Logging ............45 Setup>Dashboard ........................ 46 Setup>AutoCal (with Valve Option) ..................46 Setup>Vars .......................... 47 Setup>Homescreen ......................48 Setup>Digital Outputs (Option) .................... 49 Setup>Analog Outputs (Option) ................... 50 viii Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 11
Service and Troubleshooting ......................82 Fault Diagnosis with Alerts ....................83 Fault Diagnosis With Dashboard Functions ................. 84 Using the Diagnostic Signal I/O Functions ................84 Fault Diagnosis with LEDs ....................84 Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 12
Frequently Asked Questions ......................97 Technical Assistance ........................98 6. PRINCIPLES OF OPERATION ......................99 Pneumatic Operation ........................101 Electronic Operation ........................102 Modules ..........................102 Power Switches........................102 Glossary ............................... 103 Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
INTRODUCTION, SPECIFICATIONS, APPROVALS, & COMPLIANCE Teledyne API’s Model N500 CAPS NO -NO Analyzer uses Cavity-Attenuated Phase-Shift (CAPS) spectroscopy to render direct, true measurements of nitrogen dioxide ). By combining direct NO measurements with highly efficient gas phase titration (GPT), the N500 converts and measures the NO sample gas component as NO , thus allowing for the measurement of total NOx.
Note: All specifications are based on constant conditions EPA DESIGNATION Teledyne API’s Model N500 CAPS NOx analyzer is officially designated as US EPA Automated Equivalent Method, Designation Number EQNA- 0320-256. The official List of Designated Reference and Equivalent Methods is published in the U.S.
Do not operate this instrument without first removing dust plugs from SAMPLE and EXHAUST ports on the rear panel. Note Teledyne API recommends that you store shipping containers and materials for future use if/when the instrument should be returned to the factory for repair and/or calibration service. See...
AREA MINIMUM REQUIRED CLEARANCE 10 cm / 4 in Back of the instrument 2.5 cm / 1 in Sides of the instrument 2.5 cm / 1 in Above and below the instrument Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
The front panel (Figure 2-1) includes two USB ports for peripheral device connections, which can be used with mouse and keyboard as alternatives to the touchscreen interface, or with flash drive for uploads/downloads (devices not included). Figure 2-1. Front Panel Layout Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
For voltage or current loop outputs to a strip chart recorder and/or a data logger. DIGITAL I/O Option For remotely activating the zero and span calibration modes. ETHERNET Connector for network or Internet remote communication, using Ethernet cable. Serial communications port for RS-232. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
AC voltage range and ensure that it is equipped with a functioning earth ground. It is important to adhere to all safety and cautionary messages. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Setup>Analog Outputs menu. Ground Ground Isolated Current Outputs I Out + I Out - I Out + User definable through the Setup>Analog Outputs menu. I Out - I Out + I Out - Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Utilities>Diagnostics>Digital Outputs menu Ground Digital Input1 = Remote Zero Cal Digital Input2 = Remote Span Cal (Digital Input3 not used) View status in Utilities>Diagnostics>Digital Inputs menu Ground Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
DTE RS232 2-3, 5-6, 9-12 232Rx 232Tx View/edit the Communications parameters in the Setup>Comm>COM1 menu. Baud rate: 115200 bits per second (baud) • Data Bits: 8 data bits with 1 stop bit • Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Remove dust plugs from rear panel exhaust and supply line fittings before powering on the instrument. Keep dust plugs for reuse in future storage or shipping to prevent debris from entering the pneumatics. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Run tubing outside the enclosure. • Calibration Gas Sources: The source of calibration gas or zero air is also attached to the SAMPLE inlet, but • only when a calibration operation is actually being performed. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
SAMPLE inlet, but only when a calibration operation is actually being performed. ENTING Vent the output of the calibrator if calibrator not already vented. XHAUST UTLET Attach an exhaust line to the EXHAUST outlet fitting and vent outside the shelter. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Section 2.5 provides setup/customization instructions. TARTUP Upon initial startup, a sequence of status screens (Figure 2-10) appear prior to the Home page (Figure 2-11). Figure 2-10. Status Screens at Startup Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Active Alerts page, a history of all alerts remains in the Utilities>Alerts Log page. Figure 2-12. Viewing Active Alerts Page If Alerts about warning conditions persist after the warm-up period or after being cleared, investigate their cause using the troubleshooting guidelines in Section 5.7.1. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
The method for performing a calibration differs slightly depending on whether or not any of the available internal valve options are installed. Follow the appropriate calibration instructions presented in Section 4. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Time settings*; and check for firmware updates when Section 2.5.7 connected to a network that is connected to the Internet. Time Zone change requires special procedures (Section 5.5). Comm View and configure network and serial communications. Section 2.5.10 Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Figure 2-14 presents an orientation to the main display screen; Figure 2-15 shows that pressing the gas name or its concentration value or a meter below displays a live plot of their respective readings. Section 2.5.6 provides configuration instructions. Figure 2-14. User Interface Orientation Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
2.5.3 provides configuration instructions). If there is a graphing icon in the upper right corner of a parameter, pressing that parameter displays a live plot of its readings as in Figure 2-16. Figure 2-16. Dashboard Page Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
To clear Alerts from the Active Alerts page, either check individual boxes to choose specific Alerts, or check the Select All box to choose all Alerts, then press the Clear Selected button. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Figure 2-19. Utilities>Alerts Log of Active and Past Alerts and Events CALIBRATION The Calibration menu is used for zero/span/multipoint calibrations and for external calibration with valve options installed. Calibration procedures are presented in Section 4. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Used to override the Ozone Generator state when needed, such as for service (Section 5.7.7.3). SETUP The Setup menu is used to configure the instrument’s various features, functions, and data log. Section 2.5 provides details for the menus under Setup. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Sections 2.5.1.1 and 2.5.1.2 providing additional details. To transfer captured instrument data to a flash drive, see Section 2.5.1.3. Figure 2-20. Datalog Configuration, New Log Page Figure 2-21. Datalog Configuration, Existing Log Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Figure 2-22. Creating a New Data Log The parameters available in the list of Log Tags include the names of Events configured in the Events page (Section 2.5.2). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
The Periodic trigger is a timer-based trigger that is used to log data at a specific time interval. Periodic Trigger requires an interval that is set to number of minutes and a start time that is set to date and clock time. Figure 2-23. Datalog Periodic Trigger Configuration Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
2. Select all or define a period from which to download the collected data. 3. Press the Download button, and when complete, as indicated in the Status field, press the Done button (changed from “Cancel”) and remove the flash drive. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Home>Setup>Events menu (Figure 2-26). Press ADD to create a new Event (refer to Figure 2-27 for details), or select an existing Event to either Edit or Delete it (Figure 2-29). Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Event state. Non-latching allows the entry in the Active Alerts screen and the internal Event state to continuously update based on the Event criteria, requiring no user interaction to clear the Alert or Event state). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Events can also be used to create customized triggers for data logging functions. The name entered in the Name field of the Events Configuration page will appear in the list of Log Tags of the Datalog Configuration page. The Data Logger is presented in Section 2.5.1. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
SETUP>DASHBOARD Figure 2-30. Dashboard Display and Configuration SETUP>AUTOCAL (WITH VALVE OPTION) Auto Cal is available for calibration checks with installed valve options (see Section 4.3). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Ozone Lamp Setpoint (Adjustment is dependent on low Titration Efficiency, Section 4.4). System Hours Total system runtime hours Titration Efficiency Used for checking or calibrating efficiency of ozone titrating NO (Section 4.4). Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Setup>Homescreen menu or from Home page using the configuration icon (Figure 2-31). Figure 2-31. Homescreen Configuration An orientation to the Homescreen was presented in Section 2.4.1, including Figure 2-14 and Figure 2-15. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Map the Digital Relays in the same manner. Navigate to the Utilities>Diagnostics>Digital Outputs menu to change the state (ON/OFF) of individual digital outputs. Figure 2-32. Digital Outputs Setup Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Recorder Offset: add a zero offset for recording slightly negative readings from noise around the zero point. • Allow Overrange: check to allow a ± 5% over-range; uncheck to disable over-range if the recording device is sensitive to excess voltage or current. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
4-20 mA 3 mA 21 mA Each range is usable from -5% to +5% of the rated span. For manual calibration adjustments, see Section 2.5.8.1 for voltage and Section 2.5.8.2 for current. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
See Figure 2-4 for pin assignments and diagram of the analog output connector. CAUTION! Do not exceed 60 V peak voltage between current loop outputs and instrument ground. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
In this case, follow the procedure above but adjust the output for the following values: Table 2-10. Current Loop Output Check % FS Voltage across Resistor for 4-20 mA 1000 mVDC 5000 mVDC Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Or choose to turn OFF handshaking. Hardware Error Enable/disable hardware error checking. Checking Enable/disable the hardware First In – First Out (FIFO) for improving data transfer Hardware FIFO rate for that COM port. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
(DHCP). Most users will want to configure the instrument with a static IP address: click the Static radio button to manually assign a static IP address (consult your network administrator and see Table 2-13 for information). Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
192.168.76.1.) that is the address of the computer used by your LAN and serves Gateway as a router to access the Internet or another network. ESSEN Configure Hessen Settings and Gas List (see Section 3.1.2). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
7. When the Status field Indicates that the USB drive has been detected, press the “Upload Configuration to Instrument” Start button. 8. When the Status field Indicates that the upload is complete, remove the flash drive. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
HESSEN Hessen protocol is supported through serial communications. The Hessen protocol is not strictly defined; therefore, while Teledyne API’s application is completely compatible with the protocol itself, it may be different from implementations by other companies. Configure the COM2 port for Hessen protocol through the Setup>Comm>COM2 menu: select COM2 Protocol and press Edit to select HESSEN, then press Accept.
TYPE 2 has more flexibility when operating with instruments that can measure more than • one type of gas. For more specific information about the difference between the two versions, download the Manual Addendum for Hessen Protocol from the Teledyne API's web site. ESSEN...
Resources in JSON format. Tag names and command strings are case sensitive. The Resources are defined in Table 3-3. The Teledyne API REST guide is a tutorial in the form of Service Note 22-002, accessible on our website under this product page’s Downloads tab.
Static IP address as soon as practical. See Section 2.5.10.3 for configuration details. NUMAVIEW™ REMOTE For remote operation and data capture through an Ethernet connection, please refer to the NumaView™ Remote Software User Guide, PN 08492, available on our website. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
See Section 4.2.2 for information regarding setup and calibration of the analyzer with Z/S Valve options. Zero air and span gases must be supplied at twice the Note instrument’s specified gas flow rate. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Alternatively, if a calibrator is available that contains a trusted source of stable ozone, e.g., Teledyne API Model T700U with certified photometer, it is possible to use that O output directly to obtain the NO concentration using GPT.
PERMEATION TUBES Teledyne API offers an optional internal span gas module that utilizes an NO permeation tube as a span gas source. The accuracy of these devices is only about ±5%. Whereas this may be sufficient for quick, daily calibration checks, we recommend using certified NO gases for accurate calibration.
Sample port, and check the reading after the Stability falls below 1.0 PPB (either in the gas graph or in the Dashboard). Otherwise, follow the steps presented in Sections 4.2.1.1 and 4.2.1.2. Figure 4-1. Multi-Point Calibration Page Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Span button. d. In the Cal Result window, press OK. 3. Press the Stop button and return to Home screen. 4. In the Dashboard, check and record the Slope(s) and the Offset(s). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Figure 4-2. Zero and Span Calibration Screens Follow the instructions in Section 4.2.1, except instead of the M-P menu, go to the Calibration>Zero menu for Zero cal and to the Calibration>Span menu for NO Span cal. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Auto Cal sequence. Figure 4-3. Auto Cal Page For each sequence, there are four parameters that control operational details: Date, Time (both in the Start field), Interval, and Duration, as presented in Table 4-1. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
After the instrument stabilizes, perform a TE adjustment in the Vars menu, if required (Setup>Vars>Titrate Eff). This adjusts the calculated NO concentration, which is then added to the measured NO concentration to calculate the corrected NO value. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Government Publishing Office at http://www.gpo.gov/fdsys/) and with Quality Assurance Guidance documents (available on the EPA website: http://www3.epa.gov/ttn/amtic/qalist.html). Give special attention to specific regulations regarding the use and operation of ambient NO analyzers using Cavity Attenuated Phase Shift Spectroscopy method. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
DO NOT press the Zero or Span buttons at the end of each operation (actual calibration), as this will reset the stored values for OFFSET and SLOPE and alter the instrument’s calibration. Alternatively, use the Auto Cal feature described in Section 4.3. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
(set up through the Data Logger, Section 2.5.1). Use NumaView™ Remote to download and review this data from a remote location. The following table, checked weekly, can be used as a basis for taking action as these values change with time. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Figure 5-1: Report Generation Page The report can also be set to generate periodically and sent to a Web services “cloud” where it is available for viewing by Teledyne API technical support personnel. Set this function with two Vars: Setup>Vars>Upload Report to Cloud: set to True.
Figure 5-3. Manual Update Page (and other utilities) 3. Insert a flash drive into a front panel USB port and wait for the Status field to Indicates that the drive has been detected. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Page 77
6. When complete, as indicated in the Status field, press the Done button, which replaces the Cancel button, and remove the flash drive. 7. Power off and restart the instrument to complete the new firmware installation. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
6. After the Time Zone is implemented first (Steps 1 through 5), then other changes to the date and/or time can be made, and recycling the power is not necessary. Figure 5-4. Time Zone Change Requirements Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
REPLACING THE AREF SCRUBBER AND FILTER ASSEMBLY The AREF filter assembly consists of the AREF DFU filter, which is a backup particulate filter, and the charcoal filter, which is an NO gas scrubber. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
6. Disconnect the power connector labeled “pump” located about 4 inches down the black/red cable coming from the top of the pump. 7. Pull pump assembly out and set aside. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
7. Replace the analyzer cover, plug the power cord back in and turn on the analyzer. 8. Carry out a span check to see if the new permeation tube works properly (see Section 4). Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
5. Once the leak has been located and repaired, the leak-down rate of the indicated pressure should be less than 1 psi in 5 minutes after the pressure is turned off. Replace the instrument cover and restart the analyzer. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
5. Once kinks and leaks are ruled out, first replace the sample flow restrictor, and if the sample flow is still out of specification, replace the sample filter. 6. If flow is still out of spec, call Technical Support. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Check for gas flow problems such as clogged or blocked internal/external gas lines, damaged seals, punctured gas lines, damaged / malfunctioning pumps, etc. 5. Follow the procedures defined in Section 2.3.4.3 to confirm that the analyzer’s vital functions are working. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
• Pressure too high at Sample Inlet. This message occurs at power on. • If it is confirmed that power has not been SYSTEM RESET The computer has rebooted. interrupted:Fatal Error caused software to restart • Loose connector/wiring Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Figure 5-8 shows the layout for the mainboard; Figure 5-9 shows the layout for the CAPS DAQ smart board, and Figure 5-10 shows the layout for the ozone tower smart module. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
1. Verify that the expected concentration is set properly to the actual span gas concentration in the CONC sub-menu. 2. Confirm that the NO span gas source is accurate. • This can be done by comparing the source with another calibrated analyzer. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Also, if the span gas is not vented at all and does not supply enough sample gas, the analyzer may be evacuating the sample line. • Ensure to create and properly vent excess span gas. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Wrong materials in contact with sample - use glass, stainless steel or Teflon materials only. • Insufficient time for purging lines upstream of the analyzer. Wait for stability is reached. • Insufficient time for calibration gas source to stabilize. Wait until stability is reached. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
The preceding sections of this manual discussed a variety of methods for identifying possible sources of failures or performance problems within the analyzer. This section describes how to determine if a certain component or subsystem is actually the cause of the problem being investigated. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
NOx valve that isn’t cycling On/Off in NOx mode. A leak in the dryer or between the dryer and the generator can reduce sensitivity and cause performance drift. Carry out a leak check (Section 5.6.5). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
IZS thermistor should be around 2500 mV. RS-232 C OMMUNICATIONS Teledyne API's analyzers use the RS-232 communications protocol to allow the instrument to be connected to a variety of computer-based equipment. Problems with RS-232 connections usually center around such things as incorrect connector configuration, incorrect software settings, improper/incomplete seating of the internal cable.
Regular maintenance procedures are discussed in Section 5.5 Note and are not listed here). Also, there may be more detailed service notes for some of the below procedures. Contact Teledyne API's Technical Support Department. WARNING – ELECTRICAL SHOCK HAZARD Unless the procedure being performed requires the instrument...
Ensure to carry out a leak check (Section 5.6.5) and a recalibration Note after the analyzer has warmed up for about 60 minutes. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
6. Carry out a detailed leak check (see Section 5.6.5.1). 7. Close the analyzer and reconnect the power cord. 8. Power up pump and analyzer and re-calibrate the instrument after it stabilizes. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
FREQUENTLY ASKED QUESTIONS The following list was compiled from the Teledyne API's Technical Support Department’s ten most commonly asked questions relating to the analyzer. QUESTION ANSWER Why does the Enter button Sometimes the Enter button will disappear if you input a setting that is...
Teledyne API Technical Support 9970 Carroll Canyon Road San Diego, California 92131-1106 USA Toll-free Phone: +1 800-324-5190 Phone: +1 858-657-9800 Fax: +1 858-657-9816 Email: api-techsupport@teledyne.com Website: http://www.teledyne-api.com/ Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
“level” in order to derive concentration and compensate for source drift. Using the CAPS technique, the amount of phase shift remains constant for a given concentration, even if the LED drifts over time. This Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
LED OFF ϑ Figure 6-2. Phase Shift Representation of Increased Concentration of NO (Black = LED State, Blue = Light build up in the absence of NO , Green = phase shifted\attenuated light). Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
NO measurement. Pressure transducers monitor the sample pressure and the cell pressure to verify pressure levels, which are used for a variety of important calculations and diagnostics. Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Soft Power switch must be used first as described above. If there is an unexpected loss of source power while the instrument is running, it will power up in the ON state when source power is restored. Teledyne API N500 CAPS NO Analyzer 08977D DCN88743...
Disk Operating System DRAM Dynamic Random Access Memory DR-DOS Digital Research DOS Data Terminal Equipment Electrically Erasable Programmable Read-Only Memory also referred to as a FLASH EEPROM chip or drive Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 106
Prevention of Significant Deterioration Poly-Tetra-Fluoro-Ethylene, a very inert polymer material used to handle gases that PTFE may react on other surfaces; one of the polymers that Du Pont markets as Teflon ® Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 107
Universal Serial Bus: a standard connection method to establish communication between peripheral devices and a host controller, such as a mouse and/or keyboard and a personal computer or laptop VARS Variables, the variable settings of the instrument Voltage-to-Frequency Zero / Span Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
NOX concentration for range 1 during zero/span calibration NOX_CONC NOX concentration for range 1 NOX_STABILITY Concentration stability SV_AREF_LOSS AREF LOSS AI_CELL_PRESSURE Cell Pressure AI_LAMP_CURRENT LAMP CURRENT AI_MEASURED_LOSS MEASURED LOSS AI_ATMOSPHERIC_PRESSURE ATMOSPHERIC PRESSURE Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Page 109
ADDR NAME DESCRIPTION AI_MEASURE_I CURRENT MEASURE AI_SAMPLE_FLOW Sample Flow SV_SERIAL_NUMBER Serial Number Holding Registers NO2_TARGET_SPAN_CONC_1 NO2 Target span concentration for range 1 NOX_TARGET_SPAN_CONC_1 NOX Target span concentration for range 1 Teledyne API N500 CAPS NO Analyzer 08977D DCN8874...
Need help?
Do you have a question about the N500 CAPS NOx and is the answer not in the manual?
Questions and answers