Page 4
NEVER use any combustible/explosive gas with this product! For Technical Assistance regarding the use and maintenance of this instrument or any other Teledyne API product, contact Teledyne API’s Technical Support Department: Telephone: +1 800-324-5190 (toll free) or +1 858-657-9800 Email: api-techsupport@teledyne.com or access any of the service options on our website at http://www.teledyne-api.com/...
Page 5
Si vous utilisez cet instrument d’une autre manière que celle pour laquelle il a été prévu, l’instrument pourrait se comporter de façon imprévisible et entraîner des conséquences dangereuses. NE JAMAIS utiliser de gaz explosive ou combustible avec cet produit! 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Page 6
PRODUCT RETURN All units or components returned to Teledyne API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid.
Page 7
Provides information about that which could either affect accuracy of instrument readings or cause loss of data. Provides information pertinent to the proper care, NOTE operation or maintenance of the instrument or its parts. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
(IR) traveling through a sample chamber filled with gas bearing a varying concentration of CO. The N300 and the N300M use Gas Filter Correlation (GFC) to overcome the interfering effects of various other gases (such as water vapor) that also absorb IR. The analyzer...
3 x isolated digital input controls (fixed) 5 x isolated digital output controls (user-definable) 3 x form C relay alarm outputs (user-definable) Temperature Range 0 - 40°C operating, 0 - 40°C EPA Designation (N300 only) Humidity Range 0-95% RH, Non-Condensing Temp Coefficient <...
<± 0.1% CO Temperature Coefficient <± 0.01% CO /°C Rise and Fall Time <60 seconds to 95% As defined by the USEPA Defined as twice the zero noise level by the USEPA 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
EPA DESIGNATION Teledyne API’s Model N300, Gas Filter Correlation Carbon Monoxide Analyzer, is designated as a US EPA Reference Method, Designation Number RFCA-1093-093 for CO measurement. The official “List of Designated Reference and Equivalent Methods” is published in the U.S. Federal Register.
Do not operate this instrument without first removing dust plugs from SAMPLE and EXHAUST ports on the rear panel. Note Teledyne API recommends that you store shipping containers and materials for future use if/when the instrument should be returned to the factory for repair and/or calibration service. See Warranty statement in this manual and Return Merchandise Authorization (RMA) on our Website at http://www.teledyne-api.com.
Sides of the instrument 2.5 cm / 1 in Above and below the instrument 2.5 cm / 1 in Various rack mount kits are available for this analyzer. Contact Sales for more information. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
The front panel (Figure 2-1) includes two USB ports for peripheral device connections, which can be used with mouse and keyboard as alternatives to the touchscreen interface, or with flash drive for uploads/downloads (devices not included). Figure 2-1. Front Panel Layout 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
For voltage or current loop outputs to a strip chart recorder and/or a data logger. Option ETHERNET Connector for network or Internet remote communication, using Ethernet cable. COM Serial communications port for RS-232 Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Teledyne API recommends that you store shipping containers and materials for future use if/when the instrument should be returned to the factory for repair and/or calibration service.
Setup>Analog Outputs menu. Ground Ground Isolated Current Outputs I Out + I Out - I Out + User definable through the Setup>Analog Outputs menu. I Out - I Out + I Out - 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Outputs menu, and viewable in the Utilities>Diagnostics>Digital Outputs menu Ground Digital Input1 = Remote Zero Cal Digital Input2 = Remote Span Cal (Digital Input3 not used) View status in Utilities>Diagnostics>Digital Inputs menu Ground Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Table 2-5 (view/edit software settings Table 2-14). WARNING – Electrical Shock Hazard Disconnect all power before performing any operation that requires entry into the interior of the analyzer. Figure 2-8. Mainboard JP1 Location and Pin Arrangements 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
• The exhaust from the analyzer’s internal pump MUST be vented outside the shelter or immediate area surrounding the instrument. • It is important to conform to all safety requirements regarding exposure to CO. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Page 31
LEAK CHECK IMPORTANT Run a leak check once the appropriate pneumatic connections have been made; check all pneumatic fittings for leaks using the procedures defined in Section 5.6.3. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
TAPI 700-Series. This type of calibrator precisely mixes a high concentration gas with zero air (both supplied externally) to accurately produce span gas of the correct concentration. Linearity profiles can be automated with this model and run unattended over night. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Instrument Zero Gas SAMPLE Chassis Generator EXHAUST Exhaust must be vented outside of shelter or immediate area surrounding the instrument. Figure 2-10. Pneumatic Connections – Basic Configuration Using Gas Dilution Calibrator 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Zero air is supplied by a zero air generator such as from the TAPI 700-Series, and • Span gas is supplied by a gas dilution calibrator such as from the TAPI 700-Series. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Zero air is supplied via a zero air generator such as from the TAPI 700-Series. • An adjustable valve is installed in the zero air supply line to regulate the gas flow. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
VENT/SPAN psig) AIR/IZS TAPI 700-Series Exhaust must be Zero Gas vented outside of EXHAUST Generator shelter or immediate area surrounding the instrument. Figure 2-12. Pneumatic Connections – Ambient Zero/Pressurized Span Option Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Page 37
NIST-SRM gas) to the SPAN inlet at 30 psig. • Zero air is supplied via a zero air generator such as a Teledyne API’s T701. • An adjustable valve is installed in the zero air supply line to regulate the gas flow.
No more than 2m long and; • • Vented outside the shelter or immediate area surrounding the instrument. A similar vent line should be connected to the VENT SPAN outlet on the back of the analyzer. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Exhaust must be vented outside EXHAUST of shelter or immediate area surrounding the instrument. Figure 2-14. N300 Pneumatic Connections – Zero/Span Option AMPLE OURCE Attach a sample inlet line to SAMPLE inlet port. The sample input line should not be more than 2 meters long.
Page 40
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line should PTFE tubing; minimum O.D ¼”; • A maximum of 10 meters long; • Vented outside the analyzer’s enclosure. • Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
ASIC ONFIGURATION INSTRUMENT CHASSIS SAMPLE G AS INLET EXHAUST GAS OUTLET WHITE CELL PUMP PRESS URE PCA PUMP CONTROLLER SENSO R FLOW METER Figure 2-15. N300/N300M Internal Gas Flow (Basic Configuration) 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
PCA PUMP CONTROLLER EX HA UST PRESS URE GAS OUTLET SENSO R PUMP FLOW METER Figure 2-16. N300 Internal Pneumatic Flow for Zero/Span Valves Option Table 2-6. Zero/Span Valve Option Operating States MODE VALVE CONDITION SAMPLE Sample/Cal Open to SAMPLE inlet...
EX HA UST GAS OUTLET PRESS URE SENSO R PUMP FLOW METER Figure 2-17. N300 Internal Pneumatic Flow for Ambient Zero/Pressurized Span Option Table 2-7. Operating States for Ambient Zero/Pressurized Span Option MODE VALVE CONDITION Sample/Cal Open to SAMPLE inlet...
Figure 2-18. Status Screens at Startup Upon any startup, this instrument should warm up for at least one hour to allow the pump, heaters, IR source, and detector to stabilize, and to provide reliable measurements. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Figure 2-20. Viewing Active Alerts Page If Alerts about warning conditions persist after the warm up period or after being cleared, investigate their cause using the troubleshooting guidelines in Section 5.7. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
The method for performing a calibration differs slightly depending on whether or not any of the available internal zero air or valve options are installed. Follow the appropriate calibration instructions in Section 4. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Time settings*; and check for firmware updates when Section 2.5.9 connected to a network that is connected to the Internet. Time Zone change requires special procedures (Section 5.5). Comm View and configure network and serial communications. Section 2.5.10 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Figure 2-22 presents an orientation to the main display screen; Figure 2-23 shows that pressing the gas name or its concentration value or a meter below displays a live plot of their respective readings. Section 2.5.6 provides configuration instructions. Figure 2-22. User Interface Orientation Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
(Section 2.5.3 provides configuration instructions). If there is a graphing icon in the upper right corner of a parameter, pressing that parameter displays a live plot of its readings as in Figure 2-24. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
(Photo-detector temperature control voltage) The drive voltage being supplied to the PHT Drive thermoelectric coolers of the IR photo-detector. /min Sample mass flow rate as measured by the flow rate sensor in the sample gas stream. Pump Flow Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
To clear Alerts from the Active Alerts page, either check individual boxes to choose specific Alerts, or check the Select All box to choose all Alerts, then press the Clear Selected button. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
The Calibration menu is used for multipoint calibrations as well as calibrations for zero and span with valve options, and for external calibration with valve options installed. Calibration procedures are presented in Section 4. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
(See Section 4.4.3). 2.4.6 SETUP The Setup menu is used to configure the instrument’s various features, functions, and data log. Section 2.5 provides details for the menus under Setup. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Figure 2-30 provides illustrated instructions for setting up a new log, with Sections 2.5.1.1 and 2.5.1.2 providing additional details. To transfer captured instrument data to a flash drive see Section 2.5.1.3. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Figure 2-30. Creating a New Data Log The parameters available in the list of Log Tags include the names of Events configured in the Events page (Section 2.5.2). Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Home>Setup>Events menu (Figure 2-34). Press ADD to create a new Event (refer to Figure 2-35 for details), or select an existing Event to either Edit or Delete it (Figure 2-37). Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Active Alerts screen and the internal Event state to continuously update based on the Event criteria, requiring no user interaction to clear the Alert or Event state). Figure 2-36. Configured Event Sample 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Name field of the Events Configuration page will appear in the list of Log Tags of the Datalog Configuration page. The Data Logger is presented in Section 2.5.1. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Access the menu to see the list of variables; select a variable to view its description; touch the Edit button to change its setting(s). 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Total system runtime hours TPC Enable Enables or disables the Temperature and Pressure Compensation (TPC) feature (For information on TPC, see Section 6.6.4). User Units Change the concentration units of measure. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Setup>Homescreen menu or from Home page using the configuration icon (Figure 2-39). Figure 2-39. Homescreen Configuration An orientation to the Homescreen was presented in Section 2.4.1, including Figure 2-22 and Figure 2-23. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Map the user-configurable Analog Outputs for either voltage output (Figure 2-41) or current output (Figure 2-42) to any of a wide variety of “Signals” present in the instrument and customize their respective configurations. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
(Table 2-12). • For Current output, press the +100 button several times to get the setting close to 4mA. 6. Press the Accept button when adjustment reached. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Each range is usable from -5% to +5% of the rated span. For manual calibration adjustments, see Section 2.5.8.1 for voltage and Section 2.5.8.2 for current. Figure 2-43. Analog Output Calibration, Voltage or Current 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
+100 button to arrive at a realistic starting point for 4mA. CAUTION! Do not exceed 60 V peak voltage between current loop outputs and instrument ground. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
When an instrument is connected to a network that is connected to the Internet, follow the instructions on this Remote Update page to check for and activate software/firmware updates. (Also refer to Section 5.3). 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Enable/disable the requirement for a password for this serial port to respond. The Security only command that is active is the request-for-help command (? CR). Stop bits Select either 0 or 1 stop bit (typically set in conjunction with Parity and Data bits). Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
A string of numbers very similar to the Instrument IP address (e.g. Default 192.168.76.1.) that is the address of the computer used by your LAN and serves Gateway as a router to access the Internet or another network. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
7. When the Status field indicates that the USB drive has been detected, press the “Upload Configuration to Instrument” Start button. 8. When the Status field indicates that the upload is complete, remove the flash drive. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Static IP address as soon as practical. See Section 2.5.10.3 for configuration details. NUMAVIEW™ REMOTE For remote operation and data capture through an Ethernet connection, please refer to the NumaView™ Remote Software User Guide, PN 08492, available on our website. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Optionally, a recording device such as a strip-chart recorder and/or data logger. • (For electronic documentation, the internal data acquisition system DAS can be used by configuring the Data Logger throught the Setup>Data Logging menu, Section 2.5.1). Traceability Standards • Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
4.1.7 DATA RECORDING DEVICES A strip chart recorder, data acquisition system or digital data acquisition system should be used to record data from the serial or analog outputs of the N300/N300M. • If analog readings are used, the response of the recording system should be checked against a NIST traceable voltage source or meter.
Wait for reading to stabilize, a. Wait for reading to stabilize. then press Stop. b. Press the Zero button. b. Check and record the c. Press Stop and check the reading. reading. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
3. Press the Stop button and return to Home screen. 4. In the Dashboard, check and record the Slope(s) and the Offset(s). (See Table 4-4 in Section 4.3 for expected/acceptable values). Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Figure 4-2. Zero and Span Calibration Screens Follow the instructions in Section 4.2.1, except instead of the M-P menu, go to the Calibration>Zero menu for Zero cal and to the Calibration>Span menu for CO Span cal. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
(Zero, Low or High), or be disabled. Table 4-1 and Table 4-2 show how to set up the operating states of each calibration or check, and Table 4-3 shows how to program the execution of each. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
High Span check when the Enabled box is also the Calibrate box is unchecked ( disables the High Span calibration and check. Table 4-2 shows how the selection boxes would be enabled/disabled for calibration checks and calibrations. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
COM ports or the contact closure inputs will override any sequence that is in progress. It is recommended that calibrations be performed using • external sources of Zero Air and Span Gas whose accuracy is traceable to EPA standards. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Before performing the pressure calibration procedure, power off the sample gas pump through the software only, and disconnect the sample gas-line vent from the rear panel sample gas inlet. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
CALIBRATION OF OPTIONAL SENSORS This Section provides the calibration setup and procedures for the O Sensor and the CO Sensor options. 4.5.1 O SENSOR CALIBRATION Set up the O pneumatics connections as follows: Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Manual Control Valve PUMP Figure 4-4. O Sensor Calibration Set Up SENSOR ZERO GAS: Teledyne API recommends using pure N when calibration the zero point of your O sensor option. SENSOR SPAN GAS: Teledyne API recommends using 20.8% O in N...
(default factory setting = 12%). EPA PROTOCOL CALIBRATION When running the N300 for U.S. EPA compliance (the N300M does not have an EPA designation), always calibrate prior to use, adhering to the EPA designation requirements for this instrument.
To perform a CHECK of the instrument’s Zero or Span Calibration, refer to Sections 4.2.1.1 and 4.2.1.2, respectively. Alternatively, Auto feature described Section 4.2.2.2 with the CALIBRATE attribute set to OFF (not enabled). 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Annually or Perform Verify Leak after any Leak Check Tight Maintenance or Repair Yes if Pneumatic Examine and As Needed lines Clean cleaned Only if cover Cleaning Clean As Needed removed Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
DATE RECORDED OPERATING FUNCTION MODE* STABILITY ZERO CAL CO MEAS ZERO CAL ZERO CAL MR RATIO SPAN CAL PRES SAMPLE SAMPLE PHT DRIVE AFTER WARM- SLOPE SPAN CAL OFFSET ZERO CAL Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
• See above, MR Ratio - Zero Cal Increasing Decreasing • See above, MR Ratio - Span Cal Decreasing Increasing SLOPE Span Cal • See above, MR Ratio – Span Cal Increasing Decreasing Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Figure 5-1: Report Generation Page The report can also be set to generate periodically and sent to a Web services “cloud” where it is available for viewing by Teledyne API technical support personnel. Set this function with two Vars: Setup>Vars>Upload Report to Cloud: set to True.
Figure 5-3. Manual Update Page (and other utilities) 3. Insert the flash drive into a front panel USB port and wait for the Status field to indicate that the drive has been detected. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Page 93
6. When complete, as indicated in the Status field, press the Done button, which replaces the Cancel button, and remove the flash drive. 7. Power off and restart the instrument to complete the new firmware installation. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
5. After the Time Zone is implemented first (Steps 1 through 4), then other changes to the date and/or time can be made, and recycling the power is not necessary. Figure 5-4. Time Zone Change Requirements Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
To change the filter: 1. Turn OFF the analyzer to prevent drawing debris into the instrument. 2. Open the N300 Analyzer’s hinged rear panel and unscrew the knurled retaining ring on the filter assembly. Figure 5-5. Sample Particulate Filter Assembly 3.
4. If the reading is < 10 in-Hg, the pump is in good condition and there are no large leaks. 5. Check the sample gas flow. If the flow is <10 cm /min and stable, there are no large leaks in the instrument’s pneumatics. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
2. Sample flow should be 800 cc/min ± 10%. 3. Once an accurate measurement has been recorded by the method described above, adjust the analyzer’s internal flow sensors (See Section 4.4.3). 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
High flows indicate leaks downstream of the flow control assembly. 5.6.6 CLEANING THE OPTICAL BENCH The N300/N300M sensor assembly and optical bench are complex and delicate. Disassembly and cleaning is not recommended. Please check with the factory before disassembling the optical bench.
M/R ratio. This message occurs at power on. If you have not cycled the The computer has rebooted. SYSTEM RESET power on your instrument: Fatal error caused software to restart 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Check for gas flow problems (see Section 5.7.5). SAMPLE SAMPLE TEMP should be close to BENCH TEMP. Temperatures outside of the specified range or oscillating TEMP temperatures are cause for concern. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Use the Utilities>Diagnostics menu to view the raw voltage of an input signal or the Setup menu to control the state of an output voltage or control signal. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
5.7.4 FAULT DIAGNOSIS WITH LEDs The following illustrations show smartboard components that can be useful in determining fault conditions. Figure 5-6. Mainboard Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Use an independent, calibrated volumetric flow meter to perform a flow check as described in Section 5.6.5. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
The N300/N300M has one main gas flow path. With a valve option installed, there are several subsidiary paths but none of those are displayed on the front panel or stored by the data acquisition system. With the O sensor option installed, a third gas flow controlled with a critical flow orifice is added, but this flow is not measured or reported.
If no leaks or loose connections are found in the fittings or the gas line between the orifice and the pump, replace the critical flow orifice(s) inside the sample flow control assembly. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
The pump module may need to be replaced if the reading is above 10 in-Hg. If the Sample Flow reading is greater than 10 cc/min, there is a leak in the pneumatic lines. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
1. Bad span gas. This can cause a large error in the slope and a small error in the offset. Delivered from the factory, the N300 Analyzer’s slope is within ±15% of nominal. Bad span gas will cause the analyzer to be calibrated to the wrong value. If in doubt have the span gas checked by an independent lab.
Inputs menu (Section 5.7.3). This parameter will vary with ambient temperature, but at C (6-7° above room temperature) the signal should be ~1450 mV. AMPLE EMPERATURE The Sample Temperature should closely track the bench temperature. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
(included in the Final Test and Validation Data Sheet shipped with the unit from Teledyne API). The most common cause of excessive noise is leaks, and running a leak check and flow check (Sections 0 and 5.6.5) will give an indication.
Problems with RS-232 connections usually center around such things as incorrect connector configuration, incorrect software settings, or improper/incomplete seating of the internal connector. Do not do anything inside the instrument without first contacting Technical Support. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Never pull out fuse drawer without ensuring that the Hard Power switch is in OFF position and power cord disconnected, to ensure there is no power to the instrument before checking/changing fuse. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
COULD DAMAGE INSTRUMENT AND VOID WARRANTY ATTENTION Never connect or disconnect PCA electrical connectors while the instrument is powered on (“hot-swap”) as this will damage the PCA and require replacement. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
CONNECTORS ON MAINBOARD Modules can be connected to any of the Mainboard’s six CANBUS connectors. Figure 5-12. Mainboard Connectors 5.8.2.2 CONNECTORS ON DC PUMP PCA Figure 5-13. DC Pump Control Board Connectors 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
IMPORTANT Never swap Circuit Gain shunts: high gain is only for the N300 and N300U; low gain is only for the N300M. 5.8.3 REMOVING/REPLACING THE GFC WHEEL When removing or replacing the GFC Wheel it is important to perform the disassembly in the following order to avoid damaging the components: 1.
Figure 5-15. Opening the GFC Wheel Housing 6. Remove the two (2) screws holding the opto-pickup printed circuit assembly to the GFC Wheel housing. 7. Carefully remove the opto-pickup printed circuit assembly. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
8. Remove the three (3) screws holding the GFC Wheel motor/heat sink assembly to the GFC Wheel housing. 9. Carefully remove the GFC Wheel motor/heat sink assembly from the GFC Wheel housing. GFC WHEEL HOUSING Figure 5-17. Removing the GFC Wheel Housing Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
12. Follow the previous steps in reverse order to put the GFC Wheel/motor assembly back together. FREQUENTLY ASKED QUESTIONS The following is a list from the Teledyne API’s Technical Support Department of the most commonly asked questions relating to the N300/N300M CO Analyzer. QUESTION...
Page 118
CO measurements are expected above 100 ppm. Call Teledyne API’s Technical Support Department for more information on this topic. This most commonly occurs when an independent metering device is What do I do if the concentration...
Toll-free Phone: +1 800-324-5190 Phone: +1 858-657-9800 Fax: +1 858-657-9816 Email: api-techsupport@teledyne.com Website: http://www.teledyne-api.com/ Before contacting Teledyne API Tech Support, please fill out the problem report form, available online for electronic submission at http://www.teledyne-api.com. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
6. THEORY OF OPERATION The N300/N300M Gas Filter Correlation Carbon monoxide Analyzer is a microprocessor-controlled analyzer that determines the concentration of carbon monoxide (CO) in a sample gas drawn through the instrument. It requires that the sample and calibration gases be supplied at ambient atmospheric pressure in order to establish a stable gas flow through the sample chamber where the gases ability to absorb infrared radiation is measured.
(Euler’s number) is the absorption path, or the distance the light travels as it is being absorbed. is the concentration of the absorbing gas (in the case of the N300/N300M, Carbon Monoxide (CO)). α...
Each cavity is mainly filled with composed gases. One cell is filled with pure N (the measurement cell) The other is filled with a combination of N and a high concentration of CO (the reference cell). Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
IR Measure to be more sensitive to the presence of CO in the sample chamber than IR Reference and the ratio between them (M/R) to move closer to 1:1 as the concentration of CO in the sample chamber increases. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Measure and IR Reference. In effect, the difference between the peak heights remains the same. is Shifted IR shining through both cells is affected equally by interfering gas in the Sample Chamber Figure 6-5. Effects of Interfering Gas on IR Measure and IR Reference Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
NTERFERENCE EJECTION The basic design of the N300/N300M rejects most of this interference at a 300:1 ratio. The two primary methods used to accomplish this are: The 4.7μm band pass filter just before the IR sensor which allows the instrument to •...
Normally the analyzer is operated with its inlet near ambient pressure either because the sample is directly drawn at the inlet or a small vent is Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
A thermal-mass flow sensor is used to measure the sample flow through the analyzer. The sensor is calibrated at the factory with ambient air or N , but can be calibrated to operate with samples consisting of other gases such as CO. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
ON state when source power is restored. 6.5.3 OPTICAL BENCH & GFC WHEEL Electronically, in the case of the optical bench for the N300 Analyzer, GFC Wheel and associated components do more than simply measure the amount of CO present in the sample chamber.
GFC Wheel and the sample chamber and just before it reaches the photo-detector to eliminate all black body radiation and other extraneous IR emitted by their various components. Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
The resulting light signal tells the analyzer whether the IR beam is shining through the measurement or the reference side of the GFC Wheel. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
SOFTWARE OPERATION The N300/N300M Gas Filter Correlation Carbon Monoxide Analyzer uses special software developed by Teledyne API, which interprets user commands via the various interfaces, performs procedures and tasks, stores data in memory and calculates the concentration of the sample gas.
This TPC feature multiplies the analyzer’s CO concentration by a factor that is based on the difference between the ambient pressure of the sample gas normalized to standard atmospheric pressure. As ambient pressure increases, the compensated CO concentration is decreased. 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Page 134
Diagnostics, the diagnostic settings of the analyzer. Disk On Module, a 44-pin IDE flash drive with up to 128MB storage capacity for instrument’s firmware, configuration settings and data DRAM Dynamic Random Access Memory Data Terminal Equipment Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Page 135
Poly Vinyl Chloride, a polymer used for downstream tubing RS-232 specification and standard describing a serial communication method between DTE (Data Terminal Equipment) and DCE (Data Circuit-terminating Equipment) devices, using a maximum cable-length of 50 feet 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Page 136
VARS Variables, the variable settings of the instrument Voltage-to-Frequency Zero / Span Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Page 137
Wheel temperature control duty cycle AI_SAMPLE_TEMP Sample temperature in degree C AI_SAMPLE_PRESSURE Sample pressure in PSIA AI_BOX_TEMP Box temperature in degree C AI_PHOTO_TEMP_RAW Detector temp drive in mV AI_PUMP_FLOW Sample flow in CCM 092910300A DCN8475 Teledyne API – Models N300/N300M CO Analyzers...
Page 138
ADDRESS DESCRIPTION AI_ATMOSPHERIC_PRESSURE Atmospheric pressure in PA Holding Registers CO_TARGET_SPAN_CONC_1 CO target span conc for range 1 CO_TARGET_SPAN_CONC_2 CO target span conc for range 2 Teledyne API – Models N300/N300M CO Analyzers 092910300A DCN8475...
Page 139
APPENDIX B - N300 Interconnects REVISION HISTORY CHASSIS GROUND FRONT PANEL ZONE DESCRIPTION DATE APPROVED 094140100 OPTION, INITIAL RELEASE, PROTOTYPE 8/9/2019 094150000, OPTION, PRES SPAN, AMB ZERO CAL, VALVE 094140000, OPTION Z/S VALVES 019320100, CBL, GRND STRAP 094160000, OPTION IZS VALVE...
Page 140
APPENDIX B - N300M Interconnects REVISION HISTORY CHASSIS GROUND FRONT PANEL ZONE DESCRIPTION DATE APPROVED INITIAL RELEASE, PROTOTYPE 8/9/2019 019320100, CBL, GRND STRAP 078640000, ASSY, 084260000 THERMISTOR, 12" PCA, MAINBOARD, 089090000 FOR SOM304 ASSY, FAN, 24VDC 078640000, ASSY, 087960100, CBL, 24" CANBUS THERMISTOR, 12"...
Need help?
Do you have a question about the N300 and is the answer not in the manual?
Questions and answers