NEVER use any gas analyzer to sample combustible gas(es). Note Technical Assistance regarding the use and maintenance of the T320/T320U or any other Teledyne API product can be obtained by contacting Teledyne API’s Technical Support Department: Phone: 800-324-5190 Email: sda_techsupport@teledyne.com...
Page 4
CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est représentée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent dans ce manuel et à...
3.4.2. Internal Pneumatic Flow ............................11 3.4.3. Pneumatic Connections to T320/T320U Basic Configuration: ................12 3.4.4. Pneumatic Connections to T320/T320U with Internal Valve Options Installed ............13 3.4.5. Pneumatic Connections to T320/T320U in Multipoint Calibration Applications ............14 4. FREQUENTLY ASKED QUESTIONS & GLOSSARY ......................15 5.
Page 6
TABLE OF CONTENTS Teledyne API T320/T320U Addendum to T300/T300M Manual LIST OF APPENDICES APPENDIX A – MENU TREES and SOFTWARE DOCUMENTATION APPENDIX B –SPARE PARTS LIST APPENDIX C - REPAIR QUESTIONNAIRE APPENDIX D - SCHEMATICS 07406B DCN7347...
(N O), respectively. The primary difference between the models is their maximum operating range; 1000 ppm for the T320 and 200 ppm for the T320U. The T320 and T320U are designed to be used for monitoring background levels of N O in the atmosphere, byproducts of combustion products and contamination of bottled gases.
Page 8
Introduction Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
SPECIFICATIONS, APPROVALS AND WARRANTY This section provides the specifications for the T320 and the T320U analyzers, and references approvals and warranty. 2.1. SPECIFICATIONS Specifications differ between the T320 and the T320U analyzers for Ranges, Noise, Lower Detectable Limit and Drift.
There is no US EPA reference method for the measurement of nitrous oxide; therefore, neither the T320 nor the T320U are designated as reference or equivalent methods. Hence, the EPA Protocol Calibration section of the Model T300/T300M Operation Manual does not apply.
The analyzer should be started and allowed to warm up; a functional check followed by calibration should be performed. The process for starting and warming up the T320/T320U is identical to that described in the Initial Operation section of the T300/T300M Operators Manual (P/N 06864).
In this case, N O measurements made with the T320/T320U analyzer, it is recommended that you use a span gas with a N O concentration equal to 80% of the measurement range for your application.
Teledyne API T320/T320U Addendum to T300/T300M Manual Getting Started 3.4.2. INTERNAL PNEUMATIC FLOW The T320 internal pneumatic flow is illustrated in Figure 3-1. The T320U internal pneumatic flow is illustrated in Figure 3-2. INSTRUMENT CHASSIS SAMPLE GAS Particulate Filter INLET...
Teledyne API T320/T320U Addendum to T300/T300M Manual 3.4.3. PNEUMATIC CONNECTIONS TO T320/T320U BASIC CONFIGURATION: The information found in the Pneumatic Connections section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320/T320U with the following changes: • Synthetic air, ultra-zero air or N should be used as the zero gas rather than a Zero Gas Generator like the TAPI Model 701.
3.4.4. PNEUMATIC CONNECTIONS TO T320/T320U WITH INTERNAL VALVE OPTIONS INSTALLED • The following figures show the pneumatic set up for T320/T320U analyzers with one of the three available internal valve options installed. For more information on these options see Section 5 of this addendum.
We recommend using high-concentration, certified, calibration gas supplied to the analyzer through a gas dilution calibrator such as a Teledyne API T700. This type of calibrator precisely mixes span gas and zero air to produce max concentration levels between 0 ppm and the concentration of the certified gas. This means that both the source of zero air and span gas must be connected to the calibrator whose output is then connected to the span inlet on the rear panel of the instrument.
FREQUENTLY ASKED QUESTIONS & GLOSSARY The information found in the Frequently Asked Questions section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320/T320U. 07406B DCN7347...
Page 18
Frequently Asked Questions & Glossary Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
OPTIONAL HARDWARE AND SOFTWARE Not all options are available for the T320 and T320U as are available for the T300 and T300M. 5.1. CALIBRATION VALVE OPTIONS FOR THE T320/T320U There three valve options available for the T320 and T320U analyzers Table 5-1.
Optional Hardware and Software Teledyne API T320/T320U Addendum to T300/T300M Manual 5.1.1. ZERO/SPAN WITH COMMON SHUTOFF VALVES (OPTION 50G) Option 50G is operationally and pneumatically similar to Option 50B (see T300/T300M manual), except that both the zero and span gases are applied to the analyzer under pressure. This option is designed to be used with bottled zero and span gases.
6.1. TEST FUNCTIONS The information found in the Test Functions section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320/T320U with the following exception(s): • The following table supersedes Test Functions Defined table of the T300/T300M Operators Manual (P/N 06864).
6.1.1. SELECTING A TEST CHANNEL FUNCTION FOR OUTPUT A4 This section supplements the corresponding section of the T300/T300M Operators Manual (P/N 06864) The Test Functions available to be reported on analog output A4 are: Table 6-2. Test Channels Functions Available on the T320/T320U’s Analog Output TEST CHANNEL DESCRIPTION...
ADVANCED FEATURES The information found in the Advanced Features section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320 and T320U we recommend that you read that section before continuing. 07406B DCN7347...
Page 24
Advanced features Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
The information found in the Hessen Protocol Gas ID section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320/T320U with the following exception(s): • There is only one default gas type programmed into the T320/T320U: the Hessen Gas ID for N O is 320. 07406B DCN7347...
Page 26
Remote Operation Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
T300/T300M Manual - P/N 06864 with the following notes and exceptions: • Delivering span and zero gases for the higher resolution the T320/T320U can be difficult. Attention must be paid to the quality of the gases, the level of contaminants in the gases as well as the history and conditioning of the gas delivery components.
Page 28
Calibration Procedures Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
EPA PROTOCOL CALIBRATION The information found in the EPA Protocil Calibration section of the T300/T300M Operators Manual (P/N 06864) does not apply to the T320/T320U as there are no USEPA reference methods for the measurement of N 07406B DCN7347...
Page 30
EPA Protocol Calibration Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
PRINCIPLES OF OPERATION The information found in the Principles of Operation section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320 and T320U (we recommend that you read this section before continuing) with the following exception: The wavelength of operation is 4.58µm versus 4.7µm for the T300/T300M.
Page 32
Principles OF OPERATION Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
The information found in the Maintenance section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320/T320U with the following exception(s): • Since a catalytic zero scrubber is not available for the T320 or T320U there is no replacement schedule for the scrubber material. •...
Page 34
Maintenance Schedule Teledyne API T320/T320U Addendum to T300/T300M Manual This page intentionally left blank. 07406B DCN7347...
TROUBLESHOOTING & SERVICE The information found in the Troubleshooting and Service section of the T300/T300M Operators Manual (P/N 06864) is applicable to the T320 and T320U. It is recommended that you read that section before continuing. 07406B DCN7347...
Page 36
Troubleshooting & Service Teledyne API T320/T320U Addendum to T300/T300M Manual P/N 06864 This page intentionally left blank. 07406B DCN7347...