Measurement Accuracy - National Instruments SCC-TC Series User Manual

Thermocouple input modules
Hide thumbs Also See for SCC-TC Series:
Table of Contents

Advertisement

Amplifier Characteristics
Transfer Characteristics

Measurement Accuracy

Thermocouple
Type
B
E
J
Temperature range is 23 °C ±5 °C.
1
2
Temperature range is 0 to 50 °C.
Total system measurement error for operating temperature within ±5 °C of calibration temperature. Includes
3
PCI/AT-MIO-16XE-50 one-year accuracy specification of 0.01% ±412 µV. SCC-TC01/02 accuracy specification of 0.08%
±5 µV, and reference junction measurement accuracy of 0.5 °C. Assumes averaging. Non-averaged, single-point reading has
an additional uncertainty (up to ±0.1 °C for J-type thermocouple).
SCC-TC Series Thermocouple Input Modules User Guide
Download from Www.Somanuals.com. All Manuals Search And Download.
Open thermocouple detection current.....250 nA max
Common-mode rejection ratio................110 dB min
Output range ...........................................±10 V max
Gain ........................................................100
Gain error................................................±0.08% max
Gain stability...........................................±0.0005%/°C max
Offset error .............................................±5 µV max (post calibration)
Offset stability ........................................±0.6 µV/°C max
Nonlinearity ............................................±0.004% max
Recommended warm-up time.................5 minutes
3
Temperature
Range (°C)
400 to 600
600 to 1,800
–200 to –100
–100 to 600
600 to 1,000
–200 to –100
–100 to 500
500 to 1,100
(assuming floating thermocouple)
Maximum
Inaccuracy (±°C)
2.5
2
2
1
1.5
2
1
1.5
14
1
1
1
1
2
1
2
1
Typical
Inaccuracy (±°C)
0.31
0.13
1.76
1
0.6
1.6
0.9
0.72
ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Scc-tc01Scc-tc02

Table of Contents