Adding the product to Bronkhorst FlowSuite
When Bronkhorst FlowSuite is started, the software will
automatically search for connected products. All found products
are displayed in the left pane.
▶ Use the refresh option to check for newly connected products.
If the product is not found, it may not be set to serial RS-232.
▶ Follow the instructions on the next page to set the product
to serial RS-232.
Commissioning
▶ Apply fluid pressure gradually to prevent pressure shocks [1].
▶ Warm up the product for at least 30 minutes before starting
measurement [2].
Adjusting Zero Point
To ensure correct measurements, adjust Zero Point in the following
situations: before first use and after remounting [3].
To start adjusting Zero Point:
▶ Open Bronkhorst FlowSuite on the PC.
▶ Select the product in the left pane.
▶ Go to the Actions menu.
▶ Press the button Auto Zero.
▶ Follow the on-screen instructions.
The product is ready for use!