▶ If in any doubt: scan the QR code (Service & Support) on the The product requires professional experience and knowledge of back cover to get in contact with Bronkhorst. installation, operational requirements of the relevant industry, and knowledge of the type of application and national requirements ▶...
▶ If in any doubt: scan the QR code (Service & Support) on the product in a non-critical process. It guides the user in controlling back cover to get in contact with Bronkhorst. and monitoring research and pilot setups by means of The user is responsible for taking the necessary safety measures Bronkhorst FlowSuite.
Page 4
FLEXI-FLOW™ Compact FF-C1xS 1. Fluid inlet 5. Status LED (controller) 9. Fluid outlet 2. USB-C port 6. Control valve 10. Serial number label 3. Address selector 7. Status LED (shut-off valve) 4. Connector 8. Shut-off valve...
Page 5
FLEXI-FLOW™ Compact FF-C1xS (ethernet) 1. Fluid inlet 5. Status LED (controller) 9. Shut-off valve 2. USB-C port 6. Address selectors 10. Fluid outlet 3. Ethernet ports 7. Status LED (shut-off valve) 11. Serial number label 4. Connector 8. Control valve...
5. +Us shut-off (24 Vdc)** ▶ Scan the QR code (Service & Support) on the back cover. 6. D1 Modbus (B/B’) or RS485-A ▶ Go to the Bronkhorst website and select your operating 7. +Us (24 Vdc) system to download Bronkhorst FlowSuite. For download 8.
Mounting location (general) To ensure reliable measurements, it is important to choose a correct mounting location. For optimal performance, observe the following guidelines: ▶ Install the product on a flat, rigid, and clean surface [1]. ▶ Avoid installation close to mechanical vibrations [2]. ▶...
Fluid line connection Electrical connection For basic setup, it is recommended to use Bronkhorst® CAUTION accessories (not included) as described in this manual. Risk of personal injury due to fluid leakage! ▶ Ensure all address selectors on the product are set to ‘0’ (zero).
Adding the product to Bronkhorst FlowSuite When Bronkhorst FlowSuite is started, the software will automatically search for connected instruments. All found instruments are displayed in the left pane. ▶ Use the refresh option to check for newly connected products. Commissioning ▶...
Product status in Bronkhorst FlowSuite The product has Bluetooth, this is disabled by default. To monitor In Bronkhorst FlowSuite, a status icon is displayed next to the the process using a smartphone, Bluetooth must first be activated. product in the left pane. When selecting a product from the left pane, more status information is given in the overview screen.
Maintenance Technical specifications Although no routine maintenance is required, to ensure that the Description Unit Value product works correctly and safely, a regular (visual) inspection is Maximum fluid pressure bar(g) recommended. Permissible fluid temp. °C (°F) 0 ... 50 (32 ... 122) Check prior to each use ▶...
(WEEE). Bronkhorst representative. Bronkhorst will take care of proper dismantling, recycling ▶ To help you quickly, make sure you have the serial number of and/or reuse. By ensuring this product is disposed of the product at hand.
Need help?
Do you have a question about the FLEXI-FLOW Compact FF-C1xS and is the answer not in the manual?
Questions and answers