Wfs Calibration With Digital Meter - Lincoln Electric LN-25 PRO DUAL POWER Service Manual

Hide thumbs Also See for LN-25 PRO DUAL POWER:
Table of Contents

Advertisement

B-7
WFS
CALIBRATION
METERS (K2613-4, K2614-1, -2 & -3 )
To calibrate the wire feed speed, before entering the
set-up menu:
Set the display to the desired wire feed speed (exam-
ple: 400 inches per minute)
Measure the actual wire feed speed (example: 405
inches per minute)
While in the set-up menu, adjust the calibration factor
as follows:
Actual WFS
=Calibration Factor Example
Set WFS
Press the set-up button to exit the set-up menu.
OPERATION
WITH
DIGITAL
405
=1.01
400
LN-25™ PRO DUAL POWER
B-7