End Point Lock-on - 0x60
This command unlocks the unit from a single point lock-on to all other points that are enabled.
When this command is issued, a new TWA start for all points.
Command packet to instrument:
          Command Code (0x60) - 1 byte
Response packet from instrument:
          Command Code (0x60) - 1 byte
VC4™ Date: - 2 bytes
VC4™ Time - 2 bytes
          Status - 1 byte
                    0x00 = Points unlocked
                    0xff = Error
Start Point Lock-on - 0x61
This command locks the unit to one specific point. When this command is issued, all other
points are disabled and locked-on point continues to monitor for concentration and TWA.
Command packet to instrument:
          Command Code (0x61) - 1 byte
          Point Lock - 1 byte
                    bits 0-1: point to lock on
                              00 = Point 1
                              01 = Point 2
                              10 = Point 3
                              11 = Point 4
Response packet from instrument:
          Command Code (0x61) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          Status - 1 byte
                    0x00 = Point locked
                    0x01 = Selected point not enabled
                    0xff = Error
VC4™
164