Honeywell VC4 User Manual page 143

4-point continuous monitor
Hide thumbs Also See for VC4:
Table of Contents

Advertisement

Get Maintenance Dates - 0x34
This command queries the maintenance items.
Command packet to instrument:
          Command Code (0x34) - 1 byte
Response packet from instrument:
          Command Code (0x34) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
          Last Power Down Date - 2 bytes
          Last Power Down Time - 2 bytes
          Last Power Up Date - 2 bytes
          Last Power Up Time - 2 bytes
          Flow Balance Date - 2 bytes
          Flow Balance Time - 2 bytes
          Optics Calibration Date - 2 bytes
          Optics Calibration Time - 2 bytes
          Date Chemcassette™ Replaced - 2 bytes
          Time Chemcassette™ Replaced - 2 bytes
          Date Int. Filter Replaced - 2 bytes
          Date Ext. Filter Replaced - 2 bytes
          Status - 1 byte
                    0x00 = No errors
                    0xFF = Error
Get Point Configuration - 0x35
This command queries an individual point for its current configuration. Command packet to
instrument:
          Command Code (0x35) - 1 byte
          Point Flag - 1 byte
                    bits 0-1: Point Number
                              00 = Point 1
                              01 = Point 2
                              10 = Point 3
                              11 = Point 4
                    bits 2-7: Undefined