bit 0: Printer port enable/disable
0 = Disable
1 = Enable
bits 1-2: Printer report format
00 = Continuous
01 = Summary
10 = Compressed
11 = Invalid
bits 3-5: Printer baud rate
000 = 1200
001 = 2400
010 = 4800
011 = 9600
100 = 19200
bit 6: Printer hardware handshaking (flow control)
0 = Disabled
1 = Enabled
bit 7: Undefined
Get Gas Table Data - 0x3C
This command allows you to view individual gas tables that are contained within the system.
Command packet to instrument:
Command Code (0x3C) - 1 byte
Gas to table to retrieve - 1 byte (0-255)
Response packet from instrument:
Command Code (0x3C) - 1 byte
VC4™ Date - 2 bytes
VC4™ Time - 2 bytes
(filled with zeroes if error)
MDA Gas Abbr. - 6 bytes (not null terminated)
Full Scale - 2 bytes
TLV - 2 bytes
LAL - 2 bytes
LDL - 2 bytes
VC4™
148
User Manual