4. The "MULTI-EJECT" dialog box will open.
Use air on multi-eject
Apply Default Initial
Ejection Volumes
Specify Initial Ejection
Volume
Additional Extra
Volume
5. If tip reuse is disabled (i.e., each tip will be used only once), deselect Use air on multi-eject.
6. Select either Apply Default Initial Ejection Volumes, or select Specify Initial Ejection Volume. If Specify Initial Ejection Volume
is selected, define the initial ejection volume.
7. Define the Additional Extra Volume.
Note: Setting this value to zero may reduce precision. It is recommended to set this volume to half the volume of the aliquot.

Define worktable setup

Plate operations

The plate operations menu provides options to change plate type, change plate function, manage tip availability, and manage
sample banks.
QIAgility Version 4.18 User Manual
Select to blow out air on the final ejection. This limits accumulation of liquid on the tip interior.
This is a default setting when tip reuse is selected, but is an optional setting when single tip
usage has been selected.
Select to apply default initial ejection volumes (i.e., 1 µL for 50 µL tips, and 4 µL for 200 µL
tips).
Select to specify initial ejection volumes. The initial ejection volume is the volume that is
aspirated and ejected into the source well before ejecting into target wells starts.
Use this option to specify the extra volume to aspirate.
06/2023
76