Product Specifications; Electrical; Battery; Environmental - BD Alaris Directions For Use Manual

Gateway workstation
Table of Contents

Advertisement

Product Specifications

Electrical

Protection Against Electrical Shock:
Class I
Supply Voltage:
115-230V, ~50-60Hz
Rating:
460VA (Maximum)
Fuses:
2 x T 4H, 250V
Consult Qualified Service Personnel for replacement of fuses .
AC Outlets:
MDI Tile: 115-230V, ~50-60Hz, 60VA
Protection against fluid ingress:
IP22 - Protected against solid foreign objects of 12 .5 mm diameter and greater, and protected against vertically falling
water drops when tilted up to 15° .

Battery

Type:
Nickel Metal Hydride
Charge time:
With no Pumps attached: Up to eight hours to 95% charge time based on battery life cycle .
With Pumps attached: Up to 16 hours to 95% charge time based on battery life cycle and an 11 Pump load .
Operating time:
60 minutes (With the AC supply disconnected, a fully charged battery has the capacity to maintain the maximum
intended configuration of the Workstation (excluding powering of infusion Pumps) and required components to run for a
minimum of 60 minutes .)

Environmental

Temperature:
Humidity:
Atmospheric Pressure:
Environmental ranges for any attached Pumps should be reviewed to ensure the system is within
specifications prior to operating, transport or storage.
Operating
Transport and Storage
+5°C to +40°C
−20°C to +50°C
20% to 90%
15% to 95%
70 to 106kPa
50 to 106kPa
BDDF00794 Issue 4
Alaris™ Gateway Workstation v1.3.x
Product Specifications
37/46