Data Communications Interface Specifications; Nurse Call Interface; Barcode Reader Interface; Serial Rs232 Interface - BD Alaris Directions For Use Manual

Gateway workstation
Table of Contents

Advertisement

Data Communications Interface Specifications

Nurse Call Interface

Workstation Connector
Plug Type: Binder 09 0978 00 03
Mating Connector
Socket Type: Binder '710' series 99 0975 100 03
Cable Type: Max . cable sheath diameter 4mm .
Isolation: 1 .5kV
Rating: 30V/1A
Description
Pin 1: NC_COM
Pin 2: NC_NC
Pin 3: NC_NO

Barcode Reader Interface

For use with BD supplied barcode reader only.
Workstation Connector Type: Binder 09 0998 00 05
Cable Type: N/A
Description
Pin 1: +5V
Pin 2: TxD
Pin 3: GND
Pin 4: RxD
Pin 5: SENSE

Serial RS232 Interface

Connector Type: D type - 9 pin (female)
Cable Type: Standard RS232 Cable: Length <1 .5m
Isolation: 1 .5kV
Data Rate: 57 .6k baud
Description
Pin 1: N/C
Pin 2: RXD
Pin 3: TXD
Pin 4: N/C
Pin 5: GND
Pin 6: N/C
Pin 7: RTS
Pin 8: CTS
Pin 9: N/C
2
3
1
3
2
4
5
1
1,3
5 4 3 2 1
9 8 7 6
BDDF00794 Issue 4
Alaris™ Gateway Workstation v1.3.x
Data Communications Interface Specifications
34/46