Resetting The Workstation; Fitting An Auxiliary Workstation; Removing An Auxiliary Workstation - BD Alaris Directions For Use Manual

Gateway workstation
Table of Contents

Advertisement

Resetting the Workstation

In the unlikely event that the Workstation needs to be reset, press and hold the
ON Status Indicator is extinguished, release the key then press again to switch the Workstation back on .
If the Workstation still fails to operate correctly after resetting, remove the Workstation from service
and contact Qualified Service Personnel.
Resetting the Workstation is also required to clear any System Fault Indicator.

Fitting an Auxiliary Workstation

1 . Start with both the primary Workstation (option 2) and the auxiliary Workstation (option 1) powered off .
2 . Connect the auxiliary connector (red ethernet port) on the auxiliary Workstation to the auxiliary connector on the
primary Workstation using the ethernet cable supplied with the auxiliary Workstation .
3 . Press the
key once, on the primary Workstation, to switch the Workstation on .
a
4 . Wait till the primary Workstation has completed the power up sequence and tile registration beep sounds .
5 . Press the
key once, on the auxiliary Workstation, to switch the auxiliary Workstation on .
a

Removing an Auxiliary Workstation

If the primary Workstation cannot be switched off for whatever reason then the Pumps fitted to the
auxiliary Workstation can be removed but the auxiliary Workstation must remain on. If the auxiliary
Workstation is turned off or disconnected it may cause the primary Workstation to activate the System
Fault Indicator.
1 . On the primary Workstation, press the
2 . On the auxiliary Workstation, press and hold the
Workstation off .
3 . Disconnect the ethernet cable from the auxiliary connector (red ethernet port) on the primary Workstation .
4 . Disconnect the ethernet cable from the auxiliary connector on the auxiliary Workstation .
key once to switch the Workstation off .
a
key for approximately four seconds to switch the auxiliary
a
BDDF00794 Issue 4
Alaris™ Gateway Workstation v1.3.x
Operation of Workstation
key for at least four seconds until the
a
30/46