Auxiliary Interface

Connector Type: RJ45
Cable Type: CAT5e Ethernet Cable: As required
Isolation: 1 .5kV
Description
Pin 1: Tx+
Pin 2: Tx–
Pin 3: Rx+
Pin 4: N/C
Pin 5: N/C
Pin 6: Rx–
Pin 7: N/C
Pin 8: N/C
NOTES:
Assumes client equipment is non-medical equipment and outside of the patient environment; may be less than 1 .5m if
1
the client equipment is medical electrical equipment in compliance with EN 60601 .
Workstation may be used anywhere inside the patient environment .
2
Either a cross-over or straight through RS232 cable is required . Refer to the Directions For Use of the device to be
3
connected for cable information .
2
BDDF00794 Issue 4
Alaris™ Gateway Workstation v1.3.x
Data Communications Interface Specifications
36/46