Error Messages

Error Message
Reason
Low battery warning
Voltage of battery is too low to operate
device
Overload
Total load exceeds device's maximum
capacity
Counting Error
Signal from loadcells too low
Counting Error
Signal from loadcells too high
Zero count over calibration zero range
+10% while power on
Zero count under calibration zero
range -10% while power on
Program Error
Fault with device software
Action
Replace batteries
Reduce weight on
measurement platform and try
again
Error normally caused by faulty
loadcell or wiring. Please
contact distributor
Error normally caused by faulty
loadcell or wiring. Please
contact distributor
Re-calibration required. Please
contact distributor
Re-calibration required. Please
contact distributor
Please contact distributor