Error Messages - marsden M-930 User Manual

Table of Contents

Advertisement

Error Messages

Low Battery
The scale's alkaline AA type batteries are flat; please replace
the batteries.
Overload
This indicates that the scale's load sensor(s) have been
overloaded. Reduce the loading and retry.
Counting Error
The signal from the load cells is too high. Please
1.
remove any weight from the scale and try to power
on again. If the scale continues to show the error
message, it indicates a fault with the electronics or
wiring.
The signal from the load cells is too low. Please
2.
remove any weight from the scale and try again. If
the scale continues to show the error message, it
indicates a fault with the electronics or wiring.
High/Low Zero Count
The scale is above its zero range. Please remove
1.
any weight from the scale and power on again. If
the scale continues to show the error message, it
indicates a fault with the electronics.
The scale is below its zero range. Check there is
2.
nothing jammed underneath the scale and power
on again. If the scale continues to show the error
message, it indicates a fault with the electronics.
EEPROM Error
This indicates there is a fault with the scale's software and is
normally caused by a fault with the load cell or wiring. Contact
your local service representative.
EU Authorized Representative:
Distributor:
EU Importer:
Manufactured by:
Marsden Weighing Machine Group,
Unit 1, Genesis Business Park, Sheffield Road, Rotherham,
UK, S60 1DX
Marsden Weighing Machine Group Ltd,
The Black Church, St. Mary's Place, Dublin 7, Dublin,
Ireland, D07 P4AX
Charder Electronic Co., Ltd.
No.103, Guozhong Rd., Dali Dist.,
Taichung City 41262 ,Taiwan (R.O.C.)
Page 15 of 16

Advertisement

Table of Contents
loading

Table of Contents