Introduction Thank you for purchasing a Marsden professional medical scale. This is a precision Class III weighing instrument and considerate use will result in many years of accurate weighing. The scale has a maximum load capacity of 300kg which must not be exceeded.
If any inaccuracies occur, please contact your local dealer or service partner. • If you are in the UK, service contracts are available from Marsden to keep your scale accurate and reliable for longer. Call 01709 364296 for more information.
Explanation of Graphic Symbols Separate collection for Caution, consult waste of electrical and accompanying electronic equipment, documents before use in accordance with Directive 2002/96/EC Manufacturer of medical Manufacturing year of device medical device Carefully read user manual before Medical electrical installation and usage, equipment with Type B and follow instructions...
Wall Mounting the Scale You can hang the Patient Transfer Scale on the wall either horizontally or vertically, using the hooks provided. Your hooks will need to be positioned exactly 1400mm or 610mm apart, centre to centre, depending on orientation. Use the Patient Transfer Scale and the hooks to determine the positioning of the hooks on the wall, and mark drill holes.
We recommend fully charging the scale before first use. To fully charge the scale please allow 8 hours. When the low battery indicator on the LCD display indicates that the scale needs recharging, move the M-999 to a location where it can be charged. You will find the charging port on the underside of the scale.
Setting up the Scale With the scale switched off, press and hold until the display shows , Then press 3 times. You will enter Setup for auto power-off, buzzer and backlight. Auto off time: 30m/60m/off Buzzer: On/Off Backlight: ON/OFF Auto power-off: Buzzer on/off: Backlight on/off: To confirm the settings, please press HOLD/UNIT when...
Preparing to use the Scale The Patient Transfer Scale should be used in line with current moving and handling policies. Essentially, it should be used in the same way you would use a transfer board, taking into account of course that you will need to pause for a few seconds during the transfer process, to allow the scale to capture the patient’s weight.
Operation: Basic Functions Switching the Scale on To switch the scale on, press the ON/OFF/ZERO button. When the display shows 0.0 the scale is ready to use. Switching the Scale off Press and hold ON/OFF/ZERO button for three seconds to power the scale down Setting the Scale to Zero If for any reason the scale shows a reading other than 0.0 it can be reset to zero.
Operation: Advanced Functions Using the Hold Function The scale’s Hold function stabilises the weight reading on the display, allowing you to take an accurate reading without fluctuations. With a patient on the scale, press HOLD/UNIT. The display will show ‘HOLD’. NB: Alternatively, you can activate the Hold function before the patient is on the scale.
Guidance and manufacturer’s declaration-electromagnetic emissions The MEDICAL SCALE M-999 is intended for use in the electromagnetic environment specified below. The customer or the user of the MEDICAL SCALE M-999 should assure that it is used in such an environment. Emission test...
Page 12
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MEDICAL SCALE M-999 is used exceeds the applicable RF compliance level above, the MEDICAL SCALE M-999 should be observed to verify normal operation. If abnormal performance is observed, additional measures my be necessary, such as re-orienting or relocating the MEDICAL SCALE M-999.
Distributor: Unit 1, Genesis Business Park, Sheffield Road, Rotherham, UK, S60 1DX EU Importer: Marsden Weighing Machine Group Europe Ltd, The Black Church, St. Mary's Place, Dublin 7, Dublin, Ireland, D07 P4AX Charder Electronic Co., Ltd. No.103, Guozhong Rd., Dali Dist., Manufactured by: Taichung City 41262 ,Taiwan (R.O.C.)