Calibration Testing; Disposal; Technical Specifications - ADC Adtemp 433 User Manual

Non-contact infrared body thermometer
Hide thumbs Also See for Adtemp 433:
Table of Contents

Advertisement

10. Calibration Testing

This thermometer is calibrated at the time of manufacture. If the thermometer
is operated in accordance with these instructions, periodic readjustment is
not required. ADC recommends checking calibration on an annual basis or
whenever clinical accuracy of the thermometer is in question.

11. Disposal

The thermometer contains no hazardous materials. Remove the batteries
before disposal. Discard both thermometer and batteries in accordance with
national or local regulations.

12. Technical Specifications

Measurement unit:
Operating mode:
Measuring site:
Reference body site:
Measurement range:
Accuracy:
Display resolution:
Three-color backlight:
(Color alarm)
Note: Surface mode is always green backlight.
In Body mode,93.2°F to 95.7°F (34.0°C-35.4°C) is green backlight
Facial indication:
Auto power off time:
Measuring time:
Measuring distance:
14
°F / °C
Adjusted mode (Body mode)
Direct mode (Surface mode)
Forehead
Axillary
Body mode: 93.2°F to 109.4°F
Surface mode: 32.0°F to 212.0°F
Body mode:
93.2°F to 94.8°F: ±0.5°F
95.0°F to 107.6°F: ±0.4°F
107.8°F to 109.4°F: ±0.5°F
Surface mode: ±3.6°F
0.1°F / 0.1°C
95.9°F to 99.2°F (35.5°C to 37.3°C):
Green (Normal temperature)
99.3°F to 100.4°F (37.4°C to 38.0°C):
Yellow (Slight fever)
100.5°F to 109.4°F (38.1°C to 43.0°C):
Red (High fever)
Happy Face: 95.9°F to 99.1°F (35.5°C to 37.3°C)
Sad Face: 99.3°F to 109.4°F (37.4°C to 43.0°C)
≤18 seconds
≤2 seconds
0.04in - 5.9in (0.01cm - 15cm)
(34.0°C to 43.0°C)
(0°C to 100.0°C)
(34.0°C to 34.9°C:±0.3°C);
(35.0°C to 42.0°C:±0.2°C);
(42.1°C to 43.0°C:±0.3°C)
(±2°C)

Advertisement