2. Press the up "▲" or down "▼" button to move between Celsius
and Fahrenheit scale.
3. Press the SET button to confirm the new setting.
Please note: The factory default is Fahrenheit.
F2 Fever-Alert Threshold
The fever-alert threshold is the temperature at which the backlit display
color turns red. The default fever alert is 100.6°F (38.1°C).
To change the threshold:
1. With the instrument on, press the SET button twice (until the display reads F2).
2. Press the down "▼" button to decrease by 0.2° F. Press the up "▲"
button to increase 0.2° F.
3. Press the SET button to save.
F3 Silent Mode
Silent mode will keep the unit from beeping.
To turn silent mode on or off:
1. With the instrument on, press and hold the SET three times until the display
reads F3.
2. Press the down "▼" button or up "▲" button to toggle between on or off.
3. Once the desired mode is selected, press the SET button to save.
Factory Reset:
reset. With the unit on, press and hold the
in the display. This will also clear any saved readings from memory.
Adjustable settings can be cleared by performing a factory
MODE
button until
RES
appears
11